NAOJ GW Elog Logbook 3.2
1) Glueing Related Things
1-1) Checked the current situation.
One magnet and one stand-off (wire-breaker) should be glued on the BS mirror.
1-2) Got an adhisive of MasterBond EP30-2
Via Hirata-san I got a set of MasterBond EP30-2.
Now it is storaged at TAMA center room.
1-3) Remaining missing part is weight scale.
Takahashi-san will give us. Yuefan, please receive it from Takahashi-san.
1-4) Glueing work will start from 13:30 of 7th November.
2) Suspension Related Things
2-1) Suspension JIG for double pendulm:
Takahashi-san will send back to TAMA from KAGRA site.
An expected delivered date is 9th November (Wed).
2-2) By using dummy mass, Tatsumi will set the hanging wires.
I confirmed that tungsten wires are existing at TAMA.
Expected date for this wotk is 13:30 of 11th November.
2-3) Scheduled date to be install the BS suspension is 16th November.
I did some calculations on the displacement-noise properties of PR3 by using the yaw-noise data measured in April this year from Akutsu-san on the respective OpLev. These calculations should show how the outcome of a length-sensing OpLev-QPD would be influenced by the noise of itself and its surroundings.
It seems by using a lens with f=500 mm we would reach the requirement of measuring at least 1um shift of the mirror. The big feature around 1Hz is due to the Kumamoto earthquake that happend in the period of data-taking.
However, this are just short-term spectra and still it is not quite clear whether such a lens can be installed.
please note that by using a lens with a smaller focal length, the noise spectrum would increase with a factor (900-f)/f, where 900 is the distance between mirror and lens (in mm as f is given in mm too).
Simulations have been done assuming a misplacement of the WAB of 2mm in x and y direction (z is the direction of the main beam) and 1 degree rotation around x and y axis.
The assumptions of shift and rotation are based on the maximum tolerable misplacement of any baffle in KAGRA.
Result: no important impact on the sensitivity of the interferometer has been found.
However, the effect on the power due to the rotation turns out to be 10000 times stronger than due to the shifting!
I ordered the attached parts for SHG housing.
Expected deliver date is 28 October, 2016.
The installation of an "upgraded" version of the OpLevs toward a low-frequency yaw- and length-sensing kind of OpLev has to be done for all main mirrors in KAGRA. However, the PR2 chamber is different from all the other chambers as there is only one viewport available for the (external) OpLevs.
In order to measure the reflected OpLev beam, we either have to install the beam collimeter inside the PR2 chamber so that the reflected beam goes through the only viewport outside, or to install a mirror in the inner wall (proper positioned), so that the beam, coming from outside, is subsequentally reflected directly toward the viewport.
I checked therefore the feasiblity of the second option by using the actual (as I hope) drawings of the PR2 chamber and the payload. It seems that, taken only the drawings, there is no reason why it shouldn't work. I prepared some pictures to see the paths of the OpLev beam (green line) within the chamber. By installing a mirror, it may be possible to guide the reflected beam toward the viewport.
However, there is still an open question how much space the additional structure on the inner wall (small mirror + mount) will take and whether or not it will collide with the main beam passing PR2 chamber toward the BS.
I packed our SHG housing for modification by a company.
Both PZT (mirror) box and heat sink are remained at NAOJ.
I upload a PDF file of
drawings and Parts list for SHG housing
I got a picture from Manuel.
Sakai-kun set the new computer inside the clean booth, plugged all the instrument and installed the software to control the absorption system.
The new computer is a PC desktop with Windows 7 64bit in Japanese. The license doesn't allow to change the language, but we could install Labview 2016 in English, at least.
We use a GPIB to USB adapter to control and read the Lockin amplifier. We found a library for the sr830 here, and used the examples to build our VIs.
The main VI is "Stanford Research 830 Acquire Measurement - X Y DC Freq.vi" and it uses the subVI "Read X Y DC Freq.vi"
The subVI "Read X Y DC Freq.vi" sends one single "SNAP ? 1,2,6,9" command to the lockin and reads the output through the GPIB port. The output of this subVI is a cluster of data with X,Y,DC and Frequency values.
The main VI initializes the lockin and set the parameters. Then uses the subVI "Read X Y DC Freq.vi" in a timed while loop (100ms for each loop), and put the read values in a shared variable. This VI can also save data in a file.
Keeping this VI running in background, we have the shared variable updated each 100ms, so we can run other VIs to use the values of the shared variable to show real-time charts or to make scans.
The "VI Show X Y DC Freq.vi" gets the values of the shared variable each 100ms and plots the X, Y, AC, Phase, DC and Chopper frequency.
We have two sapphire samples with the dimensions of Tama mirrors.
To measure them, I fixed a horizontal translation stage at the optical board and placed the mirror mount on top of it.
The first Idea was to make a scan by moving the translation stage manually (by using its micrometer screw) and taking a measurement every 5mm.
Then I realized that I could replace the micrometer screw with one of the 3 step motors of the XYZ translation stage of the PCI system, and doing so, I could make an automatized scan of the samples with the original software of the PCI system.
Before making the measurements I made a calibration scan of the bulk reference and found the usual value R=0.5W-1. This value is for silica, to use this calibration for sapphire samples we have to add a factor of 3.34. This correction factor was provided by the SPTS company (Alexandrovski)
The position of the Imaging Unit is corrected by 25.9mm according to the formula SampleThickness*(n-1)/n elog entry 294
I manually moved the tama sapphire sample in order to have the pump at the center of the sample.
The scan gives a good SNR and the result is 50ppm/cm for both the samples with some zone with higher absorption.
I had some trouble with the OXIDE laser 1064nn. When I change the computer to control it, I got a communication error with the serial port RS232.
This is the correct procedure to restart the laser and reestablish the communication with the computer.
1. Please turn off the all the elements. (PC, main switch of Laser, key switch of laser) 2. Disconnect USB-serial port from the PC. 3. turn of PC 4. Connect USB-serial port to PC and check corresponding COM port at device manager. 5. Start software 6. set corresponding COM port 7. set save file for log (This process must be done, otherwise the software does not work properly.) 8. Turn on main switch of the laser ==> after a few moment the communication between laser and PC would be established. ==> If the communication would work properly, turn on the key switch.
The accuracy of previous measurements could be affected by stray light coming from the high power pump laser.
I carefully checked the photo-detector ad I found two small filters inside a threaded pipe attached in front of the detector. They are supposed to suppress the stray light signal.
Those filters are Heat Absorbing Glass:
- 3 mm-thick KG-3 glass. It filters out 1064 nm http://www.edmundoptics.com/document/download/352659
- 2.5 mm-thick R-60 glass. It is the red-color filter used to block most of daylight. http://www.edmundoptics.com/optics/optical-filters/longpass-edge-filters/longpass-glass-color-filters/66043/?print=Pdf
In order to check if the pump stray light is well stopped by the filters, I switched off the probe (to avoid to have any true absorption signal, in purpose) and see if there is a different signal with and without the pump laser.
Since switching the probe OFF makes the DC signal almost 0, I don't divide the AC signal by the DC.
I took 1h of data with the pump ON; and 1h of data with the pump OFF. I attach the plot of the raw AC signal (X and Y) in the two cases, red and blue clouds of points.
The standard deviations of the X and Y signals are 1microV in both clouds. The difference between the means of the two clouds is 0.2microV.
Comparing with the order of 50microV of last absorption measurements, I conclude that pump stray light don't contribute to the signal.
with Mathematica software, I derived the Image Unit position correction formula:
SampleThickness*(n-1)/n
I used the ABCD matrixes of the absorption bench system and the equations for the q parameter.
I cannot attach a .nb file, so I attach the pdf of it
According to the basic theory of the PCI method, the heated area of the sample makes a phase shift in the probe beam; this perturbation is small and can be treated as a gaussian beam which interferes with the main beam; the maximum of the interference is detected when the PD is at the Rayleigh length of the gaussian perturbation, which can be calculated using the waist of the perturbation, which is the pump beam size.
The approximation of the perturbation to a gaussian beam is valid at first order, but for a fine tuning of the detector position, it might be not a good approximation. I consider this because, when I correct the position for the thick sample (as in elog entry 291), I notice that the calibration value is not the same (as expected from the simulation). A possible explanation might be that when I put the thick sapphire sample and correct the Image unit position, the detector is not in the interference maximum anymore. So, I maximize the signal as a function of the detector position experimentally, by moving the Image Unit with the micrometer screw. I do it for the reference sample alone, and for the reference sample with the sapphire sample behind it. Then I compare the two maxima positions in order to find the best position correction for thick samples.
First two plots show several scans of the reference sample for each position of the Image Unit, with and without tama sapphire sample. In last plot, I took the middle value of each scan and plot it as a function of the Image unit position. 35mm is the closest position of the IU to the sample, 0mm is the furthest IU position. To move it further it's necessary to unmount the IU micrometric translation stage. The theoretical distance between the maximums is 26mm. and the maximums should have the same value.
The plot shows maximums position accuracy of about 5mm, but in the case of reference + tama sapphire, it's not clear wether the maximum is below the position 0mm on not. The problem is the maximum value, it should be the same but for the reference alone the value is 0.1 and in the other case is 0.04. More than a factor of 2
I repeated the measurement of entry 289, but this time, I put the tama-size sapphire sample in a position such that the probe beam is crossing it but the pump beam is not, so I avoid any back reflection of the pump. I also correct the position of the Image Unit according to the formula I wrote in entry 290.
I moved the base micrometer by the distance L x (n-1)/n - 1mm away from the sample. L is the path inside the sapphire sample, which is the thickness 60mm divided by cos(6°), 6° is the probe incidence angle, n is the sapphire refractive index 1.76. Therefore the displacement is 25.05mm
The attached plot shows the comparison of the two scans.
The result is not as good as in the simulation. The DC is different because there is an additional reflection when I put the second sample, but the AC/DC should be equal in the two cases.
I think I have to figure out why. Maybe the positioning should be slightly different, so I will try to find he optimal position.
According to the basic theory of the PCI method, the heated area of the sample makes a phase shift in the probe beam; this perturbation is small and can be treated as a gaussian beam which interferes with the main beam; the maximum of the interference is detected when the PD is at the Rayleigh length of the gaussian perturbation, which can be calculated using the waist of the perturbation, which is the pump beam size.
The approximation of the perturbation to a gaussian beam is valid at first order, but for a fine tuning of the detector position, it might be not a good approximation. I consider this because, when I correct the position for the thick sample (as in elog entry 291), I notice that the calibration value is not the same (as expected from the simulation). A possible explanation might be that when I put the thick sapphire sample and correct the Image unit position, the detector is not in the interference maximum anymore. So, I maximize the signal as a function of the detector position experimentally, by moving the Image Unit with the micrometer screw. I do it for the reference sample alone, and for the reference sample with the sapphire sample behind it. Then I compare the two maxima positions in order to find the best position correction for thick samples.
First two plots show several scans of the reference sample for each position of the Image Unit, with and without tama sapphire sample. In last plot, I took the middle value of each scan and plot it as a function of the Image unit position. 35mm is the closest position of the IU to the sample, 0mm is the furthest IU position. To move it further it's necessary to unmount the IU micrometric translation stage. The theoretical distance between the maximums is 26mm. and the maximums should have the same value.
The plot shows maximums position accuracy of about 5mm, but in the case of reference + tama sapphire, it's not clear wether the maximum is below the position 0mm on not. The problem is the maximum value, it should be the same but for the reference alone the value is 0.1 and in the other case is 0.04. More than a factor of 2
I got a formula to correct the positioning of the detection unit.
SampleThickness*(n-1)/n - 1mm
In the case of 60 mm-long Sapphire the shift is 60*0.76/1.76 - 1 = 24.9 mm
I applied to the last simulation and it looks working.
I tried to reproduce experimentally the situation of last simulations (Elog entry 288), but the data are a bit confusing. I think I'm missing something.
Here is what I did:
I made a scan of the bulk reference sample (blue data in the plot), and then I placed the tama-size sapphire sample in front of it, at about 15mm, and repeated the scan (black data in the plot).
Pictures of the setup: 2,3
With the sapphire sample, I noticed that:
- the signal is very low
- the phase signal has a strange shape.
This phase shape makes me think that the sapphire sample might reflect the transmitted pump beam back and then heat the reference sample again. This would change a lot the signal shape. So I placed the sapphire sample further, at about 45mm, and a bit tilted (about 3°) so the reflected beam would not go back on the measured point. Then I repeated the scan (red data in the plot). Picture 4. Then I tilted more the sample (about 6°) and repeated the scan (green data). I couldn't tilt the sample more because, otherwise, the probe beam would go out of the prism mirror.
Every time I moved the sapphire sample, I had to tune the focusing lens to center the probe beam on the detector.
I feel confused by the fact that changing the position of the sapphire sample make so different signals.
In order to calculate the calibration in the case of thick samples, I simulated the scan of the bulk reference silica sample, then I simulated the same thing but adding a 6cm-thick sapphire sample on the probe path.
I already calculated the probe beam size on the detector for different sample thicknesses (Elog entry 263)
In the first plot, there is the scan of the bulk reference sample (red line), and also the same scan but with a 60mm thick sapphire after the sample (blue line). Adding 60mm of sapphire after the sample changes the optical path of the probe and makes a different signal.
The calibration value is taken at z=2mm, and there is a factor of 5 of difference between the two cases.
The second plot is a 2x2 matrix of plots and it shows the probe beam profile at the detector, when the sample is at z=2mm. First column of plots is the beam profile. Second column is the interference pattern, from which the AC signal is calculated. The first row is the case with only the bulk reference sample and the second row is the case with the bulk reference plus the tama-sized sapphire sample after it. The white rectangle is the profile of the photodetector. The unit of axis is m, the photodetector is 1mm large.
I think it is necessary to adjust the reimaging of the detection unit. I will try to get better signal in the simulations by changing some distances among the components and also changing the focal lengths of the lens and of the sphere.
I will also reproduce the same experimental configuration of this simulation, putting the tama-size sapphire sample after the reference sample, and making a real scan. This will also be a test to see how reliable is the simulation.
I got a formula to correct the positioning of the detection unit.
SampleThickness*(n-1)/n - 1mm
In the case of 60 mm-long Sapphire the shift is 60*0.76/1.76 - 1 = 24.9 mm
I applied to the last simulation and it looks working.
with Mathematica software, I derived the Image Unit position correction formula:
SampleThickness*(n-1)/n
I used the ABCD matrixes of the absorption bench system and the equations for the q parameter.
I cannot attach a .nb file, so I attach the pdf of it