ManuelMarchio - 17:24, Wednesday 17 August 2016 (294)
Comment to Simulation of bulk reference sample scan. Adding a thick sapphire sample on the probe path (Click here to view original report: 288)
with Mathematica software, I derived the Image Unit position correction formula:
SampleThickness*(n-1)/n
I used the ABCD matrixes of the absorption bench system and the equations for the q parameter.
I cannot attach a .nb file, so I attach the pdf of it
Non-image files attached to this comment