LOG-IN
Displaying reports 2641-2660 of 3128.Go to page Start 129 130 131 132 133 134 135 136 137 End
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 16:02, Monday 10 July 2017 (412)Get code to link to this report
Filter cavity locking loop characterization

In the past days we tried to characterize the locking loop of the filter.

The loop transfer function for the filter cavity (sketched in figure1) is compose by different blocks

  • G1 [Hz/V]  = piezo actuator                      
  • G2 [Hz/Hz]  = SHG                                  
  • G3 [Hz/W]  = cavity
  • G4 [W/V]  = photodiode + demodulation
  • H [V/V]= servo 
 
if we define G = G1*G2*G3*G4  the open loop trasfer function is simply  H*G
 
In the loop scheme are shown the points where we can read the signal and the points when we can inject noise. By choosing the appropriate combination of observation and injection points we have tried to measure different parts of the loop transfer function. In particular
 
1)      H*G   ->  OPEN LOOP TF
  • inject noise on perturb 
  • measure EPS1/EPS2
NOTE:  We perform this measurement with a swept sign. (See picture 2) It allows to measure the UGF and the phase margin. The measure is not good at low frequency where the gain on the loop is higher. Unfortunately at these frequency where are not able to inject enough noise to dominate the error signal without unlocking.
 
2)       H   -> ELECTRONIC TF
  • inject noise on perturb
  • measure piezo mon/EPS2
NOTE: We performed the measure with a swept sine (See picture 3). Unfortunately I was not able to find a way to monitor the coherence between the two channels while performing a swept sine, so I don't know how much we can trust the measurement. It seems to be flat after the cavity pole (1.5 kHz) as it should be.
We have also perform the TF without injecting additional noise and assuming that the laser was sufficiently high. A result of the measure is plot in figure 4. I have also measures the coherence between the two channels ( shown in picture 5) which shoud tell in which regions the measuremts is more reliable.
 
3)         G  
  • inject noise on RAMP
  • measure  EPS2/piezo mon
NOTE: The blocks composing G are basically frequency independent up to few tens of KHz except for the cavity which should have a pole at 1.45 KHz.
Being able to fit the pole frequency would allow a measurement of the cavity finesse [ f_p = c/( 4*L*F)]
Also in this case, the amount of noise we could inject without unlocking was not high enough to provide a clear measurement ( we tried with with noise and swept sign). The obtained TF is shown in picture 6. it is not possible to extrapolate a value for the cavity pole.
 
In the last picture there is a scheme of the rampeauto done by Pierre prat with a summary of the gain of each channel.
Images attached to this report
412_20170710084417_07.png 412_20170710084439_tfopenloop.png 412_20170710084502_31.png 412_20170710084517_04.png 412_20170710084529_cohe.png 412_20170710084628_27.png 412_20170710084706_rampeautosummary.jpg
Comments related to this report
EleonoraCapocasa - 07:43, Wednesday 26 July 2017 (536)

The amplitude of the loop transfer functions plotted so far are actually the square of the real amplitude. The problem comes from the way I treated data saved by the spectrum analyzer. Each file is composed of 3 columns: frequency, real part (a) and imaginary part (b) of the TF.  Of course amplitude and phase are recovered by doing:

Amplitude = sqrt (a^2 +b^2)

Phase = angle (a+i*b)

Due to an oversight, I had replaced the sqare root with the absolute value in the amplitude computation. This explain the unexpected behaviour (1/f^2 instead of 1/f) of the openloop TF around the UGF. 

We will upload soon new TFs measurements (taken by Yuefan and Marc on monday night) properly plotted.

R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 15:02, Monday 10 July 2017 (428)Get code to link to this report
Occasional excess of noise in local controls

The activity of locking characterization of the past days pointed out some issues that are worth to be reported.

As expected the stability of the lock is strongly affected by the performance of local controls. The occasional presence of spikes or excess of noise has been observed on the local controls of BS, PR and INPUT mirror.
 
In the following table I report the normal angular variation for each suspension that can be used as reference values to detect excess of noise. 
 
  open loop (urad) closed loop (urad)
PR yaw 1-2 0.5-1
PR pitch 4-8 2-3
BS yaw 3-4 2-3
BS pitch 5-6 4-5
INPUT yaw        2-3 1-2
INPUT pitch 4-5 3-4
END yaw 2-3 1-2
END pitch 4-5 3-4
 
The value of the angular variation in micro radiant is reported on the top of each error signal plot in the labview control VIs and it is computed measuring the difference between the maximum and the minimum of the error signal over a second and multiplied it by the respective calibration. 
 
PR
We observed spikes on pitch and yaw signal on the morning of the 28th june for the first time.  Since this spikes where also present when disconnecting the signal in input to the ADC we suspecting the ADC itself or the softwere. We changed the ADC with a spare one from the labview control of the west arm without notice any improvement. ( We put back the original one.)
The spikes has disappeared the day after and I have never observed them again.
 
BS
Similar kinds of spikes are are sometimes appearing (bringing the motion up to 15-20 urad). They disappear when I disconnect the input cableto the ADC and are not present on PR signals which is using the same ADC. Maybe they are not coming from the electronics.
 
INPUT 
Excess of noise can be observed periodically. It bring the mirror motion up to 20-30 urad making impossible to keep a proper alignment. I have observed that they disappear switching on and off the  laser of the INPUT local control. I have changed the laser power supply but it didn’t solve the problem.
 
END
Nothing to report
 
My statistics is too small to extrapolate a trend for these noise occurrences but if you spend few hours working on the locking you are likely to have some noisy periods of BS and or INPUT. 
 
Investigation on the origin on the issue should be done in order to fix it. The possibility to record and store long periods of data will help this.
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 01:46, Saturday 08 July 2017 (429)Get code to link to this report
IR and green beam simultaneously aligned in the cavity

After a long alignment work we were able to make the IR and the green beam flashing in the cavity at the same time. In this video are shown superposed flashes in trasmission both of the green and the IR (the green beam as been cut at the second 5).

In this configuration we were able to lock the cavity on the green TEM00 but we coudn't check the IR condition since there was too much green light transmitted by the dichroic mirror before the camera installed on the end bench. 

After solving this issue and installing the AOM we will start to look for the frequency shift needed to have both the IR and the green resonant on the fundamental mode.

R&D (FilterCavity)
Print this report.
YuefanGuo - 15:50, Friday 07 July 2017 (526)Get code to link to this report
AOM calibration and green power fluctuation
We tried to calibrate the AOM on the mode cleaner path. With a 100mm lens to focus the beam, followed the operation manual. At first, we put the AOM around the beam waist position and adjusted the beam position at the input and output port to have the highest transmission rate before we powered it up. And also we set the position of the beam on the screen we put as close as we can from the bench edge, so this will be our 0 order position.

Then we powered up the AOM, there was the other orders appearing, the one close to the 0 order is the 1st order which we need. We put a PD at the 1st order, connected it to the oscilloscope and started to turn the AOM to have maximum power in the 1st order. After finding the best position of the AOM, we tried to change the RF power we were sending to increase the power more. But when we tried to do this, we found out the power fluctuation is about 40-50% percent, it was much more than the tolerance.

We put PD both in the infrared path and the green path to check the fluctuation of the green is coming from the infrared or not. Then we found out since the PD aperture is quite small, so it will increase the degree of uncertainty we saw. So we changed the PD to the powermeter which has larger aperture. This time the infrared is quite stable and the green is fluctuating but much less than before. We guessed that the fluctuation of the green is coming from the changing of the SHG cavity alignment when people moving around or touching the bench. We are going the install the MZ and stabilize the green.

Then other thing we checked is that the SHG loop gain we are using now is 50. But with this gain,if we changed the power sent to the PZT, the cavity cannot pull the error signal back to zero when the cavity is locked. If we increase the gain, until it reaches 10000, there will be no oscillation appear, the cavity will be locked more stable, but one problem is that it will be quite easy also lock on the other mode. We will try to find a the optimal gain considering also the filter cavity.

KAGRA MIR (Absorption)
Print this report.
ManuelMarchio - 16:27, Thursday 06 July 2017 (525)Get code to link to this report
Status report

- In order to increase the pump power safely, we cover the laser path with aluminum black walls and black paper. 

- We also cover the probe part in order to stop air flow from hepa filters which is a possible noise source. Kuroki-san helped me on this tasks.

- I put a small translation stage below the half ball in the imaging unit in order to adjust better the alignment and I tried many times the alignment until I found a good signal for the surface reference sample, comparable with the one of last year. Also the bulk reference sample gives almost the same signal as last year.

- The parts for the 1310nm probe laser were delivered, I glued the golden prism mirror on a half inch post, I'm waiting for the glue to cure under the neon lamps

Images attached to this report
525_20170705145827_surf.png 525_20170705145831_bulk.png 525_20170706092407_screenshotfrom20170706162334.png 525_20170706092418_screenshotfrom20170706162152.png 525_20170706092431_screenshotfrom20170706162120.png 525_20170706092439_screenshotfrom20170706162057.png 525_20170706092449_screenshotfrom20170706162018.png
R&D (FilterCavity)
Print this report.
YuefanGuo - 12:02, Tuesday 04 July 2017 (524)Get code to link to this report
Analog signal receiver changed
Yesterday we found out the analog signal sent through the fiber from the end room had some problem. No matter what signal we sent, it was always -7.5V. At first we thought maybe the problem was from the fiber, so we decided to change the fiber, but the receiver(pic 1) we are using now have a special port, other fibers cannot be used. Other possibility is the receiver or the transmitter had some problem, so we took the transmitter from the west end to the south end. Sent a signal from the west end transmitter, the original fiber and the south end receiver, with this configuration we have a offset about 70mV. So we also changed the receiver to the west end receiver, now everything is fine.
Images attached to this report
524_20170704050057_img3899.jpg
DECIGO (General)
Print this report.
EleonoraCapocasa - 00:50, Tuesday 04 July 2017 (523)Get code to link to this report
Comment to FIlter cavity experiment: Offset in the filter cavity reflected signal (Click here to view original report: 521)
[I report about the investigations done by Pierre Prat about the offset in the error signal]
 
The mixers ideally have a zero offset when the LO and RF signals are in quadrature, if it is not exactly the case, they can have an offset that can be a few mV, varying with the frequency of LO.
 
We saw this afternoon that the 1.9 MHz low-pass filter (LBP1.9) was not terminated by a 50 Ohms as it should be. We had an offset of the order of 200mV on the output monitor EPS1. By charging the filter with a 50 Ohm load, the offset reduced to 50mV at the EPS1 output.
 
Between input DETECT F (error signal) and outputs EPS1/EPS2 (error monitors) there is a fixed gain (independent of the attenuation) of 15.9.  See attached picture.
 
The offset at the output of the mixer (measured with the oscillloscope) is therefore about 3mV, which is normal. 
 
Pierre has verified that, once the filter BLP1.9 is loaded by 50 Ohm, this offset is stable and does not depend on the ground conditions. 
Images attached to this comment
523_20170703174745_signal.png
R&D (FilterCavity)
Print this report.
Matteo Barsuglia and Eleonora Capocasa - 22:51, Sunday 02 July 2017 (522)Get code to link to this report
Calibration of the PDH error signal

The PDH filter cavity signal has been calibrated injecting a line at 28 kHz (above the ugf ~ 10 kHz of the loop) on the “ramp” input of the electronic servo. The ramp input is summed to the PZT correction signal.

The amplitude of the 28 kHz line in Hz is obtained using the formula:

S_Hz = V_RMS  (V) * sqrt(2)*100*2e6 Hz/V = 1.25e-6*sqrt(2)*100*2e6 = 353 Hz 

Where V_RMS is the line amplitude measured by the Agilent spectrum analyser. The factor sqrt(2) is obtained to pass from the V_RMS to the line amplitude (the factor has been also experimentally verified looking the same line with the spectrum analyser and the oscilloscope).

The factor 100 is the reduction of the PZT_moni output . 2e6 Hz/V is the calibration of the PZT after the SHG.

Measuring the line at 28 kHz in the error signal and compensating for the cavity frequency pole is it possible to find the calibration factor K in V/Hz. The formula used is :

S_V  = K(V/Hz)* S_Hz /sqrt(1+ (f/f_0)^2)  

where f_0 = 1.5 kHz and S_V = sqrt(2)*38.9e-3 V 

--> K = 2.9e-3 V/Hz

which seems to be in agreement with the calibration obtained looking the PDH signal when the cavity is freely swinging. In that case we see a peak-to-peak of the PDH of ~ 4 V for 1.5 kHz of the cavity line which correspond to a K = 2.7e-3 V/Hz. Note that when the cavity is freely swining we have also rining effects which can perturb this measurement. 

We have also checked that reducing the frequency of the line sent to the PZT (with the same amplitude) to 14 kHz, the amplitude of the line of the error signal is multiplied by 2, as expected given the cavity pole. A more quantitative analysis (fully taking in account the effect of the loop) is necessary to check the position of the cavity pole.

Another test was to increase the amplitude of the line by a factor 10, thus having a 29 kHz line with amplitude of 3 kHz (two times the cavity width of the cavity). The cavity stays locked and the calibration factor measured is the same with the one measured with the line with an amplitude of 300 Hz. Increasing further the amplitude of the 28 kHz line to ~ 7 kHz (4 times the cavity linewidth) makes the lock more fragile, and sometimes the cavity unlocks. Moreover, an oscillation with a frequency of ~ 1 Hz appears in the error signal (but it is not accompanied with a similar oscillation in the transmitted power).

R&D (FilterCavity)
Print this report.
Matteo Barsuglia - 22:40, Sunday 02 July 2017 (521)Get code to link to this report
Offset in the filter cavity reflected signal

We observe that the PDH filter cavity signal has an offset of ~ 170 mV. See picture.

The offset is present even when the 78 MHz signal sent to the EO modulator is swithed off (and the 78 MHz sent to the local oscillator is ON). When both signals are OFF, we see a slowly varying offset between 200 mV and -200 mV, which also have an higher frequency oscillation. To be investigated. 

Images attached to this report
521_20170702153911_img3171.jpg
Comments related to this report
EleonoraCapocasa - 00:50, Tuesday 04 July 2017 (523)
[I report about the investigations done by Pierre Prat about the offset in the error signal]
 
The mixers ideally have a zero offset when the LO and RF signals are in quadrature, if it is not exactly the case, they can have an offset that can be a few mV, varying with the frequency of LO.
 
We saw this afternoon that the 1.9 MHz low-pass filter (LBP1.9) was not terminated by a 50 Ohms as it should be. We had an offset of the order of 200mV on the output monitor EPS1. By charging the filter with a 50 Ohm load, the offset reduced to 50mV at the EPS1 output.
 
Between input DETECT F (error signal) and outputs EPS1/EPS2 (error monitors) there is a fixed gain (independent of the attenuation) of 15.9.  See attached picture.
 
The offset at the output of the mixer (measured with the oscillloscope) is therefore about 3mV, which is normal. 
 
Pierre has verified that, once the filter BLP1.9 is loaded by 50 Ohm, this offset is stable and does not depend on the ground conditions. 
R&D (FilterCavity)
Print this report.
Matteo Barsuglia and Eleonora Capocasa - 22:38, Sunday 02 July 2017 (520)Get code to link to this report
Measurement of the PZT correction spectrum

We have measured the spectrum of the PZT correction signal sent to the laser when the cavity is locked, using the output PZT_mon (1/100 of the PZT correction signal). The spectrum is in the attached plot.Since in this region the gain of the loop is very high, the signal is proportional to the cavity length/frequency noise. 

The calibration is 1 MHz/V (given by the manufacturer). 

at 100 Hz we have ~ 700 nV/sqrt(Hz) corresponding to 70 Hz/sqrt(Hz), at 1 kHZ we have 100 nV/sqrt(Hz) corresponding to 10 Hz/sqrt(Hz)

The shape of the spectrum is compatible with the free running laser noise ~ 7-10 kHz /f  Hz/sqrt(Hz) up to a few kHz. According to aother measurement, after ~4 kHz the spectrum is limited by a flat noise, which is compatible with the noise of the 100 kOhm resistor at the output of the PZT_moni signal. For f<10 Hz probably the mirror control noise and the seismic noise are limiting the spectrum. 

We also see several 50 Hz harmonics. It is not clear if this harmonics can be reduced rearranging the grounds and if they have an impact on the RMS of the error signal of the filter cavity locked. To be investigated.

Images attached to this report
520_20170702152620_img3173.jpg
R&D (FilterCavity)
Print this report.
Eleonora Capocasa and Matteo Barsuglia - 15:25, Friday 30 June 2017 (475)Get code to link to this report
Filer cavity lock characterization campaign

Summary of yesterday night work (thu 29-->fri 30). The goal was to make a characterization campaign for the cavity lock, in order to make it more stable.

1) Beam stability

In the past we observed an evident jitter of the beam. From a comparison of the spectra we were convinced that this was caused by the residual motion of BS and PR. In the past days we where able to improve the stability by improving the local control filters (a dedicated entry will follow). 

We observed that the beam direction (observed by misaligning the input mirror) was drifting and we decided to test a new strategy to keep the mirror position. We change the local control filters in order to avoid to gain at low frequency (we changed a pole at 0.1 Hz with a double zero at 0.1 Hz and we controlled the mirror position not by adding an offset of the loop but simply sending a DC signal to the coils.

We coudn't see a major improvement in the performances.

We also observed the intermittence presence of spikes in the error signals from BS and PR which makes difficult to keep the cavity alignment.

Eventually the old controls (with integrators at low frequency) were restored.

2) Laser servo gain transfer function 

We have set the gain of the servo in order to have ~10 kHz bandwidth. See the transfer function in fig.1. (in 1/f^4 mode)

At a first look, the TF behaves as expected. The data have been stored in the floppy disk and they will be compared with the model. The phase margin at ~10 kHz is about 40 degrees.

The transfer function has been measured with the Agilent 35670A spectrum analyser, with a swept sine with 50 mV ptp. 

3) Servo parameters 

- modulation depth = 1 V pp at 78 MHz (reduced with respect to before). This should correspond to a modulation depth of m= 0.185 rad.

- LO = 8.5 Vpp at 78 MHz (increased with respect to before)

- Demodulation phase = 111 deg

--> With this data the error signal is 3-4 V ptp, for a transmited signal of ~ 3-4 V depending on the alignment of the cavity  (note that we did not checked the green laser power yesterday night)

- attenuation of the input signal =9.1 

- PZT gain = 0.7

- thermal control gain = 3 

- Threshold on the transmitted signal ~ 2 V

4) Auto-relock

With this configuration the cavity automatically locks when the transmitted power crosses the resonance. When the cavity unlocks, it relocks automatically. Note that the servo is always in the 1/f^4 configuration. The video shows the cavity locked, then the input mirror is on purpose misaligned, then it is re-aligned and the cavity re-locks.

5) Stability 

During yesterday night lock the cavity was very stable. The plots 2 and 3 show the transmitted power (in cyan) and the error signal (in yellow) for 500 s. No actions were performed to realign the cavity on the second plot. Max transmitted power was ~ 4 V. 

Images attached to this report
475_20170630073320_tf1.png 475_20170630075035_longterm1.png 475_20170630075042_longterm2.png 475_20170630075207_pzt.png
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 11:16, Thursday 29 June 2017 (503)Get code to link to this report
Filter cavity first lock

On Tuesday 27th june we managed to lock the laser on the filter cavity length. 

In the first attachment there is a plot of the transmitted power during the lock acquisition, in the second there is a picture of the transmitted beam when the cavity is locked. A short movie of the the lock acquisition can be seen here.

https://www.dropbox.com/s/pinj73ewrk9kgk0/locking.mp4?dl=0

Images attached to this report
503_20170629041426_lock3.png 503_20170629041441_lock.png
KAGRA MIR (Absorption)
Print this report.
ManuelMarchio - 16:50, Wednesday 28 June 2017 (513)Get code to link to this report
Imaging unit alignment

Friday, June 23, 2017

Following the following procedure, I aligned the Imaging Unit for the HeNe probe beam.

1) move IU micrometer closer to the end of translation which would give you enough translation range to move the whole IU farther away in case you test thick objects;

2) place front lens of the IU to get the probe beam at its center, then place gold-coated half-ball approximately at the focus of the front lens;
3) with a knife edge half-close the probe beam at a location of a Rayleigh length of perturbation behind the pump/probe crossing point;
4) place a paper screen at the detector position to observe the probe spot reflected by the half-ball (may place the screen at a longer distance to have a larger probe spot);
5) changing the distance between the front lens and half-ball find a moment when you see a sharp image of the knife edge.

That will complete rough alignment of the IU. The fine tuning is done by maximizing AC signal coming from the surface calibration piece. For that, try different micrometer positions around one you started with. For every position you have to center the probe (maximize the DC), maximize AC if needed. The maximum R should be close to the original R for the surface calibration. Then make scan with the bulk calibration piece.

According to the theory the signal is maximum when the detector at the Rayleigh length of the perturbation, experimentally we can check this changing the position of the blade and aligning again the imaging unit and measuring the signal. So, in order to maximize the signal, I repeated the procedure changing the position of the blade from 18mm to 12mm and 6mm but I got a lower signal, so I aligned it back to 18mm.

The absorption signal of the reference is similar to original value (the one we had since we bought the system) even if I changed by the 20% the waists of pump and probe,

Parameters: LD current = 0.8A, power without sample = 33mW

Images attached to this report
513_20170628093935_surfrefjune23.png 513_20170628093940_bulkrefjune23.png 513_20170628094809_screenshotfrom20170624191701.png 513_20170628094813_screenshotfrom20170624191633.png 513_20170628094817_screenshotfrom20170624191608.png 513_20170628094820_screenshotfrom20170624191541.png
R&D (FilterCavity)
Print this report.
YuefanGuo - 14:54, Wednesday 28 June 2017 (518)Get code to link to this report
Few works did with the electrical board
1. Video board

We changed the layout of the end bench, added one more mirror after the beam splitter to divide the beam into two, one is received by a screen and a camera was set to look at this screen, the other goes directly into the CCD. The mirror we used has the maximum reflection when it is putting in 45 degree, so we turned it a bit in order to have large enough power also for the transmission. The one received by the screen we took it as a reference for the alignment of the cavity and the CCD is to find a good mode matching. In this configuration we will need three images sending back from the end room at the same time(two on the end bench and one for the second target), the electrical board we are using now for the video only has two channels. So we took the same board from the west end, but it seems one channel of that board is broken, but it is enough for our current need. The fiber used to send the video signal now is '1-13,1-14,1-15,1-16', each board need two fibers. Also each monitor only can receive two channels, so we also took the monitor at the second target of west end to the central room.

2.Analog signal board

On the reflection path of the beam splitter on the end bench we put a PD before and use the receiver box to send this signal to central room for locking the cavity. But since the the aperture of the PD is limited, the spectrum we saw will also be effected by the alignment of the beam. So yesterday we changed the PD to PSD in order to have the information of the beam position. But in this way, we need to send back three analog signals together. We found two board for this and one of them have four channels. We tested the board with a sine wave sent from the end room board, and received it in the central room. The fiber they used for this board is too short to connect into the rack('1-11,1-12'), but the fiber system of TAMA is too complicated, so we just simply changed with longer spare fibers, now we are using '3-13,3-14' for this board.

From the board it seems we can change the gain and offset of each channel, but when we sent a 4V peak to peak signal, it pretty hard to see if it is changed or not, so maybe we can only changed in very small scale. Also it seems four channels all have different setting offset and gains, but we need further check about this.
R&D (FilterCavity)
Print this report.
YuefanGuo - 14:13, Wednesday 28 June 2017 (517)Get code to link to this report
Laser current changed
After the installation and the adjustment of the input mirror finished,we tried to align the beam in order to receive green light at the end bench. We tried for pretty long time but even we looked directly at the second target with eyes, we can only see the reflection from the tube on the edge of the target which is more reflective than the central part.

So we decided to increase laser current to have higher power in infrared and also in green, the current we used before is 1.040A which gave a 8mW-green power at the end of the bench, now we increased the current to 1.2A which gives out a three times green power around 27mW at the end of the bench.

With this power we easily found the beam at the second target and also received it at the end bench.By looking through the window of the end chamber, we better adjust the beam on the end mirror and let it pass more or less the center of it, luckily we got the flash of the cavity with the first try.

The other thing we did before closing the chamber is that we sent the beam out of the chamber to the corridor again after the BS, tried to superpose the green and infrared both in the near field and the far field with the last two infrared steering mirror on the bench. Although considering the air fluctuation in the corridor, two beams moved a lot, but we did our best to make them overlap at 300m.
R&D (FilterCavity)
Print this report.
YuefanGuo - 18:12, Monday 19 June 2017 (512)Get code to link to this report
Telescope adjusted
With the data measured at the end of the bench, we found the beam waist and position. Then we did the simulation with the 2-inch mirror, PR suspension, input and output mirror. The position and the focal length of these mirrors are as below:(The origin is set as the steering mirror which is 85cm from the end of the bench)

2-inch mirror: f=-30cm z=4.51m
PR suspension: f=3m z=7.235m
Input mirror: f=-218.35m z=12.135m
End mirror: f=-218.35m z=312.135m

With these values, the simulation result shows that the beam will keep diverging in the arm which is not what we want(pic1). So I did some calculation, we need to move the 2-inch 3.9cm backward to get the beam waist around 150m inside the arm. (pic2)

The other reason why we should move 2-inch mirror is after we installed the real input mirror, we got green reflect back from the from surface of that mirror. But it seems the reflected beam is much larger the original one.

Before moving the mirror, we set two reference point, one is the position of the beam in front of the input mirror, the other is the beam reflected back to the bench. Also after the input mirror we put two aluminum mirror to reflect the beam into the corridor to check the beam propagation. At first we only tried to move the picomotor of the 2-inch, but even we when we finished all the range, nothing changed. So we moved by hand little by little. Every time we moved, we recovered the beam with pitch and yaw of the 2-inch on two reference point and check the beam size at the entrance of the corridor and bench. After moving about 3cm, the reflected beam size seems fine, so we followed the beam inside the corridor, found it has beam waist around 150m, and the size is from 1cm to 2cm which is acceptable.

Then we took the reference of the BS chamber like what we did to the PR, with two aluminum mirror, the beam transmitted by the 2-inch and the BS suspension have been reflect out of the chamber.



Images attached to this report
512_20170619111122_i3cimgpshorig.png 512_20170619111209_wholepathfin.png
KAGRA MIR (Absorption)
Print this report.
ManuelMarchio - 16:11, Monday 19 June 2017 (511)Get code to link to this report
Comment to HEPA filter check for replacement (Click here to view original report: 426)

The filter's fan part was fine, so it was enough to buy only the filter part. New HEPA filters were delivered to Tama last friday.  I washed the prefilters with water, I cleaned the fans, replaced the filters and placed them back to the top of the absorption bench clean booth.

Images attached to this comment
511_20170619085204_p6190460.jpg 511_20170619090509_p6190022.jpg
R&D (FilterCavity)
Print this report.
RaffaeleFlaminio - 10:50, Monday 19 June 2017 (509)Get code to link to this report
Vacuum in EM2, Gate valve opening and pressure
On Monday June the 12th after lunch I started the evacuation in the EM2 chamber using the large rotative pump.
After 30'-45' the pressure was at 100 mTor so I started the turbo pump.
On Friday evening (June 17th) the pressure in the EM2 chamber reached 3.1e-7 Torr.
On the other side of the gate valve the pressure was 1.1 e-7 Torr.
So I open the large gate valve and immediately after the valve between the turbo pump and the tube
(the valve between the turbo pump and the EM2 chamber was already open).
The pressure went up a little bit in the 1e-6 Torr range and then came down again.
This morning (Monday June the 19th) the pressures are:

EM2 chamber = 1.6e-7 Torr
EM2 tube = 2.3e-7 Torr
Middle of the tube: 4.7e-8 Torr
Middle of the tube: 2.7e-8 Torr
NM2 tube = 8.6e-8 Torr (this sensor was not working but is working again now)
NM2 chamber = in air

Overall it looks reasonable to me even if there is probably some out-gassing coming
from the EM2 chamber.
KAGRA MIR (Absorption)
Print this report.
ManuelMarchio - 17:14, Thursday 15 June 2017 (508)Get code to link to this report
VI bug solved

Setting the limits at the translation stage generated a bug in the part of the VI where it sends the positioning commands: if the position exceeds the axis range, the VI goes in loop. I fixed it making a subVI "Read limits axis.vi" that reads the limits from the translation stage controller and set the minimum and maximum positions in the property node of the position controls of the VI.

Images attached to this report
508_20170615101300_propertynodes.png
Non-image files attached to this report
R&D (General)
Print this report.
YuefanGuo - 15:07, Thursday 15 June 2017 (507)Get code to link to this report
Telescope of the mode cleaner cavity
I measured the beam transmitted by the beam splitter of the MZ, did the fitting and found the beam waist position of that beam.

The origin was set at the front surface of the beam splitter. The beam waist size is w0=51.2um and waist position is z=-9.5cm.

Then I took this as the initial value to calculate the lens we are going to use for the telescope. So the condition is, the MC input mirror is 60cm away from the origin, and the beam waist should be around the position of the input mirror and has a size of 277um.

I let the program chose the lenses with focal length we have now and got this result.

L1: f1=100mm z1=5cm
L2: f2=200mm z2=37.5cm
L3: f3=200mm z3=55cm

So with this design the final beam waist is w1=279um, z1=60cm, which is the result we want.

The other thing is before we were using a 200mm lens to focalize the green beam we distracted from the Faraday to the PD. But yesterday we found out, if we want to give power supply and the signal output, we need more space for cable. But the position of the PD before was too close to the post of the steering mirror. So today I changed the lens into a 150mm one, and moved the PD 5cm forward.
Images attached to this report
507_20170615080642_mctele.png