LOG-IN
Displaying reports 2561-2580 of 3128.Go to page Start 125 126 127 128 129 130 131 132 133 End
KAGRA MIR (Absorption)
Print this report.
ManuelMarchio - 04:10, Thursday 21 December 2017 (615)Get code to link to this report
Pump beam waist at different powers

see the attached document

Non-image files attached to this report
R&D (FilterCavity)
Print this report.
YuhangZhao - 22:14, Wednesday 20 December 2017 (614)Get code to link to this report
Locking of filter cavity again
After the installation of shield case and step around, we reconnect the cable. This cause the beam goes away. Now we can lock it again like figure 1.

While I check the first iris, I find the pattern on it is really similar with the pattern we get from infrared. So I think the problem should come from the bench. Refer to figure 2.

But the locking is not stable enough, I will check the phase of two signals sending to EOM. Refer to figure 3.
Images attached to this report
614_20171220141420_lock1220.jpg 614_20171220141437_iris1.jpg 614_20171220141448_signal.jpg
R&D (FilterCavity)
Print this report.
AkihiroTomura - 18:47, Wednesday 20 December 2017 (613)Get code to link to this report
Mode cleaner telescope 1
I measured dimensions of the green beam after Mach-Zehnder to design a telescope for a mode cleaner. One of attached pictures shows a result. The x-axis of this graph starts from the output of EOM. 
 
An eigenmode of the mode cleaner was calculated using values and a formula shown in a picture. A cavity length was roughly estimated from the number of holes on the optical table because I could not find it in e-logbook.
 
Based on these information, I searched for a optimal lens pair for telescope. One example is shown in a picture. In this case, mode matching factor for vertical and horizontal axis of the beam are 99.37 % and 99.56 %, respectively. 
Images attached to this report
613_20171220102445_20171220gausbeamfit.jpg 613_20171220102834_20171220092640911ios.jpg 613_20171220104537_newsol1.jpg
Comments related to this report
MatteoLeonardi - 16:41, Thursday 21 December 2017 (616)
I upload the drawings of the MC (made by Uraguchi-san) as a reference for dimensions and specs.
AkihiroTomura - 01:07, Wednesday 10 January 2018 (623)
This is a correction of the entry 613.
In the entry 613, I made a miscalculation. This entry is a correct one.


I re-calculated optimal lens pairs for a telescope for the mode cleaner (green) with accurate dimensions (see here for the drawing). Figures attached show lens pairs that seem acceptable and the parameters I used.
In the case of the first figure, the mode matching factors are 99.718% and 99.601% for vertical and horizontal axes, respectively.
In the case of the second figure, the mode matching factors are 100% and 99.727% for vertical and horizontal axes, respectively.


The lens with 75.6 mm focal length that we already have is one without any surface coating thus low transmissivity. So I think the first case is preferable.
R&D (FilterCavity)
Print this report.
AkihiroTomura - 13:32, Wednesday 20 December 2017 (612)Get code to link to this report
Caution about AOM alignment

After the installation of the coverage on optical bench, we are now checking the situation.

Note that our AOM is very easily misaligned by touching the cable.

Temporarily, I put a post right next to the cable in order not to make it move a lot (see a figure attached).

Images attached to this report
612_20171220053042_20171219064718302ios1.jpg
R&D (FilterCavity)
Print this report.
MatteoLeonardi, YuhangZhao - 13:41, Tuesday 19 December 2017 (611)Get code to link to this report
Step around the optical bench
Yesterday the step was mounted by the company around the optical bench.
In order to allow their work all the cables from the bench were disconnected and the racks were moved away.
After the company work the cables are being reconnected (ongoing).
Images attached to this report
611_20171219054144_img20171218wa0006.jpg
R&D (FilterCavity)
Print this report.
YuhangZhao,AkihiroTomura - 11:55, Friday 15 December 2017 (610)Get code to link to this report
irises realignment and checking infrared beam
During yesterday and the day before, we mainly did two things.

1. For leaving space for the shield case, we moved two irises close to PR. Due to one of them is used to fix the lens rail, so we need to adjust the lens again to get good mode matching. Adjusting that lens costed a period of time, and then we can lock the cavity as it before.

2. Then we improvised a tripod to hold the camera, which is used for the near iris in the vacuum tube. Then we adjusted its position. Note that the camera we used here if found on the optical bench in the central room.

3. While we can see the EM transmitted signal, we rise up both of two irises one by one. And we cannot see a clear infrared on the iris. At the same time, the transmitted signal was blocked partly.

By the way, I found the transmitted signal of EM is on the edge of EM chamber view port.
(You can see the infrared checking video by clicking the following link.)
https://drive.google.com/open?id=1CqVLnQrDSaECMgJzHP2Fivabp3ip_E8T
https://drive.google.com/open?id=18AbcXlJ9LMBlW7SOZ7-MdkWTKyISZRya
R&D (FilterCavity)
Print this report.
YuhangZhao - 14:04, Sunday 03 December 2017 (609)Get code to link to this report
About the second iris's camera, SHG and infrared beam
Today I tried to find infrared beam again. But I found two things.

Firstly, I checked the camera for the second iris. I brought two cameras and one cable around the second iris back. We have a PR's camera, so I used it to check. The method is to replace one thing at a time. In this way, we can diagnose the replaced component. I found the cable is fine but two cameras are broken.
Broken camera 1: SONY NO.92980
Broken camera 2:SONY NO.161274

Secondly, I found the drifting of SHG is really serious.(with the driven voltage set around 50V). Within only 30 min or less, the power decreased from 50mW to 30mW and final 10mW. The measurement point is ahead of filter cavity's EOM.

Finally, I took a picture of the beam which is sent to our filter cavity. You can see from the attached photo. The beam is a combination of a large bright beam point and a small weak point. I cannot make it a rough circle by just adjusting the turing mirror or local control offset.
Images attached to this report
609_20171203060432_infrared.jpg
R&D (FilterCavity)
Print this report.
YuhangZhao - 01:31, Saturday 02 December 2017 (608)Get code to link to this report
Finding the infrared beam
I'm trying to find the infrared beam during these two days. I was trying to find it in-between two half of the halos.

Firstly, about the losing of the camera of the second target(or irises or aperture). I checked the link and cable. I also changed the driver for this camera. All of them didn't bring back our camera. But I found there was a gap on the case of camera.(As you can see in the figure one.) Maybe this is one reason, so I will check its capability further more.(Since we can see the scattered light or halo, I think there is no need to use the second irises. But brining it back will be a good thing for the future.)

Secondly, about the cable in the end room. I almost stumble because of the cable around the end room's bench. The reason is one of the cables is really short, which makes it suspended away the ground. And this happened on Thursday. On Friday, I cannot find the transmitted signal from the end mirror. I checked many things and finally realized where the problem would be. If you read previous part carefully, you can see the problem comes from the short cable. It causes the misalignment of the QPD. So I think we should do something to avoid this happens again.

Thirdly, I can find something really similar with yuefan's previous infrared beam gazing. You can find the video I took from the link in the bracket. But I tried to change the local control offset for each mirror, finding it's really different from the behavior of green beam. For example, the changing of this possible infrared beam does't follow the changing of the local control offset very well. Sometimes it looks really like the halo after changing the local control offset. I need to confirm this strictly, but the signal here is really weak! I feel the confirmation is really difficult.
(https://drive.google.com/file/d/12OmtNB93hNnNigcgu_0SNxJNy5pxSIPq/view?usp=sharing)

Fourthly, Tomura-san found we can use the sensor card to sense the infrared beam in the end room. I used it to check the light from EM chamber output port to the camera. I found almost one third of the beam is lost because the alignment on the end room bench. So I changed the alignment a little bit. The change includes one lens, one 45 degree mirror(Actually I don't know what this mirror for) and the camera. Now we can see the whole two half of halos by changing the angle and height of camera. You can see the video by clicking the attached link as following.
(https://drive.google.com/file/d/1MsWtaURQJLrXH1PSE4-g_qudrmI73w9c/view?usp=sharing)

Fifthly, Tomura-san suggested there is possibility that something is blocking on the infrared. Because you can see a black line in the spot of reference infrared beam. But I check the layout of our experiment, it's not the beam sending to cavity. However, the reference which is sending to cavity has also something like defect. The defect is the combination of a small spot and a larger spot. It's really like the scenario we can see from the infrared transmitted signal of EM, where there is a bright halo and a darker halo.(You can see this combination from the second video)

Sixthly, I checked the bright halo by changing the local control offset. As you can see from the attached second and third photo, there is angle changing of halo. And the halo is flashing. I suspect this halo can be the what we want to find. I will try to figure it out soon.

In the end, I can ride the bicycle very well now! I need to go to end room and go back many times to find the signal, which drives me to ride bicycle many times. This is really an useful skill and a good way to calm down or feel balance. &( ^___^ )&
Images attached to this report
608_20171201162920_camera.jpg 608_20171201172829_halo1.jpg 608_20171201172837_halo2.jpg
KAGRA MIR (Absorption)
Print this report.
ManuelMarchio - 20:40, Friday 01 December 2017 (607)Get code to link to this report
New linear stage assembly

I screw 4 short pedestals below the M-UMR12.80 linear stage for the HeNe imaging unit, I clamped it with 4 forks on the breadboard to be parallel to the beam.
Since there was not enough space for the forks, I had to move the hole breadboard by a few cm. 
I did the same with the other linear stage UMR12.40 for the 1310nm probe. Since it is imperial, I had to look for a set of 4 imperial screws in ATC.
All the parts I used, I washed them in the ultrasonic bath in ATC. Except for the stages themselves because the bearings have grease, so better not to wash them.

Images attached to this report
607_20171201123949_10.jpg 607_20171201123956_101.jpg 607_20171201124002_11.jpg
KAGRA MIR (Absorption)
Print this report.
ManuelMarchio - 19:52, Thursday 30 November 2017 (606)Get code to link to this report
Pump laser alignment

after moving alle the optics of the IPC I had to realign the pump laser beam on the pinhole (at the reference position of the cross point).

First I adjusted the height of the beam on the optical table from the IPC to the periscope (128mm)
Then I removed the last focusing lens on the optical breadboard (after the periscope).
I used two positions of the pinhole along z (max and min position limits), and maximized the power transmitted by the pinhole. I moved back and forth the translation stage on those positions. On the closest position I adjusted the second stearing mirror, on the further position I adjusted the periscope mirror. For the first rough alignment I also used an optical cross.
Now the beam is parallel to the z axis of the translation stage. Then i put the last focusing lens back on its mount and I adjusted the lens position to have the maximum power transmitted by the pinhole.

Since I had optimized the IPC, now, in order to have 30mW on the imaging unit table with 0.8A of laser current, the half wave plate have to have an angle of about 22deg (before optimization it was 0deg)

Then I made a scan of the reference sample with the HeNe Imaging unit, and when I went to optimize the maximum of the AC scan (adjusting the pump focusing lens position) I found that it was already at the maximum.

Images attached to this report
606_20171130115130_20171130scanrefsurf.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 16:14, Thursday 30 November 2017 (605)Get code to link to this report
Align the infrared beam2
Today I tried to use the camera to search for the infrared beam. And I took a video.

In the video, we can see the whole halo. But if you check something in the halo, you will find there is only something like the fiber.

I attach the video as following.

https://drive.google.com/open?id=18PApeToBRHC-tDisKg--beNVbhRKi0-E
R&D (FilterCavity)
Print this report.
YuhangZhao,TomuraAkihiro - 21:36, Wednesday 29 November 2017 (603)Get code to link to this report
Align the infrared beam
Today I and Tomura-san tried to do something about aligning the infrared beam.

Firstly, we try to lock the cavity with green beam. The height peak is 1V. After locking the transmitted signal is also 1V. We didn't change anything else but the EOM for filter cavity.

(Because we need to put a case on the top of our bench. The base plate of EOM for filter cavity excesses the boundary. So we change the base plate angle and align it agin.)

Secondly, we follow the procedure of yuefan. We made sure the laser hit on the right position of the membrane of PR chamber. And we also change the height and angle of camera. Then we can see the signal as attached. We supposed that it's the halo as yuefan mention in the procedure. We tried to moved the camera around to find the laser point and we failed. Note here, we can see only a part of the halo. That means we can see only the shape as a sector. I think the reason maybe the incident angle. And I tried to change the incident angle, but find nothing more. I will try to figure out how to find the beam tomorrow.
Images attached to this report
603_20171129133524_finealignment.jpg 603_20171129133538_lockingacquisition.jpg 603_20171129133550_infraredcamera.jpg
KAGRA MIR (Absorption)
Print this report.
ManuelMarchio - 18:51, Wednesday 29 November 2017 (602)Get code to link to this report
IPC optimization

Keeping the high-power laser current at 0.8A we have 420mW right after the fiber output.
Today the best I could get from the IPC was a power maximum and minimum (as a function of the rotation angle of the halfwaveplate) of:

Right after the IPC
407mW (max)
15mW (min)

On the imaging unit breadboard (after all the optics on the path, including the chopper)
190mW max
3mW min (comparable with the power-meter sensitivity)

In order to get 30mW (the power we usually use for calibration), I increased the laser current up to 1.3A

In order to notice a waist difference (in the absorption signal) between low and high current, we have to be able to have 30mW with high current. I think we have to get the IPC better optimized, or, if not possible, we need to add another IPC 

Images attached to this report
602_20171129105009_img20171129171734.jpg
KAGRA MIR (Absorption)
Print this report.
ManuelMarchio - 13:49, Wednesday 29 November 2017 (600)Get code to link to this report
Alignment of red probe and IPC optimization

With Matteo, we worked on the red probe last week

before replacing the linear stage with the longer one, we checked what's the maximum signal we could get in this condition
we moved the linear stage to find the max of the AC/DC signal. As usual, everytime we moved the linear stage, we maximize the DC

- Pump laser current= 0.78, power= 30mW

- Reference sample surface

- DC max=3.15V

- AC/DC max =0.061

-------------------------------------------------------------------------

Pump laser polarization

With Kuroki, we put a cube PBS right after the output fiber and measured the p and s polarization powers with the high-power power-meter
laser current 0.8A
420mW total power
325mW s-polarization
6-10mW p-polarization (power meter not very sensitive at this low power)

-------------------------------------------------------------------------
Input Power Control

we put the half wave plate in front of the output fiber to rotate the polarization into p. Then we adjusted the angle of the first plate PBS to minimize the reflected beam.
Result: minimum of reflected (p-polarized) = 19mW  (transmitted400mW)
The strange thing is that the angle is about 64deg, much larger than 56deg

Images attached to this report
600_20171129053703_img20171128191945.jpg
R&D (FilterCavity)
Print this report.
AkihiroTomura - 13:32, Wednesday 29 November 2017 (601)Get code to link to this report
Comment to Beam Radius Measurement (Click here to view original report: 597)

Vertical: w0 = 2.67e-5 m, z0 = 4.73e-2 m

Horizontal: w0 = 2.77e-5 m, z0 = 4.62e-2 m

where w0 is the spot size at a beam waist and z0 = a position of the beam waist.

Images attached to this comment
601_20171129053039_20171124gaussianbeamfitting.jpg
KAGRA MIR (Absorption)
Print this report.
ManuelMarchio - 09:49, Tuesday 28 November 2017 (599)Get code to link to this report
Pump laser Power vs Current

I put a power meter right in front of the output fiber of the pump laser and characterized the laser power as a function of the current.

Non-image files attached to this report
R&D (FilterCavity)
Print this report.
YuhangZhao - 00:39, Tuesday 28 November 2017 (598)Get code to link to this report
Something in the temperature and humidity chamber
Please check the attached file.
Non-image files attached to this report
R&D (FilterCavity)
Print this report.
ZhaoYu-hang, TomuraAkihiro, - 19:02, Friday 24 November 2017 (597)Get code to link to this report
Beam Radius Measurement
This week we measured the beam radius after EOM. The measurement started from 100mm after EOM. We took 6 points and used Gaussian beam formula to fit. The result is shown as below.

The beam waist is 2.61667926e-05(m). We set the position of EOM as origin, the position of beam waist is 4.72853283e-02(m).

For details, please refer to the figure attached below.
Images attached to this report
597_20171124110225_beamfitting.png
Comments related to this report
AkihiroTomura - 13:32, Wednesday 29 November 2017 (601)

Vertical: w0 = 2.67e-5 m, z0 = 4.73e-2 m

Horizontal: w0 = 2.77e-5 m, z0 = 4.62e-2 m

where w0 is the spot size at a beam waist and z0 = a position of the beam waist.

R&D (FilterCavity)
Print this report.
AkihiroTomura,YuhangZhao,MatteoLeonardi - 11:46, Wednesday 22 November 2017 (596)Get code to link to this report
Filter cavity locking after FC pump shutdown

Three days after FC pump shutdown on last Saturday, we checked the filter cavity locking system. Soon the whole system recovered properly and the cavity was locked. Note that the transmission power while being locked was improved after we changed the demodulation phase. Now it is around 8.0V while it was 4.0V before.

 

Preparing for installation of coverage on the optical bench, we are changing positions of some optical components. We changed the main laser position slightly. Fortunately, it did not cause problems. One thing we want to notice is that we now have only two clamps fixing laser onto the bench because of lack of space. We do not recommend you to touch laser body.

R&D (FilterCavity)
Print this report.
MatteoLeonardi, ManuelMarchio - 10:21, Monday 20 November 2017 (595)Get code to link to this report
FC pump system recovery
After the planned electrical shutdown of 2017/11/18, the FC pump system has been switched on with success yesterday evening. Note that, due to the shutdown, all the general switches close to each pump station were switched off.
After two days without pumping the pressure in the tube close to the IM was approx. 5mbar, while close to EM it was approx 0.5mbar.