LOG-IN
Displaying reports 1941-1960 of 3128.Go to page Start 94 95 96 97 98 99 100 101 102 End
R&D (FilterCavity)
Print this report.
Federico Paoletti - 20:09, Saturday 15 June 2019 (1412)Get code to link to this report
Comment to HUGE 50 Hz on END signal after connecting picomotors driver to the power line (Click here to view original report: 1392)

You could use some remote "smart" PDUs as done in virgo.


See as example

https://tds.virgo-gw.eu/?content=3&r=15139

(the name of the PDU is ENERGENIE EG-PMS2-LAN, see attached PDF file)

Non-image files attached to this comment
R&D (FilterCavity)
Print this report.
YuhangZhao - 01:29, Friday 14 June 2019 (1411)Get code to link to this report
Estimation of phase noise in our nowadays situation for Squeezer

Based on the measurement we did before, we have dark noise of CC PD. We use this PD to lock the green pump phase, at the same time, we have bandwidth lower than 80Hz. So we could use this dark noise to evaluate the phase noise we have for coherent control loop 1. The calibration method is to use pk-pk value when we scan the green phase, which corresponds to radians of pi/2. In the case of that entry, 25mV corresponds to pi/2. The estimated phase noise(RMS) is attached as follows.

While it will be better to measure the spectrum and do RMS integration. And according to Emil thesis, this 200mrad of phase will degrade ~15dB of squeezing to ~5dB. And it seems to be close to the squeezing situation we saw in the past few days.

Images attached to this report
1411_20190613182857_figure1.png
Comments related to this report
NaokiAritomi - 18:26, Tuesday 25 June 2019 (1422)

According to Emil's thesis P. 42 figure 2.10 or P.48 figure 2.13 (b), 200 mrad of phase noise seems to degrade 15 dB of squeezing to almost 0 dB of squeezing. Our situation seems around 100 mrad of phase noise. Did you consider the effect of control bandwidth when you calculated rms phase noise? As you noted, it's better to measure the spectrum and integrate it within control bandwidth.   

R&D (FilterCavity)
Print this report.
NaokiAritomi - 20:26, Thursday 13 June 2019 (1410)Get code to link to this report
GRMC alignment and lock with new green phase shifter

[Aritomi, Eleonora P, Yuhang]

Today we replaced a green phase shifter with a new thick phase shifter (picture 1). After re-alignment of GRMC, we got good mode matching (picture 2) and succeeded in locking of GRMC with lower gain. Green power is 146mW before GRMC and 90mW after GRMC when MZ is maximized. So transmission of GRMC is 62% which is good.

Lock of green phase with BAB transmission seems stable with this new phase shifter.

We'll measure transfer function of green phase lock loop with new green phase shifter tomorrow.

Today we replaced a green phase shifter with a new thick phase shifter.
(We will take the picture later)
After re-alignment of GRMC, we got good mode matching (picture 1) and succeeded in lock of GRMC with lower gain. Green power is 146mW before GRMC and 90mW after GRMC when MZ is maximized. So transmission of GRMC is 62% which is good.
Today we replaced a green phase shifter with a new thick phase shifter.
(We will take the picture later)
After re-alignment of GRMC, we got good mode matching (picture 1) and succeeded in lock of GRMC with lower gain. Green power is 146mW before GRMC and 90mW after GRMC when MZ is maximized. So transmission of GRMC is 62% which is good.
Images attached to this report
1410_20190614125254_img7540.jpg 1410_20190614125301_img7537.jpg
R&D (FilterCavity)
Print this report.
YuhangZhao - 11:13, Thursday 13 June 2019 (1409)Get code to link to this report
The first characterization of misalignment and mode mismatching for filter cavity(IR/BAB)

Eleonora P, Yuhang, and Aritomi

By keeping green always locked, we measured the power of all the higher order modes we could see from camera. The reason to keep green locked is that we have 'two different FSR' for IR. We found a lot of higher order modes. But the main higher order mode is HG10. This means our alignment needs to be improved for yaw direction. The power for each mode is reported as follows.(picture attached insequence without HG10 and TEM00)

AOM frequency Transmission power Mode
108.93233MHz  180mV HG10
109.02030MHz 62mV HG50
109.02068MHz 64mV  
109.02125MHz 54mV  
109.03569MHz 90mV TEM00
109.20801MHz 54mV HG80
109.20911MHz 54mV HG81
109.22574MHz 76mV HG30
109.22632MHz 56mV LG11
109.41733MHz 56mV  
DC offset 48mV  

Some modes I didn't write name seem so-called ince-gaussian modes. Maybe it comes from the combination of yaw misalignment and mode-mismatch.

We found the second HG10 at 109.4313MHz, which means FSR is 0.499MHz which is in agreement with our expectation.

We will do alignment of yaw for the next step.

Images attached to this report
1409_20190613070942_2.jpg 1409_20190613070950_3.jpg 1409_20190613071000_4.jpg 1409_20190613071009_6.jpg 1409_20190613071016_7.jpg 1409_20190613071022_8.jpg 1409_20190613071028_9.jpg 1409_20190613071034_10.jpg
KAGRA AOS (General)
Print this report.
SimonZeidler - 17:45, Wednesday 12 June 2019 (1408)Get code to link to this report
Relocalisation of Scatterometer

Akutsu-san, Tanioka-san, Simon

It has been a long time since I wrote something in this logbook! smiley

Anyway, we (that means AOS) are now relocating the scatterometer, which is still in one of JASMINE's laboratories at the ATC, to our lab on the first floor of ATC.
The reason is mainly that the JASMINE group needs the space in their lab and removed some desks which we used for the scatterometer's PC.

On the other hand, all of our optics-related stuff is more or less already in that lab where we are going to put the scatterometer in (including the back-scatterometer). So, it seems logical to relocate it there.

The first step in this week was therefore to find a suitable place in the lab, and we decided to use Torii-san's former space for that.
So, we cleaned it up and moved a theoretically usable clean-booth (still without walls) into the free space and a black-painted optical table underneath that booth (see pictures).

By doing all the cleaning-up thing, we discarded a lot of old carton-boxes and plastic garbage.
Now, it looks much more usable.

Next step will be to move the actual scatterometer.

Images attached to this report
1408_20190612104442_img1372.jpg 1408_20190612104458_img1373.jpg 1408_20190612104510_img1371.jpg
General (General)
Print this report.
YuhangZhao - 11:18, Wednesday 12 June 2019 (1407)Get code to link to this report
Water leakage in the tunnel of TAMA

The day before yesterday, we found a water leakage point in the middle station of TAMA arm (north, the arm we are using). See entry #1403. Yesterday I found it was filled fully by the leaked water.

Images attached to this report
1407_20190612041817_wechatimg5.jpeg
R&D (FilterCavity)
Print this report.
YuhangZhao - 23:38, Tuesday 11 June 2019 (1406)Get code to link to this report
Lock both green and BAB(after OPO transmission) together

Eleonora P, Yuhang, Matteo, and Aritomi

Several months ago, we have already found the channel for AOM amplification was broken. And the work we did yesterday was without AOM working. Fortunately, Eleonora C remembers where is the old RF amplifier. And today we first implemented the old RF amplifier(ZHL-2). The amplification factor of it is 16dBm. The optimal RF signal for AOM should be 27dBm(while it converts too much to the first order). so usually, we use a lower value. So I give 7dBm from the signal generator and amplify it by 16dBm to have totally 23dBm of driving signal to AOM. The conversion efficiency is about 80% now.

Then, for sure, we need to do the alignment for green again. And then also for IR.

Another very important thing is the amplification of BAB(after OPO transmission). In the beginning, we were thinking to lock it with the same method for CC locking. But we observed noise level brought by the beat between BAB and CC. And this noise is almost 200 times larger than the coherent control error signal we have. So it is not possible to lock with CC beam. By following Matteo's suggestion, we used the leakage power from OPO s-pol transmission through PBS and to p-pol locking PD. Then we feedback this signal after giving an offset. And then we can basically lock the green phase and have a more stable IR beam going inside the filter cavity.

After all the work above, we could lock both green and IR again by changing the AOM driving frequency. We found make higher order modes and they are listed as follows.

TEM00 109.03325MHz
HG10 108.932MHz
LG01 109.123MHz

The task of tomorrow will be to maximize the TEM00 by moving the last two steering mirrors on the bench for IR. And measure the height of each higher order modes. Also, maximize mode matching. We will follow the entry.

Images attached to this report
1406_20190611163814_lg01.jpeg 1406_20190611163829_hg10.jpeg 1406_20190611163842_tem00.jpeg
R&D (FilterCavity)
Print this report.
YuhangZhao - 10:41, Tuesday 11 June 2019 (1405)Get code to link to this report
Alignment of the amplified BAB into filter cavity

EleonoraP, EleonoraC and Yuhang

We achieved the matching of the BAB (IR probe beam co-aligned with squeezing) into the filter cavity.

In order to align BAB into the FC, we mainly followed the IR alignment procedure we did more than one year ago (entry #646). The main steps of this procedure are to align the cavity for the green, recover the reference on PR chamber, maximize IR reflection from the input mirror, not to look for the beam on the first target and check instead the second target and the end camera.

However, since BAB is only 250uW, it is too weak to see. So we used the green pump to amplify it. The maximum green we can give is obtained setting the offset on the MZ control servo at 4.3 and at that time we have roughly 30mW of IR going to FC. This measurement of power is done while we scanned the green pump phase. So actually we are sending a repeated segment of a sine wave.

By using this 30mW and following the procedure above, we found the flashes of IR. Check attached video.

As you can see in the video, this beam is too bright so, in order to avoid saturation, we decide to reduce the power to around 10mW. In this case, we need to use temperature of 7.202 for OPO, MZ offset of 4.1 and lock p-pol PLL on 150MHz.

After achiving the alignment with the amplified BAB we confirmed that flashes cannot be seen without amplification. (BAB power ~250 uW)

Next steps are to optimize the matching and to use AOM to make green and IR both resonant.

Non-image files attached to this report
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 20:31, Monday 10 June 2019 (1404)Get code to link to this report
FC GREEN transmission into DAQ

I used one of the few ADC channels still availabe to acquire the filter cavity green transmission. Channel name is K1: FDS-FC_GR_TRA

For the moment it is recorded as an epic channel, since I wanted to display it on the medm screen.

The attached picture show today's lock streches (Even it doesn't seem so, no realigment has been done during this period and the lock was stopped on purpose)

Since the trasmission PD is actually a PSD, I plan to acquire also the X and Y signals in order to  possibly correlate beam motion to suspesion resonances and understand what we should improve.

Images attached to this report
1404_20190610131148_greentra.png
General (General)
Print this report.
EleonoraCapocasa - 19:16, Monday 10 June 2019 (1403)Get code to link to this report
Water leakage from the ceiling of the north-south arm in the mid-arm station

As entitled. 

It is probably due to the heavy rain of these days. You can check the video. We put a bucket to collect the water.

Non-image files attached to this report
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 18:35, Monday 10 June 2019 (1402)Get code to link to this report
USB PCI-Express also doesn't work on new DGS computer

As reported in entry #1267 we could not setup the new DGS computer wtith gentoo linux HDD, as it was not able to detect the USB.  We bought a PCI express board to be used as USB driver.

Last week I installed it but I still couldn't make the PC to detect USB. I think the board might not be compatible with linux. We should look for another one or consider to change PC.

Images attached to this report
1402_20190610121722_5d0d5683b2594d20bfccab0a220250cd.jpeg
R&D (FilterCavity)
Print this report.
EleonoraPolini - 17:42, Monday 10 June 2019 (1400)Get code to link to this report
Installation of IR injection telescope

[Eleonora P, Yuhang]

We installed the IR injection telescope on the bench and we pre-installed the reflection telescope (putting all the optical mounts on the bench without fixing them).

We used the two 750 mm focal length lenses present in the lens box in the lab. Since we saw the beam focusing before the cavity, after the telescope, we thought one of the two focal lengths to be wrong. In fact the second one was actually wrong and it was replaced by another lens of 750 mm focal, located in the lab (Newport KPX121AR.18, PCX LENS, BK 7).

We put HR Laser Line mirrors, BK 7, ALTECHNA Co. Ltd.

In fig 1 the photo of the injection telescope on the bench. We used two of the new small mirror mounts.

Images attached to this report
1400_20190610104134_injectiontelescopeinstallation.jpg
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 17:15, Monday 10 June 2019 (1399)Get code to link to this report
Effect of rotary pump vibrarion reduced on BS

We notice that the pipe of the rotary pump close to BS was directly touching the vacuum pipe, transmitting a lot of vibrations.  We have moved the pump of  few centimeters so that the pipe was not touching anymore.

The vibration of BS in the region of 20 Hz seems improved. (See pic 1)

Images attached to this report
1399_20190610101532_screenshotfrom20190607171827.png 1399_20190610103319_4f21c91cfafd4a168faa8e31c7c4a57d.jpg 1399_20190610103327_6e52e4be3bab4eb09dec5db9682b2e08.jpg
R&D (FilterCavity)
Print this report.
YuhangZhao - 21:23, Friday 07 June 2019 (1398)Get code to link to this report
mirror check for squeezing path

EleonoraP and Yuhang

Today we checked the PBS and the two following mirrors for OPO transmission. PBS has a transmission of 0.2% while we measured reflection more than incidence.  And the two mirrors are fine as well. We didn't use wrong coating components or they have some issues.

R&D (FilterCavity)
Print this report.
YuhangZhao - 10:58, Friday 07 June 2019 (1396)Get code to link to this report
The measurement of squeezing before and after cleaning optics

Chien-Ming and Yuhang

After the last very good measurement of squeezing, we start to think about improving coherent control 1 PD. Then we measured squeezing again. The result is attached in figure 1 and 2. As you can see the low frequency is covered by classical noise while high frequency seems fine. However, the squeezing level was only 4.9dB at that moment.

Then Chien-ming suggested cleaning optics. And then we checked optics together. We found optics are tremendously dirty. Especially one of the lenses is very dirty on one side of the surface. And for others, there are lots of dust on the surface of the mirror. However, after the cleaning of optics, the balance of BS is destroyed. After the alignment of this BS, we measured squeezing again. However, we didn't improve the squeezing. We observed squeezing level almost the same with last time. See attached figure 3 and 4.

There is some consideration after this measurement:

1. The fluctuation of measurement is about 4dB pk-pk. Is this coming from the measurement device?  

2. The low-frequency noise of the second measurement of squeezing is much higher than the first one. We also know that for the second measurement, the balance of the SQZ part is very bad. The unbalance signal on PD is about 50mV. Although we know this balance is not as important as the balance of local oscillator, we still need to make it stay at a decent level.

3. To keep a clean clean room is really important for squeezing to avoid some potential losses.

4. We also tried to change coherent control power, but we noticed that after we decrease coherent control power we may also increase phase noise. This measurement is better done when we have a better coherent loop. Also with some characterization of the coherent control loop.

Images attached to this report
1396_20190607130347_wholef.png 1396_20190607130352_highf.png 1396_20190607130405_wholef.png 1396_20190607130418_highf.png
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 23:22, Thursday 06 June 2019 (1394)Get code to link to this report
Suspensions spectra with vacuum pump off in the central area

[Yuhang, Eleonora]

Today we temporarily swiched off both the rotary and turbo pump in the central area (close to BS).

The attached plot show the comparison of the spectra of the suspensions in the central area, with pumps on and off.

We see that BS is strongly affected by the pumps vibration while PR and INPUT seem not.

[Note that at the time when I developed local control for the first time, the pump where swiched off].

The pumps have been switched off for about 3 hours ( from 5PM to 8PM).

Images attached to this report
1394_20190606125646_vacuumpumpoff.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 22:26, Thursday 06 June 2019 (1395)Get code to link to this report
Some number for filter cavity locking in June 2019

Yuhang, Aritomi, EleonoraP, and EleonoraC

On 5th of June 2019, we locked the filter cavity again after almost half a year we haven't done this. We checked several references including irises on the bench, target on the film of PR chamber, target on the film of BS chamber, the first and second target inside the filter cavity vacuum tube. To recover these checking points, we used pico-motors with almost no failures. But it seems there are still some problems with end mirror pico-motor. Also, we observed some large mirror motion after we moved pico-motor of BS. But it went back to static after several minutes.

The alignment is also done with the standard procedure. It is first to misalign input mirror and find filter cavity transmission. Then center it on the screen. Then bring back input mirror and make injection and reflection overlap. Finally, align end mirror to have TEM00 dominant. After the lock, change the control point of each mirror to see if the transmission goes up to maximize.

Power:

Before AOM 36.5mW
Before EOM 212mW
Before MZ 166mW
To filter cavity 34mW
Before FC GR ref PD 136uW

Signal:

FC tra PD 4V
FC GR ref PD DC 80mV
FC locking error signal 260mV

The demodulation phase for filter cavity locking is 65deg.

The setting of FC locking servo:

Input attenuation 3.6
Gain 1 4.0
Images attached to this report
1395_20190606152639_wechatimg428.jpeg 1395_20190606171052_fcpdtragreen.jpeg
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 19:36, Thursday 06 June 2019 (1393)Get code to link to this report
DAC recovery after death

Today suddenly DAC died. Since I was working around the DGS I supposed I accidentaly disconnect some cables, but I check carefully and everything seemed fine.

We rebooted the standalone computer and the problem was solved. Note that in order to switch it on we had to disconnect and reconnect the power cables and to disconnect the timing singnal (pic 1). Timing cable was reconnected after the restart.

Images attached to this report
1393_20190606123552_timing.jpeg
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 19:27, Thursday 06 June 2019 (1392)Get code to link to this report
HUGE 50 Hz on END signal after connecting picomotors driver to the power line

We observed that connecting end room picomotor driver to the power line in the same mupliple socket of OPLEV laser and PSD brings a HUGE 50 Hz in the signals.

Since picomotors are only used temporarily we have now disconneted them. We may try to connect them to another socket next time that we need them.

Comments related to this report
Federico Paoletti - 20:09, Saturday 15 June 2019 (1412)

You could use some remote "smart" PDUs as done in virgo.


See as example

https://tds.virgo-gw.eu/?content=3&r=15139

(the name of the PDU is ENERGENIE EG-PMS2-LAN, see attached PDF file)

R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 19:11, Thursday 06 June 2019 (1388)Get code to link to this report
Local control for END mirror with new DGS

Local control on END mirror were implemented.

Pic 1 Pitch TF

Pic 2 YawTF

Pic 3 Comparison between open and closed loop spectra.

Pic 4, 5 photon model and damp for pitch  

Pic 6, 7 photon model and damp for yaw

I found a minimum of the sensing coupling when input signals (pitch and yaw) are rotated of -0.06 rad. So I updated accordingly the rotation matrix.

I used the following driving matrix  (pitch, yaw -> coils)   [0, 1, 0, 1; 1 0 -1, 1].  Note that the coil disposition is not the usual one!

For a comparison with the old control check entry #238.

Images attached to this report
1388_20190606121047_endpitchtf310519.png 1388_20190606121052_endyawtf310519.png 1388_20190606121117_endspectra300519.png 1388_20190606121558_endpitchdamp.jpeg 1388_20190606121607_endpitchmodel.jpeg 1388_20190606121615_endyawdamp.jpeg 1388_20190606121622_endyawmodel.jpeg