R&D (FilterCavity)
EleonoraCapocasa - 17:15, Monday 10 June 2019 (1399)
Effect of rotary pump vibrarion reduced on BS

We notice that the pipe of the rotary pump close to BS was directly touching the vacuum pipe, transmitting a lot of vibrations.  We have moved the pump of  few centimeters so that the pipe was not touching anymore.

The vibration of BS in the region of 20 Hz seems improved. (See pic 1)

Images attached to this report
1399_20190610101532_screenshotfrom20190607171827.png 1399_20190610103319_4f21c91cfafd4a168faa8e31c7c4a57d.jpg 1399_20190610103327_6e52e4be3bab4eb09dec5db9682b2e08.jpg