LOG-IN
Displaying reports 1961-1980 of 3128.Go to page Start 95 96 97 98 99 100 101 102 103 End
R&D (FilterCavity)
Print this report.
NaokiAritomi - 18:55, Thursday 06 June 2019 (1391)Get code to link to this report
Power of BAB transmission from filter cavity
 
I roughly calculated power of BAB transmission from filter cavity. When the injection of BAB is 200uW, transmission power at PD is 1.2uW. If we use parametric amplification, we can increase it by a factor of 20. Here is the assumption.
 
BAB injection: 200uW
Injection loss: 10%
Transmission of filter cavity: 1%
Transmission of IR line filter in End bench: 70%
BAB transmission at PD: 1.2uW  
R&D (Cryogenic)
Print this report.
SatoshiTanioka - 18:04, Thursday 06 June 2019 (1390)Get code to link to this report
Scattered light and vibration of chamber

Today I did alignment work to see the flash of transmitted beam using PD.
I though that PD detected transmitted beam, but it turned out that it was scattered beam...

Then I re-aligned so that the reflected beam reach to FI, and scanned laser frequency with monitoring reflected beam.
At that time, the reflcted beam power was fluctuated.
Actually, it was caused by vibration of cryostat chamber.
So far I have used long pedestal to hold silicon mirrors, and this may affect the vibration of cavity.
Using fixed spacer may reduce the vibration effect, but it might be better to consider about vibration isolation system.

Images attached to this report
1390_20190606105114_img9334.jpg
R&D (Cryogenic)
Print this report.
SatoshiTanioka - 20:06, Wednesday 05 June 2019 (1389)Get code to link to this report
Install PD and alignment work

I installed a PD for monitoring transimtted beam of optical cavity with one STM. (forgot to take a photo...)
Though the transmitted power is very weak, the PD detects the transmitted beam.

Then I did alignment and scanned the laser frequency.
However, I have not seen any fringe so far.

I will continue to align the beam with scanning laser frequency and try to find good alignment.

R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 13:00, Wednesday 05 June 2019 (1387)Get code to link to this report
Suspensions behavior during EQ

Taking advatage of the the DAQ I playbacked the data and check the oplev signal during a 4.9 Earthquak occured in Chiba prefecture, last saturday.

https://earthquaketrack.com/quakes/2019-05-31-22-58-08-utc-4-9-20

The EQ is cleary seen by all the sensors. Some dof ( in paricular  PR and END pitch) didn't go back to the original position.

The data on DAQ are referred to UTC but the time of standalone PC is not well sincronized (about 3 min ahead the correct one).

Images attached to this report
1387_20190605060035_eq310519.png
R&D (FilterCavity)
Print this report.
ChienMingWu - 01:13, Wednesday 05 June 2019 (1386)Get code to link to this report
Improvement of 14MHz PD of coherent control

Chien-Ming, Yu-Hang,

We want to reduce the coherent noise coupled into the homodyne by reducing the intensity of CC beam entering the OPO. However, this action will cause the demodulation 14 MHz error signal (for coherent control of the pump beam phase) to become smaller. So, we tried to increase the gain of the photodetector at 14MHz by changing the capacitor CX and the amplifier resistors R1, R2, see Fig.1.

We let the capacitor CX become a plug-in base and test some combinations of capacitors. Currently we find the biggest signal at 14 MHz when using CX as 150pF+33pF. What is puzzling is that when we replace this combination with a single 180 pF capacitor, the signal is reduced by 8 dB on the spectrum analyzer. So, we are now using 150pF+33pF, and R1=500 ohms R2=5k ohms. The S/N now is improved 5 times more as shown in Fig.2.

Images attached to this report
1386_20190604181309_114mhzpd.png 1386_20190604181313_2snisimproved.png
R&D (Cryogenic)
Print this report.
SatoshiTanioka - 19:20, Monday 03 June 2019 (1384)Get code to link to this report
Silicon mirror installation and rough alignment

I installed another silicon mirror inside the cryostat.
Then I did rough alignment of input beam with 2 STMs and the reflected beam reached to FI.

Tomorrow, I will scan the laser frequency and try to find transmitted HOMs.

Images attached to this report
1384_20190603121629_setup.jpg 1384_20190603121633_silicon.jpg
R&D (FilterCavity)
Print this report.
YuhangZhao - 15:30, Monday 03 June 2019 (1383)Get code to link to this report
Some activity of coherent control PD

Yuhang and Chien-Ming

As we said before, we need a better PD for the coherent control1 lock. (PD set at OPO reflection) So we talked and tested several things.

1. Change the amplification factor of the op-amp. We tried two combinations of (500Om, 50kOm) and (500Om, 7.5kOm).  For the first combination, we tested the signal, but the amplitude and SNR are the same as before. While for the second combination, the amplitude is smaller while the SNR is a bit better.

We checked the spec of AD8057. The maximum gain we should give for this op-amp is 10. The further increase of the gain will degrade the bandwidth. And this is exactly what we see for increasing further the amplification factor.

It seems that we are also reaching the saturation current of the op-amp.

2. We also tried to use the RF amplifier, because maybe there is an optimal input level for the mixer to make it work better. But it seems RF amplifier increases both signal and noise.

3. We also tried to change the capacitor of the PD circuit. But we are still trying.

R&D (FilterCavity)
Print this report.
EleonoraPolini - 13:31, Monday 03 June 2019 (1382)Get code to link to this report
Green path FI installation

[Eleonora P, Yuhang]

We replaced the FI's stage along the green path, after a hole has been drilled on it as explained in entry #1367

In fig.1 the picture of the stage on the bench.

Images attached to this report
1382_20190603063032_20.jpeg
R&D (Cryogenic)
Print this report.
SatoshiTanioka - 17:56, Saturday 01 June 2019 (1381)Get code to link to this report
Silicon mirror installed and beam profile measurement

I installed one silicon mirror, which was labeled "No.1", inside the cryostat and did rough alignment.
Then I measured beam profile for mode matching.

R&D (Cryogenic)
Print this report.
SatoshiTanioka - 18:45, Friday 31 May 2019 (1380)Get code to link to this report
Installation work

Today, I re-installed some optics for higher-order-modes.
I wanted to put a iris for the mark of alignment, but I could not find it...
From tomorrow, I'm going to install silicon cavity inside the cryostat.

R&D (Cryogenic)
Print this report.
SatoshiTanioka - 19:35, Thursday 30 May 2019 (1379)Get code to link to this report
Optical layout changed

I changed optical layout to achieve better alignment.
I will update the schematic figure of optical layout later...
Anyway, the alignment seems ok so far.
 

R&D (FilterCavity)
Print this report.
EleonoraPolini - 17:24, Thursday 30 May 2019 (1378)Get code to link to this report
Robustness of injection telescope considering the error on beam waist position

As shown in entry #1363, the error on the beam waist position is large ( around 8%).

I checked if the solution #3 of entry #1366 would be good also considering the border positions of beam waist in the error interval, running again the simulations (fig 1, 2, 3) and analyzing the robustness on Python.

Both focal lengths are 750 mm.

I summarize the results in the following table:

z0 (m) pos lens 1 (m) pos lens 2 (m) pos lens 1 at min mismatch (m) pos lens 2 at min mismatch (m) min mismatch (%)
0.089 0.420 0.673 0.424 0.673 0.9
0.081 0.420 0.673 0.4276 0.683 1.8
0.097 0.420 0.673 0.410 0.6755 1.6

where the positions of the lenses is referred to the PBS after the OPO.

I computed on Python the position of the two lenses corresponding to the condition of minimum total mismatch.

In fig. 4, 5 and 6 are shown the robustness of the telescope with z=0.081 (min position), 0.089 (real position), 0.097 (max position), that are still pretty good (less than 10% mismatch).

Images attached to this report
1378_20190530101050_081.png 1378_20190530101100_096.png 1378_20190530101109_097.png 1378_20190531093840_injmin081.png 1378_20190531093851_injmin089.png 1378_20190531093902_injmin097.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 00:12, Thursday 30 May 2019 (1377)Get code to link to this report
Need to have better PD for CC1 locking if we want to reduce CC laser power and phase noise

[Yuhang, Chien-Ming]

Today we tried to decrease CC laser power. But when we just decreased it to half, we couldn't lock the CC1 loop. 

The cc error signals that we can use to lock and cannot be used to lock are attached in Fig1.

The RMS dark noise for CC1 and CC2 PD is measured and attached in Fig2. We can see that the CC1 PD has RMS dark noise almost 5 times higher than CC2 PD. 

We also suspected that this noisy PD for CC1 locking brings also phase noise.

Images attached to this report
1377_20190529171218_cc1.png 1377_20190529171227_cc2.png
R&D (Cryogenic)
Print this report.
SatoshiTanioka - 23:13, Wednesday 29 May 2019 (1376)Get code to link to this report
Operation check of FI for TEM00 and some modifications

I installed a FI for TEM00 PDH locking, and confirmed that one can pick off reflected beam with simple setup (QWP and mirror).
The attached picture shows the reflected beam picked up by FI.

Then, I continued the installation of optics.
However, I found that the height of FI was not good, which is not the one I mentioned but already installed one before.
So I modified the alignment.
Actually, the beam height is still higher than the target height, though I cannot adjust anymore.

So I am planning to install STMs in front of the EOM, which can adjust the beam height.

Images attached to this report
1376_20190529160637_fi20190529.jpg
R&D (FilterCavity)
Print this report.
YuhangZhao - 22:44, Wednesday 29 May 2019 (1375)Get code to link to this report
Measurement of 5.5dB SQZ down to 50Hz after homodyne improvement

[Yuhang, Chien-Ming, EleonoraP]

After the improvement of homodyne suggested by Chien-Ming, we measured squeezing again.

This time, we could see only two peaks for homodyne shot noise limited down to 10Hz when LO incident. The 34Hz peak is related to some unknown device(maybe air-conditioner). The characterization of this 34Hz peak is done by Federico and Irene. While another peak is a 50Hz power line. (We didn't perform a low-frequency high-resolution measurement, we may see some peaks at low frequency in that case. We will do it soon.)

For SQZ, we could see the noise floor(5.5dB SQZ) down to 10Hz with several narrow peaks. But notice that now there is still 2.4mV DC signal(for SQZ path) not balanced on homodyne. We could possibly improve CMRR for SQZ path soon.

For anti-SQZ, we could see 13.25dB. And it is also very flat.

The result is shown in the attached figure. We didn't pay attention to the measurement bandwidth of anti-SQZ and LO-shot noise. But we may do it again after the refinement of BAB balance.

Images attached to this report
1375_20190529154545_shotnoise.png 1375_20190529154552_5dbsqz.png
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 16:26, Wednesday 29 May 2019 (1372)Get code to link to this report
INPUT MIRROR cotrolled with new DGS

The local controls for the input mirror have been implemented in the new DGS.

As usual,  we amplified and filtered the PSD signals for pitch and yaw with a SR560 (2nd order lowpass, cutoff 100 Hz, gain 100)

TF for yaw and pitch are reported in pic 1 and 2. (injected noise: 5000 and 7000 counts respectively)

As for PR and BS the pitch TF and coherence is not very good.

Comparison between open and closed loop spectra are shown in pic 3 (yaw) and 4 (pitch)

I attach the filters values (model and damp for both yaw and pitch):

Fig 5  yaw model

Fig 6  yaw damp

Fig 7 pitch model

Fig 8 pitch damp

I will compute the calibration soon.

Images attached to this report
1372_20190529092004_inputyawtf209519.png 1372_20190529092010_inputpitchtf290519.png 1372_20190529092115_inputyawspe290519.png 1372_20190529092120_inputpitchspe290519.png 1372_20190529092139_inputyawmodel.png 1372_20190529092147_inputyawdamp.png 1372_20190529092152_inputpitchmodel.png 1372_20190529092355_inputpitchdamp.png
R&D (FilterCavity)
Print this report.
ChienMingWu - 22:27, Tuesday 28 May 2019 (1370)Get code to link to this report
Homodyne DC balance

Chien-Ming, Yu-Hang, and Aritomi

Yesterday we adjusted the BS horizontal angle to make the LO beam reach the DC balance at homodyne detector. We measured the CMRR, it is better than 80 dB at 1 kHz.

However, in this case, if we let that BAB and LO have a good overlapping, we need to change the optical set up of BAB a lot.

So today we keep the BAB beam in its original condition and adjust the BS angle to make the BAB beam reach the DC balance at homodyne. Then we adjust the LO beam to achieve a good overlapping with BAB beam. In this case, when the BAB is DC balanced, the homodyne DC offset of the LO shows -10mV.

We repeatedly adjust the BS angle and the alignment of both beams a little bit, finally, we make the LO beam become DC balanced and the DC Offset of BAB is 2.4 mV.

R&D (FilterCavity)
Print this report.
NaokiAritomi - 19:27, Monday 27 May 2019 (1369)Get code to link to this report
Alignment method of homodyne
 
[Aritomi, Chien-Ming, Yuhang]
 
Our BS for homodyne (BSW41-1064) is R = 49.14513016% (p-pol) and T= 50.49834442% (p-pol) when the injection angle is 45 deg. That means we have 1.35% unbalance when the injection angle is 45 deg. So far we made the injection angle 45 deg and actually we had 74 mV homodyne DC offset even if the homodyne is aligned. Each PD has 5V so the unbalance is 1.48%. To reduce this unbalance, the injection angle should be below 45 deg.
 
The alignment method of homodyne will be as follows. The rough sketch is also attached.
 
1. inject BAB to homodyne PD along the hole line of optical bench
2. adjust the angle of BS so that homodyne DC of BAB is 0
3. align BAB to AMC with two  mirrors after homodyne BS
4. align LO to AMC with two mirrors before homodyne BS
 
Our BS for homodyne (BSW41-1064) is R = 49.14513016% (ppol) and T= 50.49834442% (ppol).
That means we have 1.35% unbalance when the injection angle is 45 deg. So far we made the injection angle 45 deg and actually we had 70 mV homodyne DC offset even if homodyne is aligned. Each PD has 5V so the unbalance is 1.4%. We have to make the injection angle below 45 deg to get 1:1 beam splitting ratio.
Our BS for homodyne (BSW41-1064) is R = 49.14513016% (ppol) and T= 50.49834442% (ppol).
That means we have 1.35% unbalance when the injection angle is 45 deg. So far we made the injection angle 45 deg and actually we had 70 mV homodyne DC offset even if homodyne is aligned. Each PD has 5V so the unbalance is 1.4%. We have to make the injection angle below 45 deg to get 1:1 beam splitting ratio.
Images attached to this report
1369_20190527122659_img7475.jpg
R&D (Cryogenic)
Print this report.
SatoshiTanioka - 19:21, Monday 27 May 2019 (1368)Get code to link to this report
Cryostat 4K shield

Takahashi-san, Sato-san, Tanioka

We removed the 4K shield of cryostat in ATC in order to ship it to KEK.
Some insulator are detached from 4K shield to remove screws.

Images attached to this report
1368_20190527122128_shield4k.jpg
R&D (FilterCavity)
Print this report.
EleonoraPolini - 17:51, Monday 27 May 2019 (1367)Get code to link to this report
Green path pre-alignment and FI stage pre-installation

[Yuhang and EleonoraP]

First we aligned the green beam before PR chamber.

Then we changed the stage of Faraday Isolator with a smaller one, in order to make space for the wind shields. We put some schotch to cover the borders of the FI, to avoid damages (fig 1).

We found a problem: not enough space for the mirror under the FI (fig 2). The black plastic edge of the mirror mount is 3.4mm while the beam reflected by the FI crystal is 5 mm far from the stage edge (fig 3).

Possible solutions:

1. make a hole in the mount in order to make space for the mirror. Size of the hole: 30 mm (diameter) and 12 mm (depth) in fig 4.

2. build a different stage with other components, in fig 5.

Images attached to this report
1367_20190527104206_img20190527wa0003.jpg 1367_20190527104251_img20190527wa0001.jpg 1367_20190527104922_fcgreenrefclipping.png 1367_20190527105015_36pm.jpeg 1367_20190527105025_img20190527172411.jpg