Warning: Undefined variable $s in /wwwsto01/wwwusers/gw-elog/osl/classes/DAO.php on line 959
NAOJ GW Elog Logbook
LOG-IN
Displaying report 1-1 of 1.
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 19:27, Thursday 06 June 2019 (1392)Get code to link to this report
HUGE 50 Hz on END signal after connecting picomotors driver to the power line

We observed that connecting end room picomotor driver to the power line in the same mupliple socket of OPLEV laser and PSD brings a HUGE 50 Hz in the signals.

Since picomotors are only used temporarily we have now disconneted them. We may try to connect them to another socket next time that we need them.

Comments related to this report
Federico Paoletti - 20:09, Saturday 15 June 2019 (1412)

You could use some remote "smart" PDUs as done in virgo.


See as example

https://tds.virgo-gw.eu/?content=3&r=15139

(the name of the PDU is ENERGENIE EG-PMS2-LAN, see attached PDF file)