LOG-IN
Displaying reports 1261-1280 of 3044.Go to page Start 60 61 62 63 64 65 66 67 68 End
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 18:59, Thursday 11 June 2020 (2093)Get code to link to this report
the first card in AA 0 module is broken

Yuhang, Eleonora

On Tue evening 9/06 we found out that some of the ADC channel were not working fine and one of the two power supplies was delivering the maximum current (3 A) and the saturation led was on.

I switched it off and on but after ~30 min it stated to saturated again. A strong smell of overheated plastic was present around the ADC/DAC modules. I switched it off for the night.

Yesterday after confirming that DACs were working fine we check the ADC channels one by one. Here what we found:

 

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
no sign  no sign  no sign  strange

strange

strange OK OK  OK  OK

strange

OK OK OK OK OK

 

We suspected the first card of the AI module (channel 1-8). We extracted the module form the rack and we confirmed that the first card had a very strong smell of burnt but we couldn’t individuate the broken component (we coudn't remove the card from the module to check the bottom part). We disconnected the card from the alimentation and reconnect again the AI module to the power supply. No saturation occurred since then. We will consider how to replace it. 

Note that ADC channel 11 even if is not connect to the broken card shows a strange feature (see pic 3). This channel is now used to acquire OPLEV INPUT SUM and it is not used in any control loop but the problem should be investigated.

Images attached to this report
2093_20200611115855_ai1.jpeg 2093_20200611115902_ai2.jpeg 2093_20200611115914_ai3.jpeg
R&D (FilterCavity)
Print this report.
YuhangZhao - 13:18, Wednesday 10 June 2020 (2091)Get code to link to this report
Measurement of FDS with additional Faraday isolator on 20200602

Yuhang, Eleonora

Last week we re-measured FDS with the new Faraday isolator.

The cavity was very well aligned (both in IR and green), mirror motion was small, but we could not go below shot noise at low frequency.

The squeezing at high frequency is 2.5dB (1dB less than usual). We suspect the Faraday is introducing more losses than we expected(elog 2055).

The back scattering seems quite low but we suspect that this is due to the low motion of suspended mirrors and not to the FI. In fact, in a previous measurement we observed high back scattering also with the new Faraday isolator(elog 2069).

Images attached to this report
2091_20200610061804_figure1.png
R&D (Cryogenic)
Print this report.
SatoshiTanioka - 19:01, Wednesday 03 June 2020 (2090)Get code to link to this report
Lock trial

Today I tried to lock the laser to the folded cavity with SR560 and succeeded to do that though it lasted about 10 sec.

In addition, I estimated the finesse from the linewidth of transmitted beam.
First, I measured the cavity length with a ruler and it was about 82 mm.
Then I scanned the laser frequency and measured the linewidth of the transmitted beam.

The free spectral range can be estimated as
f_FSR = 3e8/(2*83e-3)=1.8*109 Hz.
And the linewidth was about 1.4*105 Hz.
Therefore, the finesse can be calculated as F = 1.3*104.

Images attached to this report
2090_20200603120043_20200603cavity.jpeg 2090_20200603120051_20200603lock.jpg
R&D (Cryogenic)
Print this report.
SatoshiTanioka - 19:48, Monday 01 June 2020 (2089)Get code to link to this report
Alignment with folded geometry

Today I tried to align a cavity with folded geometry as shown in the picture.
The cavity has asymmetric cavity length where the longer one is about 60 mm and the shorter is 15 mm.

Eventually I could see the transmitted flash as attached picture.
Therefore, it may be possible to lock the cavity with a folding mirror though I have not checked the finesse nor transmitted power.

Tomorrow I will estimate the finesse with folded geometry and try to lock the laser using the servo.

Images attached to this report
2089_20200601124828_20200601cavity.jpg 2089_20200601124834_20200601.jpg
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 17:28, Monday 01 June 2020 (2088)Get code to link to this report
GR phase shifter successfully replaced

Takahashi-san, Yuhang, Eleonora

With the help of Takahashi-san we assembled the new GR phase shifter by gluing PZT and mirror on the new mount.

The GR mirror spec are shown in pic 1.

We used a glue (norland optical adhesive 61 in pic 2) that cured by 10 min exporsure to UV lamp (that lamp usually stored in 20 m lab, see manual in pic 3)

We soldered the new pzt connector(pic 5, green wire of PZT is positive) and replace the old phase shifter.

New alignment method for GRMC: since the new phase shifter doesn't have knobs to be adjusted, the alignment of GRMC now relies on the lens and one steering mirror between MZ and GRMC. Just as a memo, Yuhang moved lens position horizontally by hand. Besides, the alignment strongly depends on the voltage applied on pzt, so remember to turn on high voltage driver for green phase shifter before the lock of GRMC.

The CC1 TF (before and after replacment) is shown in pic 8. The phase noise measurement is shown in pic 6. The signal used to calibrate phase noise is shown in pic 7.

After the phase-shifter replacemnet the CC1 loop was much more stable and never got unlocked. There is 10dB attenuation for the error signal injected to the servo and the gain had to be reduced to the minimum.

Images attached to this report
2088_20200601102501_grmirr.jpeg 2088_20200601102509_glue.jpeg 2088_20200601102516_uvlamp.jpeg 2088_20200601102522_phaseshift.jpeg 2088_20200601104440_wechatimg666.jpeg 2088_20200601105118_newcc1phasenoise.png 2088_20200601105323_wechatimg667.jpeg 2088_20200601111525_comparison.png
R&D (Cryogenic)
Print this report.
SatoshiTanioka - 16:23, Sunday 31 May 2020 (2087)Get code to link to this report
Confirmation with FP cavity

This entry is log on May 28.

As I could not align the cavity with folded geometry and see the transmitted light, I decided to play with FP geometry to confirm whether the fused silica mirrors are O.K. or not.
The cavity length was set to 6.4 cm.
Actually, I could see the transmitted beam easily.

After that, I estimated the finesse by scanning the laser frequency.
The measured value was about 1.46*104 where the designed value was 1.5*104.
Therefore, the fused silica mirrors seemed to be O.K.

Next step is to lock the laser to this FP cavity and then I will try with folded cavity with shorter cavity length.

R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 20:22, Wednesday 27 May 2020 (2086)Get code to link to this report
frequency independent squeezing recovery

Yuhang, Eleonora

We check the power at different points on the bench and compared with nominal values. We only had to realign IRMC. After re-alignment its trasmission is above 80% (much larger than GRMC)

Here the current values for reference:

GR power before green EOM 280mW
GR power before AOM 51 mW
GR power before MZ 220mW
GR power after AOM 16 mW
GR before GRMC 48 mW
GR after GRMC 93 mW
GR into the viewport 14mW
GR pump before entering OPO 19 mW
IR power before IRMC 2.2mW
IRMC transmission 1.8mW

Then we adjust the setup to measure FIS. We recovered the alignment of homodyne and measured the squeezing, finding values similar as before (~ 6 dB). Plots will be updated soon. We remarked that the CC1 loop is not stable and keeps unlocking. We plan to repace soon the GR phase shifter, hoping this will impove the situation.

We also checked the alignement of BAB into FC. It was still fairly good.

R&D (FilterCavity)
Print this report.
YuhangZhao - 22:16, Tuesday 26 May 2020 (2085)Get code to link to this report
System recovery (filter cavity locked again)

Eleonora and Yuhang

Today we revovered filter cavity aligment and lock.

What we did:

1. Erase data on the disk which was already full

2. Recover reference on BS chamber by moving last green steering mirror on bench.

3. Moving PR pitch picomotor around 300 steps (with velocity of 500) up.

4. The threshold of FC lock remote lock was wrong (maybe touched accidently by someone), we followed procedure described in elog1886 and recovered it. We found out the threshold was changed from -0.5V to around -1V. We recovered it back to -0.5V, in this case, the remote lock worked again.

5. The green transmission was around 2000 but we didn't check injected green power.

6. The air-conditioner was changed to cold mode (temperature was set to be 22 deg).

R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 12:31, Sunday 26 April 2020 (2084)Get code to link to this report
Data disk full again: some data erased

The data disk was full and diaggui and similar were not working properly.

We cheked the status and deleted some old data with the procedure reported in entry #1979.

See attached the current situation of stored data.

Images attached to this report
2084_20200426053059_35.png
KAGRA MIR (Polarization)
Print this report.
PengboLi - 22:16, Sunday 22 March 2020 (2083)Get code to link to this report
Birefringence measurement results on Shinkosha S5 substrate

Simon, Pengbo

we finished the birefringence characterization of the shinkosha s5, as can be seen from the attachment(the first four figures are the result this time , the last two figures  are the results last year.), it seems the offset is even larger than before.

Images attached to this report
2083_20200322140750_figure1.png 2083_20200322140759_figure3.png 2083_20200322140808_figure2.png 2083_20200322140818_figure4.png 2083_20200322140854_figures1.png 2083_20200322140906_figures2.png
KAGRA MIR (Polarization)
Print this report.
PengboLi - 21:12, Thursday 19 March 2020 (2082)Get code to link to this report
Birefringence measurement results on Shinkosha S1 substrate

Simon, Pengbo

Today we finished the charaterization of the birefringence properties on Shinkosha S1 substrate. For the s-pol, we got 87.77 +/- 0.25 deg, which is slightly better than last year's result, 87.1+/- 0.05 deg. For the p-pol, we got 1.93+/- 0.68 deg, because we didn't do this measurment last year, there are no comparision. From the maps, it seems this time the distribution is more homogenous.

Images attached to this report
2082_20200319122548_figure1.png 2082_20200319122556_figure2.png 2082_20200319122606_figure3.png 2082_20200319122614_figure4.png 2082_20200319122629_figure21.png 2082_20200319122636_figure22.png
KAGRA MIR (Absorption)
Print this report.
SimonZeidler - 00:10, Wednesday 18 March 2020 (2081)Get code to link to this report
Comment to Absorption measurements on S1 (Click here to view original report: 2078)

Here the absorption map of the second surface.

Images attached to this comment
2081_20200317161024_map20200317xysurf2.png
R&D (Cryogenic)
Print this report.
SatoshiTanioka - 21:48, Tuesday 17 March 2020 (2079)Get code to link to this report
Comment to Viewport replacement (Click here to view original report: 2075)

The cryostat was pumped down againg and the pressure became less than 10-3 Pa.
This level is enough for the measurement at room temperature.
So I vented the chamber and opened in order to do the alignment of the cavity.

R&D (Cryogenic)
Print this report.
SatoshiTanioka - 19:40, Tuesday 17 March 2020 (2077)Get code to link to this report
Positioning of input optics

Background

Since I modified the mode matching lenses, I had to also modify the beam path.
Furthermore, in order to avoid putting the mode matching lenses between the steering mirrors, the layout should be reconsidered.

What I Did

I determied positions of input optics around the viewport.
They are a littele bit complicated as shown in the attached picture due to the limited space.
Then I aligned the mirrors in TEM00 path.

Next Step

  • Alignment of the TEM00 path.
  • Alignment of the input and output fused silica mirrors.
  • Need to purchase some mirror holders and pedestals for the HOM's paths (next FY).
Images attached to this report
2077_20200317105101_20200317.jpg
KAGRA MIR (Absorption)
Print this report.
SimonZeidler - 18:32, Tuesday 17 March 2020 (2078)Get code to link to this report
Absorption measurements on S1

Pengbo, Simon

Yesterday and today we repeated basically the measurements we did on S5 for S1. The results can be seen in the attachement.

Also for this one, it became obvious that there is a large absorption excess on the surface, This time, however, it seems that this excess is existent on both surfaces. Therefore, I had run the surface-absorption measurements on both (as the time I write this, the second one did not yet finish).
The center-map shows a distribution which is somewhat comparable to the measurements we did last year. However, the mean absorption coefficient has been increased by a factor of ~2.

The next step will be to do the birefringence measurements on both samples.

Images attached to this report
2078_20200317102534_map20200316xy.png 2078_20200317102550_dist20200316xy.png 2078_20200317102555_map20200317yz.png 2078_20200317103130_map20200317xysurf1.png
Comments related to this report
SimonZeidler - 00:10, Wednesday 18 March 2020 (2081)

Here the absorption map of the second surface.

R&D (Cryogenic)
Print this report.
SatoshiTanioka - 22:59, Monday 16 March 2020 (2075)Get code to link to this report
Viewport replacement

Preface

As the viewport attached the cryostat was not AR coated at the wavelength of 1550 nm but 1064 nm, there was a 10% loss for beam power at 1550 nm.
So I decided to replace the viewport for 1550 nm AR coated.
The cost, however, was a little bit expensive.
Therefore, I bought a AR coated window and a viewport without a window separately and they were supposed to be assembled.

What I did

I assembled the viewport as attached pic. 1.
Then I connected to the conversion flange though I was not confident to use M8 bolts and washers stored in the ATC.
After that I installed to the chamber and vented as a test.
Before venting, I measured the beam power before the viewport and after.

Result

First, the transmitted beam power was 2.73 mW against 2.74 mW input.
So the total loss was about 0.5% which is dramatically reduced compared to the previous one.
Moreover, the viewport does not have any serious leakages and the pressure reached less than 10-2 Pa.

Images attached to this report
2075_20200316123903_viewport1.jpg 2075_20200316123907_viewport2.jpeg
Comments related to this report
SatoshiTanioka - 21:48, Tuesday 17 March 2020 (2079)

The cryostat was pumped down againg and the pressure became less than 10-3 Pa.
This level is enough for the measurement at room temperature.
So I vented the chamber and opened in order to do the alignment of the cavity.

R&D (Cryogenic)
Print this report.
SatoshiTanioka - 22:58, Monday 16 March 2020 (2076)Get code to link to this report
Work on input optics

What I Did

  • Tweaking the lens and mirror position in a double-pass AOM path
  • Mode matching for TEM00 mode

Details

As I replaced a lens in double-pass AOM path, I tweaked the position of the lens and the mirror.
Current lens is f=70mm one.

Since the transmitted flash could not see due to the poor mode matching, I tried to improve the mode matching.
The lenses were replaced and f=-75mm and f=300mm ones are put now.
Their positions are about 0.226m and 0.659m from the beam waist, respectively.
After that I could get beam size about 65um at its waist where it is reasonable position for the apex mirror of folded cavity.

Next Step

  • Install some mirrors to inject the TEM00 beam into the chamber.
  • Alignment of the input and output mirrors.
  • Tweaking.
KAGRA MIR (Absorption)
Print this report.
SimonZeidler - 15:42, Monday 16 March 2020 (2074)Get code to link to this report
Absorption results on annealed S5

Pengbo, Simon

On the weekend, Pengbo finished the measurements on the annealed S5 sample.
Already last week, we recognized by doing the Z-scans that a large absorption excess must exist on one side of the sample (actually the side which has NO damages). So we decided to take actually 3 maps: in the center, on the suspicious surface and along the Z-axis. The results can be seen in the attachement.

As can be clearly seen, there is a very prominent structure on the suspicious surface which leads to an excess in the absoption data. This structure is partly visible also in the center-map, probably due to interference effects coming from the absorption excess (at least judging from the phase map). Please note that the sample for the center-map has been flipped so that the suspicious surface is on the out-going side because we had large problems in obtaining a meanigful result when the beam got influenced by the excess before reaching the targeted position.

As can be also seen from the map along the Z-axis, this excess is apparently not limited to the single surface area but has a depth which is hard to quantify (given the strong disturbance in phase and AC) but we estimate the affected depth to be several millimeter.

As a reference, the histogram for the absorption coefficient taken at the center is also given. The actual mean-value is not so far away from the measurements last year (however biased by the mentioned structure), which is at least one good news.

Images attached to this report
2074_20200316073857_20200314.png 2074_20200316073902_20200315.png 2074_20200316073906_20200316.png 2074_20200316073912_map20200314pengbo.png 2074_20200316073917_map20200315pengbo.png 2074_20200316073921_map20200316pengbo.png 2074_20200316073934_dist20200314pengbo.png
KAGRA MIR (Absorption)
Print this report.
SimonZeidler - 09:56, Thursday 12 March 2020 (2073)Get code to link to this report
Measuring absorption on annealed S1 and S5

Pengbo, Simon

We received the samples S1 and S5 back from France after their annealing.
Before we started measuring the absorption coefficient, we inspected the samples and we discovered some damages on the edges of both samples. Interestingly, the damages are only on one side (see attached photos).

The measuremetns started with S5 and are ongoing. However, before setting the sample into the sample holder, we cleaned it with FC on both sides.

Images attached to this report
2073_20200312015508_43.jpg 2073_20200312015515_24.jpg 2073_20200312015524_35.jpg 2073_20200312015528_15.jpg 2073_20200312015535_16.jpg
R&D (Cryogenic)
Print this report.
SatoshiTanioka - 23:26, Wednesday 11 March 2020 (2072)Get code to link to this report
Mode Matching

I tried to see the transmitted flash but I could not on Monday.
Indeed, the beam size was at its waist was much larger than the design value i.e., 100 um though 50 um is required.

I decided to change the mode matching lenses in order to achieve the designed spatial mode.
To be honest, the obtained minimun size was about 80 um radius.

I gonna try improving the mode matching until end of this week.

Note

The lenses are biconvex or biconcave in order to reduce the aberration of reflected beam.