LOG-IN
Displaying reports 1341-1360 of 3128.Go to page Start 64 65 66 67 68 69 70 71 72 End
R&D (FilterCavity)
Print this report.
RyutaroTakahashi - 17:14, Wednesday 17 June 2020 (2098)Get code to link to this report
Replacement of TMP in NM2

Takahashi, Yuhang, Eleonora

We replaced the TMP near NM2 from TMC1000MC (broken) to STP-1003 (new).

  1. Removed the screws for the ICF253 flange. Dismounted the old TMP.
  2. Mounted the new TMP. Set the rotational direction of exhaust to 225deg. Fastened the screws for the ICF253 flange. Final torque was 250kgf/cm.
  3. Connected the dry pump (DSP250). Started the dry pump. Opened the angle valve.
  4. Turned on the controller. It showed a fail due to different combination between the TMP and the controller. Reset the controller, then the controller went to auto tuning.
  5. Started the TMP. The TMP went to normal operation.
  6. After monitoring for two hours, opened gate valves to the 300-m pipe. The pressure became 4x10-7mbar.
Images attached to this report
2098_20200617101343_newtmpnm21.jpg 2098_20200617101351_newtmpnm22.jpg
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 14:42, Wednesday 17 June 2020 (2097)Get code to link to this report
PR oplev PSD replacement

Yuhang, Eleonora

During the "ADC trouble" we found out that PR oplev PSD was not working fine: 

- SUM channel was giving a negative output

- Y channel doesn't respond to vertical movement of the beam position on the sensor

We replaced the PSD with a spare one.

[Note that this spare was previously used for BS oplev (see elog #1947) and we changed it as we suspestected it was noisy but after the replacement we kept oberving occasional noise floor excess in BS so we confirmed that it was not PSD fault.]

After the replacement I mesaured the TF of PR (pic1) and found that it perfeclty overlaps to the previouse one, so I didn't need to modify the gain of the damping loops.

We suspect that PSD falilure can be connected to the AA burnt card. Investingations are on-going.

Images attached to this report
2097_20200617073913_prtfyaw.jpeg
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 23:05, Monday 15 June 2020 (2096)Get code to link to this report
Burnt opamps found in the broken AA card

Yuhang, Eleonora

We removed the broken card from the AI module (see elog #2093) and we found that at least 3 opamp AD8622 were burnt. (see pic 1,2)

The evidently broken opamps are: (see pic 3)

CH2 -> Oplev PR X

CH3 -> Oplev PR SUM

CH6 -> Oplev BS SUM

CH3 and CH6 are the most damaged. They corresponds to the SUM of oplev PSD for PR and BS. We remark that this two PSD uses the same power supply. 

We also remark that PR PSD is not working fine anymore (in particular channel Y and SUM).

We asked Miyakawa-san who said they have never had a similar problem in KAGRA. 

Images attached to this report
2096_20200615160445_16.jpeg 2096_20200615160457_26.jpeg 2096_20200616092716_aicard.jpeg
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 19:13, Friday 12 June 2020 (2094)Get code to link to this report
ADC works again

Yuhang, Eleonora

I changed the cable connection in order to move the channels that were connected to the broken AA card (AA 0 1-4 and 5-8) to a spare one in another AA module ( AA1 17-20 and 21-24). See Pic1.

I modified real time model accordingly. See pic2.

We tested the ADC input 1-8 and they are working fine now.  

Power supply is also ok since we disconnected the broken card inside the module, two days ago. Pic 3.

While doing ADC test we realize PSD of PR Oplev has some trouble. We will investiange it later.

Images attached to this report
2094_20200612121250_dgschan.jpeg 2094_20200612121302_rtm.jpeg 2094_20200612121605_powersupply.jpeg
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 18:59, Thursday 11 June 2020 (2093)Get code to link to this report
the first card in AA 0 module is broken

Yuhang, Eleonora

On Tue evening 9/06 we found out that some of the ADC channel were not working fine and one of the two power supplies was delivering the maximum current (3 A) and the saturation led was on.

I switched it off and on but after ~30 min it stated to saturated again. A strong smell of overheated plastic was present around the ADC/DAC modules. I switched it off for the night.

Yesterday after confirming that DACs were working fine we check the ADC channels one by one. Here what we found:

 

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
no sign  no sign  no sign  strange

strange

strange OK OK  OK  OK

strange

OK OK OK OK OK

 

We suspected the first card of the AI module (channel 1-8). We extracted the module form the rack and we confirmed that the first card had a very strong smell of burnt but we couldn’t individuate the broken component (we coudn't remove the card from the module to check the bottom part). We disconnected the card from the alimentation and reconnect again the AI module to the power supply. No saturation occurred since then. We will consider how to replace it. 

Note that ADC channel 11 even if is not connect to the broken card shows a strange feature (see pic 3). This channel is now used to acquire OPLEV INPUT SUM and it is not used in any control loop but the problem should be investigated.

Images attached to this report
2093_20200611115855_ai1.jpeg 2093_20200611115902_ai2.jpeg 2093_20200611115914_ai3.jpeg
R&D (FilterCavity)
Print this report.
YuhangZhao - 13:18, Wednesday 10 June 2020 (2091)Get code to link to this report
Measurement of FDS with additional Faraday isolator on 20200602

Yuhang, Eleonora

Last week we re-measured FDS with the new Faraday isolator.

The cavity was very well aligned (both in IR and green), mirror motion was small, but we could not go below shot noise at low frequency.

The squeezing at high frequency is 2.5dB (1dB less than usual). We suspect the Faraday is introducing more losses than we expected(elog 2055).

The back scattering seems quite low but we suspect that this is due to the low motion of suspended mirrors and not to the FI. In fact, in a previous measurement we observed high back scattering also with the new Faraday isolator(elog 2069).

Images attached to this report
2091_20200610061804_figure1.png
R&D (Cryogenic)
Print this report.
SatoshiTanioka - 19:01, Wednesday 03 June 2020 (2090)Get code to link to this report
Lock trial

Today I tried to lock the laser to the folded cavity with SR560 and succeeded to do that though it lasted about 10 sec.

In addition, I estimated the finesse from the linewidth of transmitted beam.
First, I measured the cavity length with a ruler and it was about 82 mm.
Then I scanned the laser frequency and measured the linewidth of the transmitted beam.

The free spectral range can be estimated as
f_FSR = 3e8/(2*83e-3)=1.8*109 Hz.
And the linewidth was about 1.4*105 Hz.
Therefore, the finesse can be calculated as F = 1.3*104.

Images attached to this report
2090_20200603120043_20200603cavity.jpeg 2090_20200603120051_20200603lock.jpg
R&D (Cryogenic)
Print this report.
SatoshiTanioka - 19:48, Monday 01 June 2020 (2089)Get code to link to this report
Alignment with folded geometry

Today I tried to align a cavity with folded geometry as shown in the picture.
The cavity has asymmetric cavity length where the longer one is about 60 mm and the shorter is 15 mm.

Eventually I could see the transmitted flash as attached picture.
Therefore, it may be possible to lock the cavity with a folding mirror though I have not checked the finesse nor transmitted power.

Tomorrow I will estimate the finesse with folded geometry and try to lock the laser using the servo.

Images attached to this report
2089_20200601124828_20200601cavity.jpg 2089_20200601124834_20200601.jpg
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 17:28, Monday 01 June 2020 (2088)Get code to link to this report
GR phase shifter successfully replaced

Takahashi-san, Yuhang, Eleonora

With the help of Takahashi-san we assembled the new GR phase shifter by gluing PZT and mirror on the new mount.

The GR mirror spec are shown in pic 1.

We used a glue (norland optical adhesive 61 in pic 2) that cured by 10 min exporsure to UV lamp (that lamp usually stored in 20 m lab, see manual in pic 3)

We soldered the new pzt connector(pic 5, green wire of PZT is positive) and replace the old phase shifter.

New alignment method for GRMC: since the new phase shifter doesn't have knobs to be adjusted, the alignment of GRMC now relies on the lens and one steering mirror between MZ and GRMC. Just as a memo, Yuhang moved lens position horizontally by hand. Besides, the alignment strongly depends on the voltage applied on pzt, so remember to turn on high voltage driver for green phase shifter before the lock of GRMC.

The CC1 TF (before and after replacment) is shown in pic 8. The phase noise measurement is shown in pic 6. The signal used to calibrate phase noise is shown in pic 7.

After the phase-shifter replacemnet the CC1 loop was much more stable and never got unlocked. There is 10dB attenuation for the error signal injected to the servo and the gain had to be reduced to the minimum.

Images attached to this report
2088_20200601102501_grmirr.jpeg 2088_20200601102509_glue.jpeg 2088_20200601102516_uvlamp.jpeg 2088_20200601102522_phaseshift.jpeg 2088_20200601104440_wechatimg666.jpeg 2088_20200601105118_newcc1phasenoise.png 2088_20200601105323_wechatimg667.jpeg 2088_20200601111525_comparison.png
R&D (Cryogenic)
Print this report.
SatoshiTanioka - 16:23, Sunday 31 May 2020 (2087)Get code to link to this report
Confirmation with FP cavity

This entry is log on May 28.

As I could not align the cavity with folded geometry and see the transmitted light, I decided to play with FP geometry to confirm whether the fused silica mirrors are O.K. or not.
The cavity length was set to 6.4 cm.
Actually, I could see the transmitted beam easily.

After that, I estimated the finesse by scanning the laser frequency.
The measured value was about 1.46*104 where the designed value was 1.5*104.
Therefore, the fused silica mirrors seemed to be O.K.

Next step is to lock the laser to this FP cavity and then I will try with folded cavity with shorter cavity length.

R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 20:22, Wednesday 27 May 2020 (2086)Get code to link to this report
frequency independent squeezing recovery

Yuhang, Eleonora

We check the power at different points on the bench and compared with nominal values. We only had to realign IRMC. After re-alignment its trasmission is above 80% (much larger than GRMC)

Here the current values for reference:

GR power before green EOM 280mW
GR power before AOM 51 mW
GR power before MZ 220mW
GR power after AOM 16 mW
GR before GRMC 48 mW
GR after GRMC 93 mW
GR into the viewport 14mW
GR pump before entering OPO 19 mW
IR power before IRMC 2.2mW
IRMC transmission 1.8mW

Then we adjust the setup to measure FIS. We recovered the alignment of homodyne and measured the squeezing, finding values similar as before (~ 6 dB). Plots will be updated soon. We remarked that the CC1 loop is not stable and keeps unlocking. We plan to repace soon the GR phase shifter, hoping this will impove the situation.

We also checked the alignement of BAB into FC. It was still fairly good.

R&D (FilterCavity)
Print this report.
YuhangZhao - 22:16, Tuesday 26 May 2020 (2085)Get code to link to this report
System recovery (filter cavity locked again)

Eleonora and Yuhang

Today we revovered filter cavity aligment and lock.

What we did:

1. Erase data on the disk which was already full

2. Recover reference on BS chamber by moving last green steering mirror on bench.

3. Moving PR pitch picomotor around 300 steps (with velocity of 500) up.

4. The threshold of FC lock remote lock was wrong (maybe touched accidently by someone), we followed procedure described in elog1886 and recovered it. We found out the threshold was changed from -0.5V to around -1V. We recovered it back to -0.5V, in this case, the remote lock worked again.

5. The green transmission was around 2000 but we didn't check injected green power.

6. The air-conditioner was changed to cold mode (temperature was set to be 22 deg).

R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 12:31, Sunday 26 April 2020 (2084)Get code to link to this report
Data disk full again: some data erased

The data disk was full and diaggui and similar were not working properly.

We cheked the status and deleted some old data with the procedure reported in entry #1979.

See attached the current situation of stored data.

Images attached to this report
2084_20200426053059_35.png
KAGRA MIR (Polarization)
Print this report.
PengboLi - 22:16, Sunday 22 March 2020 (2083)Get code to link to this report
Birefringence measurement results on Shinkosha S5 substrate

Simon, Pengbo

we finished the birefringence characterization of the shinkosha s5, as can be seen from the attachment(the first four figures are the result this time , the last two figures  are the results last year.), it seems the offset is even larger than before.

Images attached to this report
2083_20200322140750_figure1.png 2083_20200322140759_figure3.png 2083_20200322140808_figure2.png 2083_20200322140818_figure4.png 2083_20200322140854_figures1.png 2083_20200322140906_figures2.png
KAGRA MIR (Polarization)
Print this report.
PengboLi - 21:12, Thursday 19 March 2020 (2082)Get code to link to this report
Birefringence measurement results on Shinkosha S1 substrate

Simon, Pengbo

Today we finished the charaterization of the birefringence properties on Shinkosha S1 substrate. For the s-pol, we got 87.77 +/- 0.25 deg, which is slightly better than last year's result, 87.1+/- 0.05 deg. For the p-pol, we got 1.93+/- 0.68 deg, because we didn't do this measurment last year, there are no comparision. From the maps, it seems this time the distribution is more homogenous.

Images attached to this report
2082_20200319122548_figure1.png 2082_20200319122556_figure2.png 2082_20200319122606_figure3.png 2082_20200319122614_figure4.png 2082_20200319122629_figure21.png 2082_20200319122636_figure22.png
KAGRA MIR (Absorption)
Print this report.
SimonZeidler - 00:10, Wednesday 18 March 2020 (2081)Get code to link to this report
Comment to Absorption measurements on S1 (Click here to view original report: 2078)

Here the absorption map of the second surface.

Images attached to this comment
2081_20200317161024_map20200317xysurf2.png
R&D (Cryogenic)
Print this report.
SatoshiTanioka - 21:48, Tuesday 17 March 2020 (2079)Get code to link to this report
Comment to Viewport replacement (Click here to view original report: 2075)

The cryostat was pumped down againg and the pressure became less than 10-3 Pa.
This level is enough for the measurement at room temperature.
So I vented the chamber and opened in order to do the alignment of the cavity.

R&D (Cryogenic)
Print this report.
SatoshiTanioka - 19:40, Tuesday 17 March 2020 (2077)Get code to link to this report
Positioning of input optics

Background

Since I modified the mode matching lenses, I had to also modify the beam path.
Furthermore, in order to avoid putting the mode matching lenses between the steering mirrors, the layout should be reconsidered.

What I Did

I determied positions of input optics around the viewport.
They are a littele bit complicated as shown in the attached picture due to the limited space.
Then I aligned the mirrors in TEM00 path.

Next Step

  • Alignment of the TEM00 path.
  • Alignment of the input and output fused silica mirrors.
  • Need to purchase some mirror holders and pedestals for the HOM's paths (next FY).
Images attached to this report
2077_20200317105101_20200317.jpg
KAGRA MIR (Absorption)
Print this report.
SimonZeidler - 18:32, Tuesday 17 March 2020 (2078)Get code to link to this report
Absorption measurements on S1

Pengbo, Simon

Yesterday and today we repeated basically the measurements we did on S5 for S1. The results can be seen in the attachement.

Also for this one, it became obvious that there is a large absorption excess on the surface, This time, however, it seems that this excess is existent on both surfaces. Therefore, I had run the surface-absorption measurements on both (as the time I write this, the second one did not yet finish).
The center-map shows a distribution which is somewhat comparable to the measurements we did last year. However, the mean absorption coefficient has been increased by a factor of ~2.

The next step will be to do the birefringence measurements on both samples.

Images attached to this report
2078_20200317102534_map20200316xy.png 2078_20200317102550_dist20200316xy.png 2078_20200317102555_map20200317yz.png 2078_20200317103130_map20200317xysurf1.png
Comments related to this report
SimonZeidler - 00:10, Wednesday 18 March 2020 (2081)

Here the absorption map of the second surface.

R&D (Cryogenic)
Print this report.
SatoshiTanioka - 22:59, Monday 16 March 2020 (2075)Get code to link to this report
Viewport replacement

Preface

As the viewport attached the cryostat was not AR coated at the wavelength of 1550 nm but 1064 nm, there was a 10% loss for beam power at 1550 nm.
So I decided to replace the viewport for 1550 nm AR coated.
The cost, however, was a little bit expensive.
Therefore, I bought a AR coated window and a viewport without a window separately and they were supposed to be assembled.

What I did

I assembled the viewport as attached pic. 1.
Then I connected to the conversion flange though I was not confident to use M8 bolts and washers stored in the ATC.
After that I installed to the chamber and vented as a test.
Before venting, I measured the beam power before the viewport and after.

Result

First, the transmitted beam power was 2.73 mW against 2.74 mW input.
So the total loss was about 0.5% which is dramatically reduced compared to the previous one.
Moreover, the viewport does not have any serious leakages and the pressure reached less than 10-2 Pa.

Images attached to this report
2075_20200316123903_viewport1.jpg 2075_20200316123907_viewport2.jpeg
Comments related to this report
SatoshiTanioka - 21:48, Tuesday 17 March 2020 (2079)

The cryostat was pumped down againg and the pressure became less than 10-3 Pa.
This level is enough for the measurement at room temperature.
So I vented the chamber and opened in order to do the alignment of the cavity.