LOG-IN
Displaying reports 541-560 of 3128.Go to page Start 24 25 26 27 28 29 30 31 32 End
R&D (FilterCavity)
Print this report.
TatsukiWashimi - 12:40, Friday 24 June 2022 (2996)Get code to link to this report
Comment to Hammering test for FC optics (Click here to view original report: 2991)
I summarized the results of the hammering test, with power lines removal.
Images attached to this comment
2996_20220624054007_plothammeringoptics.png 2996_20220624054012_plothammeringbenches.png 2996_20220624054021_plothammeringchamber.png
R&D (FilterCavity)
Print this report.
NaokiAritomi - 20:03, Thursday 23 June 2022 (2995)Get code to link to this report
Sign of END H3 magnet is flipped

[Takahashi, Aritomi, Yuhang, Michael]

Since the END picomotor didn't move, we opened END chamber and aligned FC. We aligned PR and BS to make the green at center of GV between input/BS and at the upper left of first target as reported in elog2935. After we scanned BS alignment, we found the beam at second target.

Then we checked the behavior of END picomotor with old picomotor driver by looking at the green reflection from END mirror at the second target. We connected pitch picomotor to motor A and yaw picomotor to motor B. There are two settings of joystick: SET X and SET Y. In SET X, whatever the selected motor, when we move the joystick in X direction, the beam moves in yaw and when we move the joystick in Y direction, the beam moves in pitch. In SET Y, we can select the motor and the selected picomotor can be moved by moving the joystick in Y direction. We don't understand the bahavior of SET X, but we should use SET Y of joystick. We also confirmed that all motors of old picomotor driver are working.

To obtain the END oplev signals, we brought back the Dsub cable from PEM to END oplev signals. We found that when the END oplev beam is blocked, END oplev X and Y show around 2100 and -800 counts, respectively. After we centered the oplev around these values, we checked pitch/yaw TF. The pitch TF was fine (Fig. 1), but yaw TF was strange (Fig. 2). We found that yaw TF becomes fine when we flipped the sign of EUL2COIL matrix for H3 (Fig. 3). When the sign of EUL2COIL matrix for H1 is flipped, the sign of yaw TF is also flipped. We also checked the response of H1, H3 magnets with health check script (H1: Fig. 4, H3: Fig. 5), which also shows the sign of H3 magnet is flipped. This flipping happened when we glued it in elog2900. We flipped all the sign of EUL2COIL matrix for H3 (before: Fig. 6, after: Fig. 7). 

Finally we closed the chamber. We will start evacuation tomorrow.

Important note: After we glue the fallen magnet, we should also check the sign of the magnet. If the sign of the magnet is flipped, we need to flip the sign of EUL2COIL matrix for the magnet.

Images attached to this report
2995_20220623130428_20220623endpittf.png 2995_20220623130433_20220623endyawtfbefore.png 2995_20220623130438_20220623endyawtfh3flipped.png 2995_20220623130443_endh1.png 2995_20220623130447_endh3.png 2995_20220623130452_20220623endeul2coilbefore.png 2995_20220623130456_20220623endeul2coilafter.png
KAGRA MIR (General)
Print this report.
MarcEisenmann - 18:01, Thursday 23 June 2022 (2994)Get code to link to this report
Retest of Soleil-babinet compensator

Today I repeated the characterization of an 'unknown sample' (a HWP) using motorized HWP and translation motor.

First, the calibration gave X_cal = 13.3 mm which is compatible with the previous estimation.

Then, I measured the unknown sample retardance to be 0.496 which is expected value.

Next step will be to be able to access photodiode signal on the PC to start the LabView automatization.

R&D (FilterCavity)
Print this report.
TatsukiWashimi - 17:30, Thursday 23 June 2022 (2993)Get code to link to this report
Comment to Seismic and Acoustic measurement at TAMA FC (Click here to view original report: 2989)
I compared the seismic data and the microphone data, with KAGRA at the same time.
Seismic data of KAGRA is under OMC chamber and the microphone is in the IMC refl.

The large peak at one-handed several-tens Hz of the seismic data is the self-noise. (data sheet)
Images attached to this comment
2993_20220623103249_seiskagratama20220622.png 2993_20220623103254_mickagratama20220622.png
KAGRA MIR (Scattering)
Print this report.
HomareAbe - 10:43, Thursday 23 June 2022 (2992)Get code to link to this report
Scattering measurement of Aztec sample
Abe and Dan Chen,

We try to measure scattering of Aztec 3.

We can take a picture of scattering light.
In this figure, scattering light is in the yellow box.(Blue box is reflection light by mirror surface)
Images attached to this report
2992_20220623034325_20220622aztec3.jpg
R&D (FilterCavity)
Print this report.
TatsukiWashimi - 22:07, Wednesday 22 June 2022 (2991)Get code to link to this report
Hammering test for FC optics
[Marc, YokozaWashimi]
We performed the hammering test for the FC optics, to investigate the origin of the 170Hz noise.
50n Hz peaks are AC power and its harmonics.

The Eigen frequencies of the optics are higher.
The most suspicious object was the chamber.
Non-image files attached to this report
Comments related to this report
TatsukiWashimi - 12:40, Friday 24 June 2022 (2996)
I summarized the results of the hammering test, with power lines removal.
R&D (FilterCavity)
Print this report.
TatsukiWashimi - 13:09, Wednesday 22 June 2022 (2989)Get code to link to this report
Seismic and Acoustic measurement at TAMA FC
[YokozaWashimi]

We brought the PEM sensors from KAGRA to TAMA temporally.
* Trillium compact seismometer and its cable
* Whitening filter with DC power supply and D-sub 15pin cable
* ACO microphone, ACO 4ch power supply, and BNC cable
* 3-axis accelerometer and its cable


We connected the seismometer and the microphone to the ADC ports.
* K1:FDS-ADCspear_1_OUT_DQ -> Low frequency microphone
* K1:FDS-END_in_HOR_fil_IN1 -> Seismometer East-West
* K1:FDS-END_in_SUM_fil_IN1 -> Seismometer North-South
* K1:FDS-END_in_LEN_fil_IN1 -> Seismometer Vertical

The attached plots are seismic and acoustic data.
The units are "count", not calibrated to um/s more Pa.
Images attached to this report
2989_20220622060816_dsc1689.jpg 2989_20220622060821_dsc1690.jpg 2989_20220622060827_dsc1692.jpg 2989_20220622060852_screenshotfrom20220622110048.png
Comments related to this report
TatsukiWashimi - 17:30, Thursday 23 June 2022 (2993)
I compared the seismic data and the microphone data, with KAGRA at the same time.
Seismic data of KAGRA is under OMC chamber and the microphone is in the IMC refl.

The large peak at one-handed several-tens Hz of the seismic data is the self-noise. (data sheet)
TatsukiWashimi - 09:44, Monday 27 June 2022 (2998)
I monitored the seismic level at TAMA this weekend.


Below 1Hz, microseismic band, the ASD is large on Saturday night. It is consistent with the ocean wave.
At around 3Hz, the ASD is large on Saturday night and Monday morning, and small on Sunday.
Between 10 Hz and 20Hz, the ASD is large on Monday Morning.
Around 30 Hz, sometimes strange bumps are observed.
R&D (FilterCavity)
Print this report.
YuhangZhao - 12:23, Tuesday 21 June 2022 (2988)Get code to link to this report
Comment to Relative intensity noise measurement for the pump beam sent to OPO (Click here to view original report: 2980)

A comparison of RIN, shot noise and photo detector dark noise is shown in the attached first figure for several locations on the squeezer bench. Other figures show these noises for different single location separately.

Images attached to this comment
2988_20220621051232_rincomp.png 2988_20220621051839_shg.png 2988_20220621051847_ml.png 2988_20220621051858_irmc.png
R&D (FilterCavity)
Print this report.
NaokiAritomi - 18:16, Monday 20 June 2022 (2987)Get code to link to this report
BS suspension check

[Takahashi, Aritomi, Yuhang]

Since BS suspension motion was too large and the pitch picomotor can move in only one direction, we opened BS chamber and checked suspension. Takahashi-san found that the magnet was touching the coil holder so he adjusted it.

Regarding the picomotor, the picomotor cannot move in one direction because the required force to move the picomotor in the direction was too large. To reduce the required power to move the picomotor, Takahashi-san put a spring (Fig. 1). Then the picomotor can be moved in both directions.

We aligned PR and BS to make the green beam at center of the GV between input/BS and the first target.

Before closing the chamber, we checked BS oplev transfer functions (Fig. 2,3) and response of each coil. The transfer function and coil response are fine.

Finally we closed the chamber and started the evacuation of BS chamber.

Images attached to this report
2987_20220620111635_img9214.jpg 2987_20220620111644_bspittf20220620.png 2987_20220620111648_bsyawtf20220620.png
KAGRA MIR (Scattering)
Print this report.
MarcEisenmann - 17:29, Monday 20 June 2022 (2986)Get code to link to this report
preparation of the scattering setup

Dan Chen, Marc

Today we assembled on the 75 * 90 cm optical table the frame of the dark box inside which we will perform the scattering measurement.

We also placed AZTEC #1 and #3 on holders inside the frames.

KAGRA MIR (Absorption)
Print this report.
MarcEisenmann - 17:27, Monday 20 June 2022 (2985)Get code to link to this report
AZTEC #4 absorption

Report about AZTEC #4 absorption.

It seems to be around 100 ppm/cm and some fishbone patterns are visible as well.

Images attached to this report
2985_20220620102652_xy1.png 2985_20220620102656_xy2.png 2985_20220620102659_xy3.png 2985_20220620102703_xz.png 2985_20220620102708_yz.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 16:47, Monday 20 June 2022 (2984)Get code to link to this report
Comment to Mirror mount replacement along SHG path (Click here to view original report: 2977)

A proper beam dump thorlabs LB2 (1um - 12um) is installed for the newly replaced BS mount (as shown in Fig.1).

Although we can see some green light, let's remind that it is because the camera sensor is more sensitive to green as our eye. There are proper Dichroic mirrors to remove them.

Images attached to this comment
2984_20220620094735_2449420449222702925520220620164338.jpg
R&D (FilterCavity)
Print this report.
YuhangZhao - 16:41, Monday 20 June 2022 (2983)Get code to link to this report
Comment to Relative intensity noise measurement for the pump beam sent to OPO (Click here to view original report: 2980)

Yuhang, Aritomi, Marc

A new PD (PDA36A-EC) is used (gain is chosen as 0dB), firstly DARK NOISE is measured (saved as DARKL.txt and DARKH.txt)

A LA1608-YAG lens (75mm) was always used when the beam is apparently large.

1. Measure Mephisto RIN. It is measured at the location where we pick off main laser for CC PLL LO.

The DC voltage is measured as 2.66, 2.61V and 2.67V (measured before and after spectra measurements). (saved as MAINL.txt and MAINH.txt)

2. Measure homodyne LO RIN. It is measured just after IRMC transmission.

The DC voltage is measured as 800mV, 780mV (measured before and after spectra measurements). (saved as LOL.txt and LOH.txt)

3. Measure SHG generated GR RIN. It is measured before MZ.

The DC voltage is measured as 3.14V, 3.16V, 3.24V (measured before and after spectra measurements). (saved as SHGL.txt and SHGH.txt)

This is an experiment record. An analysis of data will come later.

R&D (FilterCavity)
Print this report.
NaokiAritomi - 18:13, Friday 17 June 2022 (2982)Get code to link to this report
So many issues on FC alignment

First we replaced the broken BS oplev PSD and measured BS oplev spectra as shown in the attached figure. The noise floor of BS oplev spectra is fine, but BS suspension moves a lot. In addition, BS pitch can go down with picomotor, but cannot go up. We need to open BS chamber again...

We somehow aligned FC and could see the green flash, but green beam spot at FC transmission moves too much due to BS suspension motion.

Then we found that we cannot maximize TEM00 very well and yaw misalignment cannot be removed. 

We aligned green injection in yaw before as reported in elog2940. We doubt that this could cause some problems. So we tweaked the green injection, but the green flash still had yaw misalignment. After we brought the green injection back to the original alignment, END yaw was misaglined a lot and we had to move the END yaw with picomotor. We went to END room and moved END picomotor with a joystick. However, at some point, END yaw picomotor stopped moving and we lost green flash...

To do list for END mirror:

  • center END oplev and check END oplev spectra/transfer function
  • install a new picomotor driver to avoid going to END room for picomotor control
  • If END oplev spectra are strange or picomotor does not move with the new picomotor driver, open END chamber and check the suspension and picomotor
Images attached to this report
2982_20220617111401_20220617.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 15:23, Friday 17 June 2022 (2980)Get code to link to this report
Relative intensity noise measurement for the pump beam sent to OPO

The OPO pump relative intensity noise was measured at different power level.

1. MZ offset 4.2 (25 mW), DC voltage on oscilloscope 320mV

2. MZ offset 4.4 (35 mW), DC voltage on oscilloscope 441mV

3. MZ offset 4.0 (15 mW), DC voltage on oscilloscope 186mV

As a reference, in LIGO, the SHG RIN is about 3e-6 from 3Hz to 400Hz. Then it gradually increases to 3e-5 at 4kHz. This indicates that the control servo has 1/f slope. (alog45088)

So we have a better RIN at low frequency (limited by PD dark noise). But a worse RIN at kHz region.

Images attached to this report
2980_20220617082333_figure1.png
Comments related to this report
YuhangZhao - 16:41, Monday 20 June 2022 (2983)

Yuhang, Aritomi, Marc

A new PD (PDA36A-EC) is used (gain is chosen as 0dB), firstly DARK NOISE is measured (saved as DARKL.txt and DARKH.txt)

A LA1608-YAG lens (75mm) was always used when the beam is apparently large.

1. Measure Mephisto RIN. It is measured at the location where we pick off main laser for CC PLL LO.

The DC voltage is measured as 2.66, 2.61V and 2.67V (measured before and after spectra measurements). (saved as MAINL.txt and MAINH.txt)

2. Measure homodyne LO RIN. It is measured just after IRMC transmission.

The DC voltage is measured as 800mV, 780mV (measured before and after spectra measurements). (saved as LOL.txt and LOH.txt)

3. Measure SHG generated GR RIN. It is measured before MZ.

The DC voltage is measured as 3.14V, 3.16V, 3.24V (measured before and after spectra measurements). (saved as SHGL.txt and SHGH.txt)

This is an experiment record. An analysis of data will come later.

YuhangZhao - 12:23, Tuesday 21 June 2022 (2988)

A comparison of RIN, shot noise and photo detector dark noise is shown in the attached first figure for several locations on the squeezer bench. Other figures show these noises for different single location separately.

KAGRA MIR (Scattering)
Print this report.
DanChen - 09:11, Friday 17 June 2022 (2981)Get code to link to this report
Some report links to klog about mirror scattering measurement
Let me put some klog links related to mirror scattering measurement.
  • 16 May 2022: "Laser scattering measurement (test samples and ITMY)" link
  • 07 June 2022: "Camera calibration for scatter light measurement of mirror" link
  • 16 June 2022: "Mirror scattering estimation" link
  • 17 June 2022: "Clean up of the mirror scattering measurement set up": link
KAGRA MIR (Absorption)
Print this report.
MarcEisenmann - 23:21, Wednesday 15 June 2022 (2979)Get code to link to this report
start of AZTEC #4 absorption

With help of Yuhang we removed #3 from the translation stage.

Then, I measured the bulk reference sample at the usual position and got R = 0.611 cm/W.

I installed the 1.5 inch sample and measured its absorption at the longitudinal center.

I measured about 35 ppm/cm which is compatible with old measurements.

Finally, again with Yuhang help, we installed the #4 and started absorption measurement.

KAGRA MIR (General)
Print this report.
MarcEisenmann - 23:19, Wednesday 15 June 2022 (2978)Get code to link to this report
Soleil-Babinet compensator actuator replacement

Marc, Yuhang

Yesterday I tried to remove the 4 screws  holding one of the 2 crystal.

Once they are removed, it is possible to access 2 set screws that fix the linear translation stage.

The 4 screws were really tight and I only manage to remove 3.

With Yuhang help, we tried to remove the fourth one without success.

In the end, we broke a screwdriver head inside the screw..

Fukushima-san from ATC helped us to remove it today.

After that the linear translation stage was replaced with a motorized one and the Soleil-Babinet is now closed with 3 out of 4 screws (need to purchase the broken one).

R&D (FilterCavity)
Print this report.
MichaelPage - 01:32, Wednesday 15 June 2022 (2977)Get code to link to this report
Mirror mount replacement along SHG path

Yuhang and Michael

We replaced the mirror mount for the beam splitter that goes into the SHG. The new mirror mount is much smaller, allowing us to more easily remove ghost beams and scattered light from the squeezer bench.

The beam splitter was replaced by roughly using an iris, SHG opening and GRMC EOM as beam position references. First, the beam splitter was moved by hand to align to the references. Fine tuning allowed us to recover the SHG mode matching to slightly better than the previous alignment condition.

Reference SHG alignment is shown in figure 1 and 2. BS model number is in figure 3. The mirror mount type is shown in figure 4.
 

Images attached to this report
2977_20220614183040_20220614172255.jpg 2977_20220614183044_20220614172246.jpg 2977_20220614183048_20220614191246.jpg 2977_20220614183155_20220614172204.jpg
Comments related to this report
YuhangZhao - 16:47, Monday 20 June 2022 (2984)

A proper beam dump thorlabs LB2 (1um - 12um) is installed for the newly replaced BS mount (as shown in Fig.1).

Although we can see some green light, let's remind that it is because the camera sensor is more sensitive to green as our eye. There are proper Dichroic mirrors to remove them.

KAGRA MIR (General)
Print this report.
MarcEisenmann - 20:36, Monday 13 June 2022 (2976)Get code to link to this report
AZTEC #4 delivered

Marc, Yuhang

Today the AZTEC #4 was delivered.

We brought it to the PCI cleanroom and applied first contact on one surface.