R&D (Cryogenic)
Print this report.
YoheiNishino - 14:05, Tuesday 13 September 2022 (3063)Get code to link to this report
Comment to measured gaussian beam radius (Click here to view original report: 3057)

Sorry for my late reply.

It's better to leave the actual value of beam radius [mm] with an error.

R&D (FilterCavity)
Print this report.
MarcEisenmann - 20:00, Thursday 08 September 2022 (3062)Get code to link to this report
Preliminary FIS measurement (2)

Marc, Yuhang

First we checked the LO alignment into the AMC and found misalignment in both pitch and yaw.

We tuned the Homodyne balance and checked the visibility.

After some tweaking we measured visibility about 0.92 % which is lower than expected (99%).

We found that the BAB was heavily misaligned in the AMC and after tweaking we could reach about 95.7 % visibility.

However, because our goal is to make sure that the system is in an understood state we decided to go towards FIS measurement.

(Actually we also found out that there were large scattering noise below 100 Hz, maybe due to grass cutting.. Even if the situation improved with the better homodyne visibility, we should also try to check this measurement at lower frequency)

We measured FIS with following conditions :

Green power [mW] 25 30 35 40
SQZ phase [deg] 90 90 90 95
ASQZ phase [deg] 160 160 150 145
OPO temp [kOhm] 7.15 7.154 7.161 7.17
p pol freq /5 [MHz] 39 39 39 39

Results are computed in figure 1 where we could reach 5.9 dB squeezing with 40 mW of green power.

With the fit of ASQZ versus SQZ in figure 2, we have optical losses = ( 20.9+/-0.93 ) % and phase noise = (25.46 +/-3.14 ) mrad.

From our wiki, we expect 19.1% optical losses but taking into account the lower visibility, the expected losses become 23.6 %.

For the phase noise, we have quite large uncertainty because from 45 mW of green power the CC1 became really noisy. We will investigate this issue tomorrow.

During the alignment tuning, I made a mistake and moved the pitch of the steering mirror close to the mirror labelled 'FIS'. Before measuring FDS we should make sure to recover its alignment using the irises on the table.

Images attached to this report
3062_20220908125621_fis.png 3062_20220908130212_fit.png
R&D (FilterCavity)
Print this report.
MarcEisenmann - 12:15, Wednesday 07 September 2022 (3061)Get code to link to this report
input oplev sensing

Marc, Yuhang

We found that one SR560 that was used for the input oplev yaw was broken (increasing a lot the noise above 10 Hz) so we replaced by the one used for the length sensing oplev.

We want to find the good PSD position and first tried to measure the natural length mechanical resonance in the tilt PSD without excitation but we could not see any signal at the expected frequency (0.94 Hz).

We decided to inject white noise in the length of INPUT at this frequency.

First we used 5000 counts amplitude but we could see harmonics so we decreased the amplitude to 1000 counts.

The attached figure report the 0.94 Hz peak amplitude in the pitch and yaw spectrum while moving the tilt PSD (larger position correspond to moving the PSD closer to the INPUT mirror).

For the yaw, we can see the expected 'V shape' in addition to some constant noise which might be due to non perfectly decoupled driving.

The minimum of this 'V shape' is at about 7.3 cm which is really close to the waist position.

For the pitch, we can clearly see the 'V shape' but the minimum position is about 3 cm which seems to be really off.

We have to investigate the reason for this discrepancy.

Images attached to this report
3061_20220907051530_oplev.png
R&D (FilterCavity)
Print this report.
MarcEisenmann - 11:42, Wednesday 07 September 2022 (3060)Get code to link to this report
preliminary FIS measurement

Marc, Yuhang

In order to prepare for OPO measurement, we measured some preliminary FIS spectrum.

New measurements with better tuning are planned.

 

We checked the OPO non-linear gain after realigning the BAB and GR beam into it.

we measured without green 550 mV in transmission and with 25.6 mW of green 2540 mV which corresponds to non-linear gain of 4.47.

This is a bit lower than expected (should be about 5).

We tried to tune the OPO temperature (starting at 7.130 kOhm with 10 Ohm increment) but it seems we were already around the optimal temperature. The p pol PLL frequency in this condition was 225 MHz.

We also had to realign the GRMC and had some issue with the error signal sign before locking it.

We tuned the Homodyne balance by injecting noise into the IRMC (sine with 1.2 V Pk2Pk at 800 Hz) and trying to minimize it from the homodyne sub DC channel.

We checked the LO and BAB matching into the AMC and after some tuning could reach about 99.9 % mode-matching for each field into the AMC.

Finally we measured shot-noise and FIS.

We tried to change the phase of the FIS but could not reach more than 4 dB.

 

Possible reasons are that we forgot to optimize the OPO non-linear gain or the Homodyne visibility.

Furthermore, we had to tweak alignment at several places so we might have to retune it today to have a more stable alignment.

Images attached to this report
3060_20220907044230_fis.png
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 04:11, Wednesday 07 September 2022 (3059)Get code to link to this report
Comment to DDS reset and broken input oplev laser (Click here to view original report: 3058)

Despite the fact that they have a power knob, the laser model that we use for oplev don't allow to tune the output power. 

R&D (FilterCavity)
Print this report.
MarcEisenmann - 15:04, Thursday 01 September 2022 (3058)Get code to link to this report
DDS reset and broken input oplev laser

Aritomi, Marc, Michael, Yuhang

We reset the DDS rack, restarted the control PC and reloaded all configurations.

We checked input oplev and found that the laser output was really low (sum read by PSD was about 150 counts while usual value is about 5000).

We replaced the laser head by a new one and could reach up to 7000 counts.

To avoid the saturation, we tried to reduce the laser power, but we couldn't tune the laser power at all...

In any case, we installed 2 OD1 and could reach about 5000 counts on the PSD sum.

Comments related to this report
EleonoraCapocasa - 04:11, Wednesday 07 September 2022 (3059)

Despite the fact that they have a power knob, the laser model that we use for oplev don't allow to tune the output power. 

R&D (Cryogenic)
Print this report.
KoheiMitsuhashi - 21:55, Monday 29 August 2022 (3057)Get code to link to this report
measured gaussian beam radius

What I did : I measured gaussian beam radius and fited beam propagation using beam profiler and Python.

Why I did: trying for Pound-Drever-Hall method.

Experiments: I set up the following in a attached picture.

Result:The result of fitting is also in a attached picture.

Conclusion: I could have measured the beam radius.

  •  
Images attached to this report
3057_20220829144440_beampropagation.png 3057_20220829151751_setup.png
Comments related to this report
YoheiNishino - 14:05, Tuesday 13 September 2022 (3063)

Sorry for my late reply.

It's better to leave the actual value of beam radius [mm] with an error.

R&D (Cryogenic)
Print this report.
KoheiMitsuhashi - 17:48, Friday 26 August 2022 (3055)Get code to link to this report
resolving problems of optical element mixing.

Problems:There are some possibility that I mixed optical elements in different wavelength bands (1550nm and 1064nm) in ATC.

What I did: I checked the suspected  optical elements  used for 1550nm or not.

Subjects: Two HWP and one QWP that were suspected in 1064nm,but one HWP(named "1") is cleaned by using first contact. So I checked one HWP(named "2") and one QWP(named "3").

Experiments: I set up the following in attached pictures. In experiment for the suspected HWP(named "2"),maximum mW and mimimum mW showed in Power Meter was as follows.

  Maximum (mW) Mimimum (mW)
No ND filter 11.64 0.02
ND 0.2 filter 8.84 0.01
ND 1.0 filter 1.59 0.01

In experiment for the suspected QWP(named "3"), maximum mW and mimimum mW  showed in Power Meter was as follows.

  Maximum (mW) Mimimum (mW)
No ND filter 3.51 0.07
ND 0.2 filter 2.67 0.06
ND 1.0 filter 0.48 0.01

 

Conclusion: Based on the above results, it can be concluded that HWP(named "2") and QWP(named "3") are for 1550nm wavelength.

Next: I will check the suspected HWP(named "1").




Addition: I checked WHP(named "1"),WHP(named "2"),and QWP(named "3")again.I set up the following in additional attached pictures.Additional experiments were set up the same as before.

 In experiment for the suspected HWP(named "1"),maximum mW and mimimum mW showed in Power Meter was as follows.

  Maximum (mW) Mimimum (μW)
No ND filter 11.65 8.47
ND 0.2 filter 8.972 6.57
ND 1.0 filter 1.679 1.19

 In additional experiment for the suspected HWP(named "2"),maximum mW and mimimum mW showed in Power Meter was as follows.

  Maximum (mW) Mimimum (μW)
No ND filter 11.59 8.40
ND 0.2 filter 9.039 6.48
ND 1.0 filter 1.762 1.23

In additional experiment for the suspected HWP(named "3"),mW showed in Power Meter at point A,B,and C was as follows.

  point A point B point C
Power Meter(mW) 12.14 3.693 3.826


Based on the above results, it was concluded that HWP(named "1"),HWP(named "2") and QWP(named "3") are for 1550nm wavelength.There was no mixing.

Images attached to this report
3055_20220829081042_pointabc.png 3055_20220829151618_hwpnamed1setup.png 3055_20220829151641_hwpnamed2setup.png 3055_20220829151649_qwpnamed3setup.png
R&D (FilterCavity)
Print this report.
MichaelPage - 20:19, Wednesday 24 August 2022 (3053)Get code to link to this report
TAMA DGS upgrade - Data Concentrator/Server Debian 11 installation

I installed Debian 11 on the Data Concentrator/Network Data Server/Frame Writer computer in TAMA. In 2951 we ran across a problem where we couldn't complete the software installation for the Debian 11 install, and for some reason couldn't install a boot loader properly. The problem was that the installation medium was running in BIOS mode as opposed to the more appropriate UEFI mode. Switching to UEFI installer upon startup made the installation go smoothly as for the frontend computer.

The following options were chosen during installation:

Language - English
Region - Japan
Locales - en_US.UTF-8
Keyboard - American English
hostname - DCCentral
domain name - mtk.nao.ac.jp (default)
Network interface - eno1 Intel Corporation I210 Gigabit Network Connection
root password - normal KAGRA controls password
ops user 1000 (same password as usual)
controls user 1001 (same password as usual)
package manager - deb.debian.org
HTTP proxy - blank
 

The following drive partitioning scheme was chosen:

Disk 1 - 1 TB Western Digital ATA - for operating system and user files
512 MB - EFI System Partition (ESP) - in the failing step last time, using the BIOS mode installer would not let us install this partition. Now it's fine.

128 GB swap space
Remaining (~870 GB) - ext4 / - perhaps there's no real need to separate into /home, /var, /tmp here, so the system can just figure that out by itself

Disk 2 - 24 TB Avago MegaRAID virtual drive (4x 7.6 TB disks with RAID5 redundancy) - for mass storage of channel data
24.0 TB - ext4 /data0 - used for frame writer. In KAGRA, there are two frame writer PCs k1fw0 /data0 and k1fw1 /data1 with about 25 TB storage each, for extra data redundancy. But we only have one server PC, so I just assume we are using one frame writer.

 

Afterwards, I added ops and controls to the sudoers file and installed cdssoft version 1.0.11 (was 1.0.9 when installing frontend) from the aligo caltech Debian Bullseye repository.

*I will comment later in a bit more detail about what some of these technical terms mean

R&D (FilterCavity)
Print this report.
MarcEisenmann - 15:42, Wednesday 24 August 2022 (3052)Get code to link to this report
FC lock issues

Marc, Yuhang

We have to reload DDS2 config everytime to be able to lock.

Then we are looking at the GR_corr spectrum to investigate the driving matrix of the INPUT but we found out that the noise level is not stable at all (see figure 1).

Our reference level is in blue with injected line on INPUT_PIT_ex2 at 5 Hz and 5000 amplitude is blue curve of figure 1.

At other times, we can see noise increase up to 10 Hz or even at higher frequency.

This is true with or without line injected, with or without AA or pointing loop.

Same after restarting laser.

Not affected by closing AA or pointing loops.

After some time, the noise excess disappeared and we could start some measurements but it quickly reappeared again..

PIT_ex2 4500, GR_corr 7.52

4000, 6.70

3500, 6.47

3000, 5.97

2500, 5.38

Images attached to this report
3052_20220824074446_20220823grcorrnoise2.png
R&D (FilterCavity)
Print this report.
MarcEisenmann - 20:07, Monday 22 August 2022 (3049)Get code to link to this report
FC relock

Marc, Yuhang

We found that GR reflection was far from overlapping with injection.

This was due to huge offset on input and end because AA loop was still closed.

After opening the loop, we could see TEM00 flashes but we had to have about 100 offset in input and end pitch.

We could not lock the FC so we checked the GR power while changing the SHG temperature.

GR power after FI (mW): 41.7, 45, 35, 44, 45

SHG temperature: 3.085, 3.071, 3.06, 3.0665, 3.075

Finally, we tuned the SHG alignment and could recover more than 50 mW after the FI.

We could still not lock so we tried to tweak the servo parameters :

gr lock servo, attenuation = 0.2, gain = 8, both changed no diff

gr reflection power > 200 uW

Finally, after changing the lock threshold we locked on a sideband. Turning off the servo seems to change the sign of the error signal and we could lock the FC.

In the end, we changed sign on DDS2 by reloading the latest configuration.

KAGRA MIR (General)
Print this report.
MarcEisenmann - 20:03, Monday 22 August 2022 (3050)Get code to link to this report
birefringence measurement of AZTEC #4

I set IR beam power to 1.6mW, I checked vertical and horizontal AOI with the razor blades and got respectively -0.008 deg and -0.022 deg.

I tuned the QWP and HWP angle to minimize p polarization while injection s polarization.

I did the calibration and started measurement of #4 (was cleaned with first contact) with s polarization at the input.

R&D (FilterCavity)
Print this report.
YuhangZhao - 19:32, Monday 22 August 2022 (3048)Get code to link to this report
Optimization of driving matrix for filter cavity input mirror

Yuhang and Marc

We would like to have 'tilt driving' coupling to 'shift driving' as small as possible. To achieve this, we worked on finding optimal driving matrix for H1 and H3 for pitch today. 

At the beginning, the driving matrix for pitch is: H1 0.794, H3 1. This was optimized in the past. However, we would like to optimize it again since we glued new magnets at the beginning of this year.

To check how much coupling we have, we sent 5Hz 5000 amplitude for INPUT_PIT_ex2. We checked coherence between GR_CORR and excitation, no coherence was found.

Then we checked when driving matrix is: H1 1, H3 1. We found no coherence even in this case. We checked the connection, we found later on that actually GR_CORR is not connected.

After putting back good connection of GR_Corr, we take data when H1 and H3 are both 1. We saved as reference as blue (REF0 and REF1).

H1 0.794, GR_corr 17.964 (REF2, REF3)

H1 1,         GR_corr 30.475

H1 0.6,      GR_corr 13.38

H1 0.5,      GR_corr 15.82

H1 0.55,    GR_corr 13.82

H1 0.65,    GR_corr 13.09

H1 0.65, H2 0.02,  GR_corr 12.63

H1 0.65, H2 0.05,  GR_corr 13.61

H1 0.65, H2 0.01,  GR_corr 12.30

H1 0.65, H2 -0.01,  GR_corr 12.82

When we drive H1, we get GR_corr 87.08

When we drive H3, we get GR_corr 63.24

We found a ratio of 0.73. Then we did H1 0.73, we get GR_corr 10.71

Attached figure shows a comparison of GR_corr when H1 is 1 or 0.73.

We still have coupling as we see the number in GR_corr 10.71. We still would like to understand it better what is causing this number. One thing we will check is the coupling factor when different amplitude signal is sent to H1 and H3. We will also test this coupling number by sending excitation at a different frequency.

Images attached to this report
3048_20220823082006_20220822inputpitchh10794.png
KAGRA MIR (General)
Print this report.
MarcEisenmann - 11:51, Friday 29 July 2022 (3047)Get code to link to this report
Packing of AZTEC #1 and #3

Aso, Marc, Matteo

We put marking on AZTEC #1 and #3 to indicate the orientation of the sapphire during shaping.

One large arrow on the top pointing towards the front surface during our measurement + 2 long lines on the side at the level of the holder.

We packed the samples inside their wooden crates with foam to prevent their movements during shipping.

Images attached to this report
3047_20220729045031_topmarking.jpeg 3047_20220729045055_mirrorinbox.jpeg 3047_20220729045100_crate.jpeg
KAGRA MIR (Polarization)
Print this report.
MarcEisenmann - 17:01, Tuesday 26 July 2022 (3046)Get code to link to this report
Trial to rotate AZTEC #1 by 90 deg

Marc, Matteo

We rotated the sample by 90 degrees and fine tuned the rotation angle to minimize p polarization power at the readout.

Results are attached to this entry.

It seems that previous position was better so today we rotated the sample back to this position.

Actually, trying to minimize the p polarization after the sample, we rotated clockwise from injection part by an additional 2.54 deg.

From HWP, it seems that we are now at 0.6 deg from theta = 0 deg.

Images attached to this report
3046_20220726100130_dntheta90deg.png 3046_20220726100134_convloss90deg.png
KAGRA MIR (Scattering)
Print this report.
HomareAbe - 09:48, Monday 25 July 2022 (3045)Get code to link to this report
Mistake in calibration of scattering measurements
Abe, Dan,

We noticed a mistake in the calibration step in the scatter measurement.
We assumed that the camera was receiving all of the light emitted from the integrating sphere, when in fact it was receiving only 0.43%.
This reduces the result for all scattered light to 0.43%.
In other words, the 7/22 result is 4~7%*4.3*10^(-3)= 1.7~3.0*10^(-2)% ,instead of 4~7%.
(The reason for the wide range of values is that the intensity of the scattered light has an angular dependence.)

This means that the scattered light is about 170~300ppm.
KAGRA MIR (Polarization)
Print this report.
MarcEisenmann - 19:35, Friday 22 July 2022 (3044)Get code to link to this report
corrected AZTEC #1 birefringence

Following the solved issue of PSD saturation, I computed again calibration parameters (see figures 1 to 3).

Then, I computed polarization map with input polarization s (figure 4) and 15 deg (figure 5).

From these measurements, I could compute delta n and theta (figure 6) as well as polarization conversion losses (figure 7).

The 2 methods to estimate these losses agree well with each other and show losses at about 0.74 % !

We will fine tune the mirror rotation angle (seems we still have an offset about 2.37 deg or 4.65 arc length cord).

Images attached to this report
3044_20220722123120_calibmeasurement.png 3044_20220722123124_calibparameters.png 3044_20220722123127_iscalib.png 3044_20220722123522_pol0.png 3044_20220722123525_pol15.png 3044_20220722123529_dntheta.png 3044_20220722123535_conversionloss.png
KAGRA MIR (Scattering)
Print this report.
HomareAbe - 11:23, Friday 22 July 2022 (3043)Get code to link to this report
Calculation of Aztec3 scattering
Abe, Dan

We calibrated IR camera sensitivity and analyze the scattering data.

According to our result, Aztec3 has 4~7% scattering for IR light.
Images attached to this report
3043_20220722040832_20220722094854.jpg 3043_20220722040900_countvsenergy.png
KAGRA MIR (Polarization)
Print this report.
MarcEisenmann - 19:16, Thursday 21 July 2022 (3042)Get code to link to this report
PSD were saturating

Previous entries reported strange anti-correlation of s and p polarization power while not changing polarization.

Furthermore we had also s polarization power too constant while rotating the mirror.

Today I found out that the PSD were saturating despite neither the PSD displays nor the lockin-amplifier showing saturation.

This is because I forgot to reduce the laser power after checking the beam parameter with the razor blades.

The power is now decreased from 14 mW to 1.6 mW.

I redid the calibration measurement and started measurement with rotated sample and s polarization.

R&D (FilterCavity)
Print this report.
NaokiAritomi - 16:25, Thursday 21 July 2022 (3041)Get code to link to this report
AA recovery

I found that green reflection for AA was misaligned a lot. I aligned steering mirrors for AA to make the green beam at the center of both QPDs. DC centering loop can be locked, but AA loop could not be locked. I checked AA setting and it was completely different from previous setting in elog2850 (Fig. 1). I brought it back to the previous setting and AA loop seems working now.

To check the IR alignment, I injected BAB to FC. I could not find BAB transmission, but I found BAB reflection from FC. I will maximize the BAB reflection and recover IR alignment soon.

Images attached to this report
3041_20220721092457_aaconfig.png