LOG-IN
Displaying reports 2401-2420 of 3128.Go to page Start 117 118 119 120 121 122 123 124 125 End
R&D (FilterCavity)
Print this report.
YuhangZhao - 18:29, Friday 15 June 2018 (827)Get code to link to this report
Test lock of green mode cleaner and the mechanical transfer function of it

Participant: Marco, Yuhang and Eleonora

During the last few days, we installed the RF photodiode for mode cleaner reflection. We achieved the lock and measured the mechanical TF.

For the power supply of RF PD, we used a T to seperate the power from CMS 100-05.

For coupling the beam into the PD, we put a 75.6mm lens before PD. In order not to saturate the PD we put a OD 2 to attenuated power. After the OD, we got power around 200uW. Now, we can get the sideband information(78MHz) and create the error signal.

The demodulation frequency(78MHz) of mode cleaner is the same with the demodulation of filter cavity. The demodulation signal(LO) is provided by DDS2 board. We also maximize the error signal by select the best phase (175 degree).

We use SR560 for a temporarily lock. We set a first order pole at 3Hz with a gain of 10. Then we measured the mechanical transfer function of green mode cleaner (cavity + piezo). It is attached as Fig 1. As expected, it is flat at low frequecy. The structures due to the piezo resonances start from about 7 kHz.

Images attached to this report
827_20180615112940_mtfofgreenmodecleaner.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 17:35, Friday 15 June 2018 (826)Get code to link to this report
Polarization of beam going to green-mode-cleaner

I used the PBS to check the polarization. The reflection of PBS is 12mW while the transmission is 11uW. This means the polarization is good s-polarization.

R&D (FilterCavity)
Print this report.
MarcoVardaro - 16:51, Friday 15 June 2018 (824)Get code to link to this report
Swap AOM RF input amplification to the amplifier box

Partecipants: Marco, Eleonora, Yuhang

Before:  the AOM inpu RF signal was amplified by a +15dB ZHL-2 Minicircuit amplifier. Its input signal is generated by a Tecktronix AFG351 Frequency generator and its power level was set to +11 dBm

After:  we connect the RF generator output to the first channel of the amplifier box. This contains a +37dB amplifier, thus its input power level was reduced from +11 dBm to -10 dBm.

R&D (FilterCavity)
Print this report.
MarcoVardaro - 16:43, Friday 15 June 2018 (823)Get code to link to this report
Swap to DDS 2 Board

participants:  Marco, Eleonora, Yuhang

- Filter cavity modulation:

Before Tektronix AFG3102 Fout=78MHz Ampl=1Vpp (Measured with HP 8563E Power 2.83dBm Fout+78.017MHz)

After DDS2 Board CH0 Fout=78MHZ Full scale output (Measured wit HP: 8563E Power after amplificator 13.6 dBm)

I put a 6 dBm SMA attenuator and I set the output at 1/2 scale after it I measure 4.6 dBm at 78.3 MHz

 

- Filter cavity PDH demodualtion:

Before provided by Tektronix AFG3102 Fout=78MHz Ampl=9.5 Vpp (the mixer was a class 23 Minicircuit ZAY-3)

After DDS2 board CH1 Fout = 78MHz Full scal output. I measure 6.43 dBm after the amplifier.

We replace the mixer with a Class 7 Mixer Minicrcuit ZAD-1-1+)

 

- GrMC reflection PDH demodulation:

Now DDS2 Board CH3 Fout=78MHz Full scale output. I measure 6.43 dBm after the amplifier.

 

Note that the CH2 of the DDS 2 Board is not working

 

Phase optimization

CH0 = 0

Ch1 = 135 deg (before was 125 deg)

CH2 = not working

CH3 = 175 deg

R&D (FilterCavity)
Print this report.
MatteoLeonardi - 10:51, Friday 15 June 2018 (822)Get code to link to this report
RF amplifier box
Just to remind the amplification values of the RF amplifier box realized by Pierre Prat.

AOM channel: AMPA-B-34-20.425 -> +37dB
EOM channels: ZFL-500HLN+ -> +20dB
demod channels: ZFL-1000+ -> +13dB
PLL channels: ZFL-1000+ -> +6dB
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 21:27, Wednesday 13 June 2018 (820)Get code to link to this report
IR beam realigned into the cavity

We have checked and quickly recovered the alignement of the IR beam into the filter cavity.

R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 17:07, Wednesday 13 June 2018 (819)Get code to link to this report
Characterization of the main laser beam for PLL

Participants: Yuhang, Eleonora

We have characterized the IR beam from the main laser transmitted by the first BS. We have measured the beam starting from the 48th hole (which is about 6 cm after the lens with f = 500 mm).

The beam is quite collimated: it has a waist of 0.67 mm ad a distance z = +2.9 m from the 48th hole. In the region around the 50th hole it has a diamiter of about  3 mm.

We plan to take a pickoff of the beam in this region and use it for the PLL. We will inject it into a fiber through a collimator. Accoring to the collimator characterization reported here, we need to prepare a collimated beam with a diameter of about 2 mm.

A possible solution for the telescope is to use two lenses with f = 75 mm and f = 50 respectively,  placed at a distance of about 12.3 cm from each other.  See in the attachement the jammt simulation.

Images attached to this report
819_20180613100524_telescopemainirpll.png
KAGRA MIR (Absorption)
Print this report.
ManuelMarchio - 15:48, Wednesday 13 June 2018 (811)Get code to link to this report
Tests on the LMA coated samples

I did some tests with the 633nm probe on the LMA samples that I already measured some years ago with the original setup. elog entry: http://www2.nao.ac.jp/~gw-elog/osl/?r=141

SCANS

calibration 633nm: z=34.5mm  power 33mW; AC=0.2017V; phase=-103.8deg; DC=3.38V freq=375Hz filename=Fri, Jun 08, 2018 6-20-48 PM.txt

median filter order 5, average filter order 5

lasr current 1.3A, angle maximum IPC

lma15033 power=960mW z=34.5;  AC=0.000651 phase=-104; DC=4.52V  freq=375Hz filename=Tue, Jun 12, 2018 11-22-12 AM.txt
absorption = 16.4ppm (nominal 12.8ppm)

lasr current 4A, angle maximum IPC

lma15033 power=5090mW z=34.5;  AC=0.00305 phase=-105; DC=4.62V  freq=373Hz filename=Tue, Jun 12, 2018 3-46-17 PM.txt
absorption = 14.2ppm (nominal 12.8ppm)

lma15032 power=5130mW z=34.5;  AC=0.001407 phase=-105; DC=4.7V  freq=378Hz filename=Tue, Jun 12, 2018 4-12-54 PM.txt
absorption = 6.4ppm (nominal 4.5ppm)

 

median filter order 20, average filter order 20

lma15034 power=5130mW z=34.5;  AC=90uV phase=-110; DC=4.28V  freq=375Hz filename=Tue, Jun 12, 2018 4-49-39 PM.txt
absorption = 0.47ppm (nominal 0.65ppm)

laser current 7.5A

lma15034 power=10W  filename= Tue, Jun 12, 2018 5-18-22 PM.txt
absorption = 0.44ppm (nominal 0.65ppm)

MAPS

lma15033 power=960mW  filename = Tue, Jun 12, 2018 11-46-14 AM.txt
mean (std) on the map =  15.3 +/- 1 ppm

lma15034 power=10W  filename = Tue, Jun 12, 2018 6-08-48 PM.txt
mean (std) on the map = 0.43 +/- 0.08 ppm

Images attached to this report
811_20180613075709_15033.png 811_20180613080849_15032.png 811_20180613081729_15034.png 811_20180613084813_15033map.png 811_20180613084819_15034map.png
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 15:08, Wednesday 13 June 2018 (818)Get code to link to this report
Collimator alignment

Participants : Yuhang, Eleonora

By sending the fiber output to the collimators (PAF-X-7-C) and (PAF-X-11-PC-C) we got the output beam and we could adjust the z of the collimator to make the beam collimated, and x and y to make it go straight.

 We used beam profiler to measure the dimension of the collimated output beam:

  •  PAF-X-7-C  has a diameter of  2000um
  •  PAF-X-11-PC-C has a diameter of  3300um

We put a label on the collimator holders to specify the beam diameter that we measured  (that is the dimension of the collimated beam that they need in input).

They are now well aligned and should not be touched.

More info on the collimator mechanics and aligment procedure can be found here:

https://www.thorlabs.com/drawings/95fa854d421b0d3c-A5221257-C242-648D-DA41342A6D75B16A/PAF-X-11-PC-C-Manual.pdf

Useful tip to adjust the z of the collimator: if the beam is diverging the z screws should be turned counter clock wise, if it is converging they should be turned  clockwise.

R&D (FilterCavity)
Print this report.
MatteoLeonardi - 18:24, Tuesday 12 June 2018 (817)Get code to link to this report
Comment to Broken collimator (Click here to view original report: 809)
A screw identical to the broken screw have been found in the collimator box. This screw is likely the lock screw for the XY position of the collimator lens.
This spare screw has been used in place the broken one and the collimator has been riassembled and it seems to work.
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 16:03, Tuesday 12 June 2018 (816)Get code to link to this report
Filter cavity lock recovered

Partecipants : Yuhang, Marco, Eleonora

During the auxiliary lasers installation performed in the past week  the green beam got a bit misaligned and  the filter cavity was not locked since then.

Today,  in order to recover the lock, we have at first realigned the green beam, than we have zeroed the error signal for the local control of the input mirror since the stanford was saturating.

Note that the last time we had to zeroes the local control was more than 2 month ago.

The SHG temperature control was set to 3.175 kOhm and we injected about 8.5 mW of green light  into the viewport.

The local control offsets for the optimal alignment were found to be

  pitch yaw
BS 0.3 -0.07
PR -0.76 -0.34
IM -0.95 0.45
EM -1.73 2.82

The transmitted power is about 1.6 V.

The IR beam on the other hand is not aligned anymore, since we are going to take a pick off to be used to the aux lasers PLL.

R&D (FilterCavity)
Print this report.
YuhangZhao - 14:25, Tuesday 12 June 2018 (814)Get code to link to this report
Aux2 (p-polarization) installation and fiber coupling

Participant: Eleonora, Marco, Yuhang and Matteo

Input beam: waist of 192e-6 m at a distance -0.2 m from the laser head. Characterized here.

The input power is 232mW. After put two waveplates, the power becomes 228mW.

Telescope

First lens (f = 50) at a distance 13.5 cm from the laser head.

after the lens we measured  a beam with a  waist  of  33.6e-6 m at z = 0.188 m from the laser head

Second lens (f=175) at a distance 39 cm from the laser head.

after putting the second lens, we adjust it and make it collimated. The beam diameter now is 3800e-6m.

Faraday isolator: input power 227mW output power: 202 mW  Throughput 89%

Semireflective mirror: input: 180 mW  reflection: 174 mW trransmission 2.7 mW

Fiber collimator: input power 2.6 mW output power 2* 0.88 mW.  Power coupled 68%

R&D (FilterCavity)
Print this report.
MatteoLeonardi - 14:08, Tuesday 12 June 2018 (812)Get code to link to this report
Assembly of OPO (part 1)

In this report the first part of the assembly of the OPO cavity. For the drawing refer to file "assiemepdopov3asm.pdf". The pictures have been uploaded on drive at the following link: https://drive.google.com/open?id=1bjSdCoMSOHeWpJarAu0ZhKGuFZZME19h

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Step 1: check peltier (21) hot/cold side. In order to check which side of the peltier is the hot or cold one, the peltier has been powered up and the sides have been checked manually to determine which is the hot and which is the cold one. For this particular type of peltier, the cold side is the one with text (XF29) while the hot one is blank. See picture 1 and 2. Since we need the peltier to heat our crystal up, the cold side has been mounted in contact with the OPO assembly baseplate (6) while the hot one will be mounted in contact with the copper lower L (8). In order to ensure a good thermal contact between the components, a sheet of 0.1mm indium has been placed in between the peltier and the copper L.

Step 2: mount the thermometers (24) on the copper lower L (8) with the retainer (22, 23). See picture 3 and 4. After mounting the thermometers, bend the wiring of the thermometers and ensure the wires go in the groove, see picture 5.

Step 3: apply indium on the copper lower L and on the macor upper L (9). See picture 6.

Step 4: place the PPKTP crystal (7) in position. See picture 7. The PPKTP crystal mounted is the one with part number 7-11159316418, see picture 8. The crystal has a black dot on one side. I suppose this indicates the HR side, so the crystal has been mounted accordingly, see picture 9.

Step 5: assemble the lower and upper L on the peltier element and close the assembly with the POM bridge (5), see picture 10. Before doing this step, make sure the two POM tipped set screw (27) have been completely relased. While mountin the POM bridge be very careful about the crystal positioning. During this part one issue has been found and solved, see "Problem 1" part.

Step 6: fix the crystal position acting on the POM tipped screws. See picture 11. On this step be very careful not to brake the crystal. Alternate acting on X and Y screws, starting from the Y one. If you see the macor L starts tilting up, while acting on X, act on Y to fix the tilt. The final position of the assembly is when the peltier, and both the lower and upper L touch the left side of the POM bridge.

Step 7: mount the crystal holder retiner (25). See picture 12.

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Problem 1: during the step 5 of the assembling procedure an issue has been found. The Hexagon Socket Low Head Cap Screws (26) was not a "Low head cap screw", therefore this screw was interacting in a bad way with the POM tipped screw for X alignment, and was preventing the complete relese of this screw, necessary to start step 5 (see picture 13). This problem has been solved substituting this screw with a different screw with lower head cap (see picture 14 and 15). This solution allowed the release of the POM tipped screw for X alignment without problems.

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

List of picture:
01 - peltier cold side
02 - peltier hot side
03 - copper lower L, thermometers and retainer before assembling
04 - copper lower L, thermometers and retainer after assembling
05 - lower L with thermometers in place and termometers wires bended
06 - indium applied to the lower and upper L
07 - crystal on the lower L
08 - crystal part number
09 - black dot on the crystal side indicating the HR (??) side
10 - POM bridge assembled
11 - crystal positioning using the POM tipped screws
12 - part 1 of OPO assembly finished
13 - problem with compenetrating screws
14 - new screw compared to old one
15 - compenetrating screw problem solved

Non-image files attached to this report
R&D (FilterCavity)
Print this report.
MatteoLeonardi - 10:58, Tuesday 12 June 2018 (809)Get code to link to this report
Broken collimator
Few days ago we noticed that one of the two 11mm EFL collimators was behaving in a strange way. In particular the lens was not moving along X direction. I unmounted the collimator and found that the screw for X movement was broken (see picture 1 and 2).


NOTE: if you look for the collimator into the Thorlabs site the part number is PAF2P-11C. Thorlabs changed very recently the part number of this item from the one written on top of the collimator to this new one.
Images attached to this report
809_20180612035840_brokenfibercollimator1.jpg 809_20180612035850_brokenfibercollimator2.jpg
Comments related to this report
MatteoLeonardi - 18:24, Tuesday 12 June 2018 (817)
A screw identical to the broken screw have been found in the collimator box. This screw is likely the lock screw for the XY position of the collimator lens.
This spare screw has been used in place the broken one and the collimator has been riassembled and it seems to work.
KAGRA MIR (Absorption)
Print this report.
ManuelMarchio - 17:50, Monday 11 June 2018 (808)Get code to link to this report
damping the pump reflection

- Since the pump is now at 2° incidence, the reflection from the sample was hitting the mount of the periscope. Without moving the periscope, I collected the reflection with a mirror and I redirected it outside with another mirror through a hole in the enclosure. Then I put a blackhole damper to absorb the power. I decided to bring the power outside to avoid heat generation inside the enclosure that may cause temperature fluctuations and noise at low frequency on the absorption signal.

- I also added a shutter after the chopper, to be able to close the pump for few seconds (not more otherwise the shutter melts) and be able to move the sample away from the pump path without burning the sample mount.

- The pump path at the ground level of the optical table was covered with a black paper enclosure. I replaced it with black anodized aluminum panels to reduce the generation of dust.

Images attached to this report
808_20180611104353_372.png 808_20180611104859_371.png 808_20180611105549_whatsappimage3.png 808_20180611105819_shutter.png
KAGRA MIR (Absorption)
Print this report.
ManuelMarchio - 16:57, Monday 11 June 2018 (798)Get code to link to this report
pump angle 2deg to avoid fabry perot effect on the reference

In order to solve the problem of the pattern on the surface reference sample I had to move the pump probe by 2°. The resulting path was outside the periscope mirror, so I had to move the periscope. It also went outside the chopper, so I had to move the chopper as well. First I aligned the pump beam without the focusing lens and measured the transmitted peak from the pinhole at different Z
  Y                 Z
121.384100  55
120.684361  35
119.984621  15
angle 0.034987rad = 2.004deg

Then I put the focusing lens again and  after putting the lens maximize the AC signal on the surf ref sample. Then measured again the transmitted peak position:

  Y                 Z
121.923354 55
121.28          35
120.57          15
angle = 1.94deg

and moving the pinhole around the crossing point I maximized the transmitted power of both pump and probe at the following coordinates:
X 327.432
Y 121.255
Z   34.900

this is the new position of the crossing point

The first screenshot is the surface reference scan along Z.
The second screenshot is the surface reference scan along Y, that shows that the pattern amplitude is now of about 5%
The third screenshot is the bulk reference scan along Z.

I also made a map of the surface reference sample
Please note that the map Absorption[ppm] axis is not calibrated.

Images attached to this report
798_20180605085225_37.png 798_20180605085704_40.png 798_20180605090731_10.png 798_20180611095612_18.png 798_20180611095617_44.png
KAGRA MIR (Absorption)
Print this report.
ManuelMarchio - 16:33, Monday 11 June 2018 (805)Get code to link to this report
power measurements

Manuel, Eleonora

I checked the pump power vs laser current with the power meter PM100D with the head S145C ( Integrating Sphere Photodiode Power Sensor, InGaAs, 800 - 1700 nm, 3 W ).
It doesn't work with the chopper modulation, so I turned off the chopper. The angle of Input Power Controller (IPC) is set around the minimum of transmitted power because the power meter head has 3W of range.
Result: see the first plot.
Every time the current is changed, the laser needs some time (about 20 min) to stabilize. Since I changed the current without waiting for this time, the plot is noisy.

Then I set the laser current to 1.5A and rotated the angle of the IPC to change the power in a more stable way. I measured the power for different angles, without the chopper modulatoin, with two power meters: the PM100D-S145C and the Ophir-Vega (that has a larger range). Then we repeated the measurement with the chopper on and measured it with the Ophir-Vega to check if the ratio with and without chopper is 0.5.

angle P100D   vega                chopper
 270  93.5mW  vega 75mW     37mW
 275  139mW   vega 125mW   52mW
 280  330mW   vega 305mW   150mW
 285  640mW   vega 595mW    290mW
 290  1080mW vega 1020mW  510mW
 295  1470mW vega 1420mW  725mW
 300  1940mW vega 1840mW  915mW
 305  2320mW vega 2190mW  1095mW
 310  2570mW vega 2420mW  1215mW
 315  2680mW vega 2530mW  1270mW
 
Result: see second plot. From the linear fitting there is an offset of 5.6mW and the Ophir-Vega power meter gives 5% lower value than the PM100D-S145C.
 
Third plot is the ratio between the Ophir-Vega measurement with chopper and without chopper.
the ratio's mean is 0.49 and the standard deviation is 0.025
Images attached to this report
805_20180611091315_s145cvscurrent.png 805_20180611092644_ophirvegavspm100ds145c.png 805_20180611093328_choppernochopperratio.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 11:47, Monday 11 June 2018 (804)Get code to link to this report
Aux1(Coherent control) laser first telescope

Participant : Eleonora, Marco, Yuhang and Matteo.

After put the first lens(f = 62.9mm), I checked the beam diameter by beam profiler. The result is 101um and 88um. According to the simulation, it should be 88um. It looks fine, so I procced.

For Faraday:

    1. Rotate lambda/4 to find the maximum, then rotate 180 degree, compare to find a better angle.(146 degree)

    2. Rotate lambda/2 to find the maximum.(279 degree)

    3. Before FI the power is 272mW, after FI the power is 253mW. The coupling is 93%.

After putting the second lens, we measured the beam is collimated and has a beam diameter as 2500um.

Then we rotate the half wave plate(163.5 degree), we have s-polarization 202mW while p-polarization 100uW. Means we have 99.95% s-polarization.

After put the 98:2 beam splitter, we got 1.9mW as transmission and 217mW as reflection. The ratio between them becomes 99 : 1. The ratio change is because of the polarization.

R&D (FilterCavity)
Print this report.
YuhangZhao - 15:38, Thursday 07 June 2018 (803)Get code to link to this report
Farady design for coherent control laser

For the purpose of having 2.4mm collimated beam in diameter(as required by entry 801), we did the design.

Images attached to this report
803_20180607120638_fidesign.png
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 11:46, Thursday 07 June 2018 (802)Get code to link to this report
Comment to recheck of laser parameter (Click here to view original report: 800)

This is for AUX 1, the laser that will be used for the coherent control. A first characterization was done and reported in entry 666. It seems compatible with the new measurement.