LOG-IN
Displaying reports 1361-1380 of 3128.Go to page Start 65 66 67 68 69 70 71 72 73 End
R&D (Cryogenic)
Print this report.
SatoshiTanioka - 22:58, Monday 16 March 2020 (2076)Get code to link to this report
Work on input optics

What I Did

  • Tweaking the lens and mirror position in a double-pass AOM path
  • Mode matching for TEM00 mode

Details

As I replaced a lens in double-pass AOM path, I tweaked the position of the lens and the mirror.
Current lens is f=70mm one.

Since the transmitted flash could not see due to the poor mode matching, I tried to improve the mode matching.
The lenses were replaced and f=-75mm and f=300mm ones are put now.
Their positions are about 0.226m and 0.659m from the beam waist, respectively.
After that I could get beam size about 65um at its waist where it is reasonable position for the apex mirror of folded cavity.

Next Step

  • Install some mirrors to inject the TEM00 beam into the chamber.
  • Alignment of the input and output mirrors.
  • Tweaking.
KAGRA MIR (Absorption)
Print this report.
SimonZeidler - 15:42, Monday 16 March 2020 (2074)Get code to link to this report
Absorption results on annealed S5

Pengbo, Simon

On the weekend, Pengbo finished the measurements on the annealed S5 sample.
Already last week, we recognized by doing the Z-scans that a large absorption excess must exist on one side of the sample (actually the side which has NO damages). So we decided to take actually 3 maps: in the center, on the suspicious surface and along the Z-axis. The results can be seen in the attachement.

As can be clearly seen, there is a very prominent structure on the suspicious surface which leads to an excess in the absoption data. This structure is partly visible also in the center-map, probably due to interference effects coming from the absorption excess (at least judging from the phase map). Please note that the sample for the center-map has been flipped so that the suspicious surface is on the out-going side because we had large problems in obtaining a meanigful result when the beam got influenced by the excess before reaching the targeted position.

As can be also seen from the map along the Z-axis, this excess is apparently not limited to the single surface area but has a depth which is hard to quantify (given the strong disturbance in phase and AC) but we estimate the affected depth to be several millimeter.

As a reference, the histogram for the absorption coefficient taken at the center is also given. The actual mean-value is not so far away from the measurements last year (however biased by the mentioned structure), which is at least one good news.

Images attached to this report
2074_20200316073857_20200314.png 2074_20200316073902_20200315.png 2074_20200316073906_20200316.png 2074_20200316073912_map20200314pengbo.png 2074_20200316073917_map20200315pengbo.png 2074_20200316073921_map20200316pengbo.png 2074_20200316073934_dist20200314pengbo.png
KAGRA MIR (Absorption)
Print this report.
SimonZeidler - 09:56, Thursday 12 March 2020 (2073)Get code to link to this report
Measuring absorption on annealed S1 and S5

Pengbo, Simon

We received the samples S1 and S5 back from France after their annealing.
Before we started measuring the absorption coefficient, we inspected the samples and we discovered some damages on the edges of both samples. Interestingly, the damages are only on one side (see attached photos).

The measuremetns started with S5 and are ongoing. However, before setting the sample into the sample holder, we cleaned it with FC on both sides.

Images attached to this report
2073_20200312015508_43.jpg 2073_20200312015515_24.jpg 2073_20200312015524_35.jpg 2073_20200312015528_15.jpg 2073_20200312015535_16.jpg
R&D (Cryogenic)
Print this report.
SatoshiTanioka - 23:26, Wednesday 11 March 2020 (2072)Get code to link to this report
Mode Matching

I tried to see the transmitted flash but I could not on Monday.
Indeed, the beam size was at its waist was much larger than the design value i.e., 100 um though 50 um is required.

I decided to change the mode matching lenses in order to achieve the designed spatial mode.
To be honest, the obtained minimun size was about 80 um radius.

I gonna try improving the mode matching until end of this week.

Note

The lenses are biconvex or biconcave in order to reduce the aberration of reflected beam.

R&D (FilterCavity)
Print this report.
YuhangZhao - 18:29, Wednesday 11 March 2020 (2071)Get code to link to this report
Measurement of QPD shot noise from filter cavity reflection with filter cavity locked

We performed the measurement which is close to the real case when we use QPD. The difference is only light is adjusted into only one segment of QPD.

We could see the modulation frequency we are interested in can be seen.

Images attached to this report
2071_20200611123305_figure3.png
R&D (Cryogenic)
Print this report.
SatoshiTanioka - 22:27, Monday 09 March 2020 (2070)Get code to link to this report
Alignment work and PDH servo modification

I tried to see the transmitted flash with scanning the laser frequency, but I could see nothing.
I decided to remove the input and output mirrors and do the alignment work again.

I also modified the PDH servo.
The schematic of the servo will be uploaded on wiki.

R&D (FilterCavity)
Print this report.
YuhangZhao - 16:01, Monday 09 March 2020 (2069)Get code to link to this report
Measurement of FDS after new FI

Pengbo and Yuhang

We performed the measurement although the mirror is not very stable. As you can see from the attached figure 2. This makes the low frequency measurement of shot noise(with the corruption of backscattering) much worse.

The measurement of squeezing level is only 2.5dB.

Images attached to this report
2069_20200309080127_figure11.png 2069_20200309080319_wechatimg641.jpeg
R&D (FilterCavity)
Print this report.
YuhangZhao - 15:10, Monday 09 March 2020 (2068)Get code to link to this report
Measurement of QPD shot noise with different beam size and different power(second measurement)

We measured again with 5mW green power, but this time we make the beam smaller to see if it will decrase or not.

The result is that shot noise will decrease if the power density is higher than 47mW/mm2.

There is a discrepancy between this measurement and the one measured last time(entry 2067). I think the reason is that, the beam size is highly related to position. And this time we removed QPD and we may put it back to a slightly different position. Then it leads to this discrepancy.

We need to use a more robust telescope if we want to have a more precise measurement.

Images attached to this report
2068_20200309071659_figure1.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 21:14, Thursday 05 March 2020 (2067)Get code to link to this report
Measurement of QPD shot noise with different beam size and different power

Pengbo and Yuhang

We measured shot noise spectrum with 5, 15, 26mW seperately. For each power, we measured also with different beam size ranging from to.

The set-up is using the second segement of QPD2. We checked DC voltage with oscillscope. We also checked RF channel after an amplification of 32dB with spectrum analyzer operates in 1MHz RBW(this time we average for 10 times so that the noise vurve is smooth). The noise spectrum of RF signal is plot for each case.

1. 5mW case(attached as the first picture): the noise floor is the same for all the beam size.

2. 15mW case(attached as the second picture): the noise floor reaches maximum when the power density is below ~25 mW/mm2

3. 26mW case(attached as the third picture): the noise floor reaches maximum when the power density is below ~22mW/mm2

We didn't measure the changing point for 5mW, however, from the beam density we measred, all the measurement we did for 5mW has low power density relative to the threshold (roughly between 22 and 25mW/mm2). We should see the noise floor decrease when the beam size is smaller than 500um.

Conclusion: The QPD response will saturate and decrease if the power density exceeds around 23mW/mm2.

(Notice: the beam size in this entry is diameter)

Images attached to this report
2067_20200305131521_5mw.png 2067_20200305131526_15mw.png 2067_20200305131531_26mw.png
KAGRA MIR (Absorption)
Print this report.
PengboLi - 14:02, Thursday 05 March 2020 (2066)Get code to link to this report
Absorption results on KAGRA-size Shinkosha#7 after setup upgrade

Simon, Pengbo
 

Attach to this report I show the result on shinkosha #7 sample with different polarization input beam(S-pol and P-pol).

We can see a 1 percent difference from these two maps, which is even smaller than the difference for TAMA-size sample.

Images attached to this report
2066_20200305060051_kagras1.png 2066_20200305060054_kagrap1.png 2066_20200305060059_kagras2.png 2066_20200305060102_kagrap2.png
KAGRA MIR (Absorption)
Print this report.
PengboLi - 18:42, Wednesday 04 March 2020 (2065)Get code to link to this report
Absorption measurement on TAMA#1 and Shinkosha #7

Simon, Pengbo

After the birefringence measurement, we change to the absorption system with a controllable polarization of the laser. First, we did two XY-plane absorption measurements on TAMA#1 with different polarization, which is P-pol and S-pol. Then we choose a small area of the mirror and did another measurement under the S-pol incident beam. We can see a very clear structure in the map. Then we did another YZ-plane absorption measurement with an S-pol incident beam. The distribution is quite homogenous.

We change the sample to Shinkosha #7, then follow what we did before, checking whether the polarization might have some influence on the absorption. We already have the result of the absorption map with the S-pol incident beam. The result is almost the same compare with the former result.

Images attached to this report
2065_20200304093548_tamap1.png 2065_20200304093602_tamas1.png 2065_20200304093606_tamal1.png 2065_20200304093612_tamap2.png 2065_20200304093617_tamas2.png 2065_20200304093621_tamal2.png 2065_20200304103301_tamayz.png 2065_20200304103308_figure1.png 2065_20200304103312_figure2.png
R&D (FilterCavity)
Print this report.
YuefanGuo - 05:56, Wednesday 04 March 2020 (2064)Get code to link to this report
Measurement of the QPD did in Virgo

We did the same measurement as in logbook entry 1875, the QPD response to different size of the green beam

The measurement was done by using the green beam reflected by the green mode cleaner of the Virgo squeezer(1500W). 

First, we measured the green beam size without any lenses and saved the data file from the beam profiler.  The beam profiler has 1928*1448 pixels corresponding to active are of 7.1mm*5.3mm, we got power value at each data point, and we did a 2-d Gaussian fit to these data points. One of the fit shows in figure 1. From the beam plotted in this figure, it is clear that the beam has astigmatism, so the final beam waist size and position are quite different in two axes. Below is the number (position zero is just a random point we chose easy for the measurement) 

 

Beam waist size

Beam waist position

X

242um

0.0924m

Y

367.6um

0.6262m

By checking the 2d plot of the raw data, we found out the beam profiler is saturated (fig 2 shows the top cut shape). But in the 2d fit, we were not able to remove these points by substituting them into 'NaN' while using 'lsqcurvefit' function in Matlab, because this function needs 'double' format input.  Then to check the quality of the fit, we plot the difference between the fit the original data, result shows in figure 3. It seems the fit is fine. Since the data is very noisy, I was guessing maybe we actually got the peak of the Gaussian, and the saturation part is just the noise. 

Anyway, after measuring the beam size, we put a 50mm lens and measured the beam with the QPD in different positions. We did two groups of measurements with different power by changing the density. the results show below.

Beam size

Group 1

Group 2

X direction (um)

Y direction(um)

DC (V)

RBW/VBW/MHz (dBm)

DC (V)

RBW/VBW/MHz (dBm)

35

23

1.35

-150.5

2.344

-150.6

148

256

1.35

-148.4

2.46

-148.7

277

496

1.309

-146.9

2.401

—145.6

407

738

1.307

-146.5

2.390

-144.8

538

979

1.307

-146.7

2.383

-144.7

668

1220

1.3

-146.6

2.360

-144.6

Images attached to this report
2064_20200303215458_2dfit30cm.jpg 2064_20200303215513_data2dplot.jpg 2064_20200303215541_resfitdata.jpg
R&D (FilterCavity)
Print this report.
YuhangZhao - 16:54, Tuesday 03 March 2020 (2063)Get code to link to this report
Measurement of filter cavity reflection beam height

Eleonora and Yuhang

We measured the filter cavity reflection beam height. It is 74.5mm, which is 1mm lower than injection beam measured three months ago.

Images attached to this report
2063_20200303085325_wechatimg639.jpeg
R&D (Cryogenic)
Print this report.
SatoshiTanioka - 14:31, Tuesday 03 March 2020 (2062)Get code to link to this report
Output mirror alignment

This entry is log on Mar. 2nd.

What I did

I aligned the output mirror of the folded cavity so that the reflected beam can be picked off at the FI.
The procedure is as follows.

  1. Install the output mirror.
  2. Check the reflected beam postion by a sensor card.
  3. Insert the shim sheet between the mirror holders to adjust the direction of mirror.

Next Step

Align the input mirror as the same procedure.
Then scan the laser frequency and monitor the transmitted beam.

R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 12:00, Tuesday 03 March 2020 (2061)Get code to link to this report
Comment to Shot noise and squeezing with additional faraday (Click here to view original report: 2059)

Note that back scattering on Saturday was particulary bad. It seemed the mirrors were more excited than usual.

By the way I remember that before Christmas, whitout the additional Faraday, we managed to measure less then 2 dB above shot noise at 30 Hz. See here

R&D (Cryogenic)
Print this report.
SatoshiTanioka - 20:43, Saturday 29 February 2020 (2060)Get code to link to this report
Folded cavity alignment

What I did

  • Replaced two pedestals used in the chamber
  • Removed the output fused silica mirror and re-aligned the beam. At this moment, the cavity direction was not good as the beam reflected by the apex mirror could not reach the center of the output mirror. So I tweaked the cavity direction.
  • Re-installed the output fused silica mirror and checked the transmitted laser power.

Some results

The transmitted laser power was about 1.03 uW with 6.4 mW input power.
Assuming no loss in the apex mirror , the transmittance of the fused silica mirror is estimated about 0.016%.
In practice, there are some loss in the apex mirror and the transmittance of the fused silica mirror is slightly larger than 0.016%.
According to the spec sheet, the transmittance of the fused silica mirror is 0.017%.
So this result is reasonable and the loss in the fused silica mirror is not too large, about 30ppm with the assumption that the reflectivity of the fused silica is same as shown in the spec sheet (99.98%)
In any case, I need to compose the cavity and see the flash.

Next step

  • Tweak the alignment of the output mirror to pick off the reflected beam. In fact, I could see the reflected beam on PD by adjusting the direction of the output mirror. Shim sheet, however, is indispensable to maintain the optimal direction.
R&D (FilterCavity)
Print this report.
NaokiAritomi - 18:20, Saturday 29 February 2020 (2059)Get code to link to this report
Shot noise and squeezing with additional faraday

[Aritomi, Yuhang, Eleonora]

We measured shot noise and squeezing with additional faraday before homodyne to reduce back scattering (attached picture). 

Difference between black and blue is back scattering with faraday which is much smaller than before. Difference between  blue and red curve might be effect of CC. 

Images attached to this report
2059_20200229102102_sqz20200229.png
Comments related to this report
EleonoraCapocasa - 12:00, Tuesday 03 March 2020 (2061)

Note that back scattering on Saturday was particulary bad. It seemed the mirrors were more excited than usual.

By the way I remember that before Christmas, whitout the additional Faraday, we managed to measure less then 2 dB above shot noise at 30 Hz. See here

R&D (Cryogenic)
Print this report.
SatoshiTanioka - 18:36, Friday 28 February 2020 (2057)Get code to link to this report
Folded Cavity Installation

I re-installed a folded cavity inside the chamber with one apex silicon mirror.
Then I aligned the 2 steering mirrors to pick off the reflected beam and I could pick off the beam from the chamber.
A PD was put the transmitted side to check the transmitted power when the cavity is composed.
After that, I installed a fused silica mirror as the output mirror and tried to see the transmitted beam power.
Some amount of stray light, however, prevented me to check the transmittance of the fused silica mirror.

I will install an input fused silica mirror and try to see the flash by scanning the laser frequency.

[note]
There is an optical loss at the viewport, about 10%.

R&D (FilterCavity)
Print this report.
YuhangZhao - 15:27, Wednesday 26 February 2020 (2056)Get code to link to this report
Temperature/humidity change and suspended mirror position drift monitor

The temperature change inside and outside the bench is monitored. At the same time, the suspended mirrors position is also monitored. 

The comparison is attached.

Also the seperate measurement is attached.

Images attached to this report
2056_20200226083744_07.png 2056_20200226084835_temhumidity2.png 2056_20200226084843_trend.png
R&D (FilterCavity)
Print this report.
EleonoraCapocasa - 13:52, Wednesday 26 February 2020 (2055)Get code to link to this report
New FI on the squeezing path: work on-going

In order to mitigate back reflection noise we plan to install an additional  Faraday isolator (FI-1060-5SC-HP) on the squeezing path in reflection from FC, at the edge of the optical bench. See pic 1 and 2.

To finalize the istallation we need to modify a bit the custom support: drill holes for the screws and reduce the hight of about 2.5 mm. (ATC will do it by next Wednesday afternoon)

The Faraday should bring additional 3% of propagation losses, the associated squeezing degradation (simulated in pic 3) doesn't seem too bad.

The drawback is that the faraday will prevent to close one side of the bench cover. But so far we have never closed it.

The beam waist is ~1mm and FI aperture is 5 mm in diameter, so clipping losses shoud be negligible.

Images attached to this report
2055_20200226054344_fi.png 2055_20200226054617_fipic.jpeg 2055_20200226054625_wof.png