LOG-IN
Displaying reports 1141-1160 of 3128.Go to page Start 54 55 56 57 58 59 60 61 62 End
R&D (FilterCavity)
Print this report.
YuhangZhao - 20:19, Thursday 17 December 2020 (2314)Get code to link to this report
Mixer test with saturated 5dBm LO for different RF power

As mixers need to be operated in saturation mode, I temporarily take the amplifier channel used for IRMC demodulation to amplify the signal from DDS. DDS provides about -6dBm signals. With the mentioned amplifier, LO was amplified to about 5dBm. At the same time, the RF signal was about -15dBm.

When we scan the CCFC phase with a sine wave, the demodulated signals will deviate from sine wave if the demodulation process has problems. So I did this test with 5dBm LO (shown in attached figure 1) for different RF power (-15dBm, -9dBm, -6dBm and -3dBm). These tests are in attached figures 2 to 5. All these figures seem to provide good shape demodulated signals (sinusoidal). From these figures, we could also see that the pk-pk signal also increases with the increase of RF power almost linearly (115mV, 206mV, 288mV, 380mV).

I also checked the CCFC error signals for these cases(figure 6,7,8). They are consistent with the error signals we found in elog2308. And apparently, better SNR is achieved with -3dBm RF power.

(We could add 12dB+12dB+3dB attenuator for the -3dBm signal to simulate a factor of ~25 decreases of CCSB power)

Images attached to this report
2314_20201217121831_1.jpeg 2314_20201217121836_2.jpeg 2314_20201217121847_3.jpeg 2314_20201217121853_4.jpeg 2314_20201217121859_5.jpeg 2314_20201217121913_15.png 2314_20201217121918_6.png 2314_20201217121924_3.png
R&D (FilterCavity)
Print this report.
NaokiAritomi - 23:55, Wednesday 16 December 2020 (2313)Get code to link to this report
Comment to Improved CCFC error signal (Click here to view original report: 2308)

I fitted the measured CCFC error signal by fitting the CC detuning (with respect to carrier), demodulation phase, and starting time (first plot). In this plot, misalignment effect is not considered.

In the second plot, I added the misalignment effect in theoretical curve by fixing the mode matching to 94%.

Images attached to this comment
2313_20201216155541_20201211ccfcfitting.png 2313_20210807181918_20201211ccfcfixedmismatch.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 13:24, Wednesday 16 December 2020 (2312)Get code to link to this report
Filter cavity length control characterization with CCFC

Matteo and Yuhang

The suppression of filter cavity length noise provides stable detuning, which is vital for the production of frequency dependent squeezing. The CCFC control loop is designed to achieve this goal.

To understand better how CCFC control works, several characterization works have been done recently. They are listed as follows:

1. Figure 1 shows many length error signals and noise curves. The addition of CCFC error signal introduces length noise for GR loop at low frequency. This is validated by figure 3 and 4. The GR+IR error signal doesn't change because the filter cavity length change doesn't change.

2. Figure 2 shows correction signals. For the correction signals send to the main laser or end mirror, they are the same whether there is CCFC or not. This is consistent with the unchange of IR+GR error signal.

3. Figures 3 and 4 show FC GR TRA/REF DC spectrums. CCFC causes the GR length noise increase, which translates into intensity noise. 

Images attached to this report
2312_20201216052558_errors.png 2312_20201216052604_correction.png 2312_20201216052739_influenceongrref.png 2312_20201216052745_influenceongrtra.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 14:24, Tuesday 15 December 2020 (2310)Get code to link to this report
FC_IR_TRA spectrum (AA on/off)

In elog2231 and elog2267, a worse locking accuracy was found to be caused by AA.

Today I compared the FC_IR_TRA while AA is on or off. It seems AA induced noise increase doesn't have the same shape with FC length noise (but similar).

This noise increase is clearly visible but could be well suppressed if CCFC lock is implemented.

Images attached to this report
2310_20201215062413_irtralengthnoise.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 15:59, Monday 14 December 2020 (2309)Get code to link to this report
Comment to Improved CCFC error signal (Click here to view original report: 2308)

12dB attenuator was added for RF signal (before the 32dB amplifier)

12dB attenuation was applied to LO signal (DAC current control was reduced from 1/2(-12dBm) to 1/8(-24dBm))

Current RF amp: -15dBm

Current LO amp: -24dBm

R&D (FilterCavity)
Print this report.
NaokiAritomi - 23:33, Friday 11 December 2020 (2308)Get code to link to this report
Improved CCFC error signal

[Aritomi, Yuhang, Matteo]

We found that we still had saturation problem of CCFC RF and LO so we reduced them.

Then we measured CCFC error signal with different CCFC demodulation phase (Pic. 1). AOM FM freq is 300 mHz and deviation is 2kHz, so AOM scan speed for IR is 4kHz/(5/3 s)/2 = 1.2 kHz/s. CCFC amplitude for normalization is 28 mV. The calibration factor of CCFC error signal is determined by fitting the blue curve around 0, which is 1191 Hz/V.

We locked CCFC with 70 deg and 250 deg CCFC demodulation phase (both are I phase, but sign is opposite) and compared the locking accuracy with CCFC lock (Pic. 2). We found that CCFC locking accuracy with 250 deg is smaller than 70 deg above 1kHz. Changing CCFC demod by 180 deg means that CCSB on resonance and off resonance are swapped. CCSB noise on resonance is filtered out by cavity pole while the noise of other CCSB is not. If noise of upper/lower CCSB are different, this noise difference can happen.

IR filter is 500 gain and 30 Hz low pass filter.

Anyway now CCFC locking accuracy is below 1Hz if the calibration factor is correct. Strange thing is that locking accuracy above 10kHz is much better than BAB locking accuracy with green lock.

Images attached to this report
2308_20201211153306_ccfc20201211slowscan.png 2308_20201211153315_ccfc20201211lockaccuracy.png 2308_20201217042405_20201211ccfcoltf.png
Comments related to this report
YuhangZhao - 15:59, Monday 14 December 2020 (2309)

12dB attenuator was added for RF signal (before the 32dB amplifier)

12dB attenuation was applied to LO signal (DAC current control was reduced from 1/2(-12dBm) to 1/8(-24dBm))

Current RF amp: -15dBm

Current LO amp: -24dBm

NaokiAritomi - 23:55, Wednesday 16 December 2020 (2313)

I fitted the measured CCFC error signal by fitting the CC detuning (with respect to carrier), demodulation phase, and starting time (first plot). In this plot, misalignment effect is not considered.

In the second plot, I added the misalignment effect in theoretical curve by fixing the mode matching to 94%.

R&D (FilterCavity)
Print this report.
YuhangZhao - 00:45, Thursday 10 December 2020 (2307)Get code to link to this report
CCFC error signal at different demodulation phase

Elog2300 described optimization for CCFC error signal. To characterize better these error signals, I put measured CCFC error signal as follows.

Figure 1 is CCFC error signal at different demodulation phase, after modematching optimization.

Figure 2 is CCFC error signal at different demodulation phase, after mixer optimization.

Images attached to this report
2307_20201209164548_aftermm.png 2307_20201209164557_aftermixer.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 00:11, Thursday 10 December 2020 (2306)Get code to link to this report
CCFC lock after optimization of mode matching and mixer

Matteo and Yuhang

As reported in elog2300, we optimized mode matching and mixer. We obtained a larger CCFC error signal after that. Then we used it to lock the filter cavity length for IR. Control loop information is summarized as follows:

  1. Gain of CCFC loop: 50
  2. Corner frequency of CCFC loop: 30Hz (one order low pass)
  3. Error signal shape: figure 1
  4. Open loop transfer function (only CCFC part): figure 2 (40mVpk-pk excitation used)
  5. Error signal spectrum (loop on/off): figure 3.
  6. Calibration for error signal: AOM speed (4000Hz/2.5s)(figure 4) divided by error signal slope around zero (75mV/43.7ms)(figure 1 and 5) divided by 2 (AOM scan green to IR) : 4000*43.7/2.5/75/2 = 466 Hz/V

We could see that the CCFC method stabilized length noise for IR below ~1kHz. The IR length noise reached 2.3Hz after closing the CCFC loop. Compared with the old case, we could see that the main difference in IR length noise is around 1kHz~10kHz. The reason for this difference is still unknown. But if the CCFC error signal can go back to the old case, the CCFC loop can reduce IR length noise to less than 1Hz.

Images attached to this report
2306_20201209161218_wechatimg49.jpeg 2306_20201209161234_oltf.png 2306_20201209161257_ccfcla.png 2306_20201209161314_wechatimg50.jpeg 2306_20201209161403_wechatimg51.jpeg
R&D (FilterCavity)
Print this report.
YuhangZhao - 14:16, Wednesday 09 December 2020 (2305)Get code to link to this report
CCFC error signal offset (after mixer optimization)

By changing CCFC demodulation phase, CCFC error signal offset should change in a sinusoidal way. I checked this after the optimization of mixer. The result is in attached figure 1.

Images attached to this report
2305_20201209061639_ccfcoffset.png 2305_20201209171030_ccfcdemodfit.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 11:42, Wednesday 09 December 2020 (2304)Get code to link to this report
Histogram of CCFC error signal (after mixer optimization) and CC1 error signal

I checked the histogram of CCFC and CC1 error signal. This check is after the mixer optimization.

We could see that strange behavior of histogram reported in elog2302 disappeared.

Images attached to this report
2304_20201209034344_ccfc.png 2304_20201209034351_cc1.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 11:31, Wednesday 09 December 2020 (2303)Get code to link to this report
Optimization of mixer (ZX05-1L-S+) (used for CCFC error signal demodulation)

Matteo and Yuhang

As reported in elog2289 and elog2302, the demodulated signal from mixer ZX05-1L-S+ has strange behaviors, such as not exactly sinusoidal or strange data distribution. We realized these issues but we didn't know what is the reason.

On 2020/22/07, we checked two quadrature-phase signals of CCFC error signal while CC1 phase is scanned more than 2pi. While checking, we found these two quadrature-phase signals were not the same. Attached figure 1 shows these two signals.From this figure, the quadrature-phase signal is quite similar with sinusoidal shape while the in-phase one is quite linear between each maximum and minimum. After observing this difference, we start to investigate what is the difference between these two channels.

Comparison of these two channels:

1. The RF signals come from the same PD, the LOs come from the same channel of DDS3

2. LO signal is splitted by ZMSCQ-2-90, RF signal is splitted by ZFDC-10-1-S+

3. They use the same mixer ZX05-1L-S+ and the same low pass filter SLP-1.9+

The splitting of LO makes one LO ~11dB smaller than the other one (The splitting of LO should give identical output. However, there is difference due to frequency issue.). The splitting of RF makes one RF ~10dB smaller than the other one. (RF signal is about -3dBm before splitting)

In the end, we found the problem comes from LO. We were using ~-6dBm LO, which is smaller than the datasheet requirement. However, in practice, this mixer needs even smaller LO (-12dBm LO is used now).

Images attached to this report
2303_20201209033357_ccfcscan.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 18:51, Tuesday 08 December 2020 (2302)Get code to link to this report
Histogram of CCFC error signal's offset

When CC1 is locked and the filter cavity is detuned, the CCFC error signal only shows an offset. This field should be identical with CC1 error signal if offset is not considered. Before the optimization of the mixer, we checked the histogram of this offset. From the attached figure, we could see that this histogram has some problems (no data located in the center). It could come from an oscillation of this signal.

We should recheck it after the optimization of the mixer. 

Images attached to this report
2302_20201208105006_ccfchist.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 18:38, Tuesday 08 December 2020 (2301)Get code to link to this report
Issues about the open-loop transfer function of filter cavity GR lock

The filter cavity GR lock's OLTF may differ with the filter cavity GR+IR lock's OLTF at low frequency. Therefore, we start to investigate GR OLTF's low-frequency part.

In the attached figure, there are four measurements. Their legends are listed in the sequence of time on 2020/12/07. We could see that:

1. All measurement shows flat gain at low frequency, which is different from what we expect.

2. Morning and evening measurements' magnitude are quite different at low frequency. The reason for this difference is still unknown.

3. Measurement phases are different with/without SR560 (just passing through without gain/filters). We could see that the phase margin is better if SR560 is used.

Images attached to this report
2301_20201208103851_grlock.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 17:21, Tuesday 08 December 2020 (2300)Get code to link to this report
Optimization of CCFC error signal

Matteo and Yuhang

Based on Aritomi-san's code, I add the degradation from mode-matching to the CCFC error signal. The simulation result is in attached figure 1. From this simulation, worse mode-matching makes CCFC error signal degrade around resonance. But mode-matching doesn't affect the CCFC error signal's offset.

Based on this simulation, we sent BAB to the filter cavity and checked the mode-matching was about 0.75. We found the IR drift happened only in the yaw direction. After optimizing yaw, mode matching increased to about 0.9. When we checked the CCFC error signal's pk-pk value, we found some issues with this signal's demodulation. After optimizing the mixer, we saw an even better CCFC error signal. The comparison of CCFC error signals before and after optimization is in attached figure 2.

I compared the mm-optimized/mm-original CCFC error signal's minimum. In the simulation, the ratio is 0.64. While in measurement, it is 0.58.

Images attached to this report
2300_20201208092226_mm.png 2300_20201208092244_optimization.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 00:38, Tuesday 08 December 2020 (2299)Get code to link to this report
Comment to TAMA PSD for PR pitch show excess noise (2020.12.07) (Click here to view original report: 2298)

We took a spare PSD and replaced the old one for PR Oplev. The spectrum of PSD was measured and shown in figure 1. We can see that the new PSD has higher noise than the reference. Apart from that, the new PSD also shows different peaks, which needs to be further examined.

Images attached to this comment
2299_20201208010912_prnoiseincrease.png
R&D (FilterCavity)
Print this report.
YuhangZhao - 15:15, Monday 07 December 2020 (2298)Get code to link to this report
TAMA PSD for PR pitch show excess noise (2020.12.07)

TAMA PSD for PR pitch show excess noise again, the situation is shown in the figure 1.

Images attached to this report
2298_20201207071542_wechatimg48.jpeg
Comments related to this report
YuhangZhao - 00:38, Tuesday 08 December 2020 (2299)

We took a spare PSD and replaced the old one for PR Oplev. The spectrum of PSD was measured and shown in figure 1. We can see that the new PSD has higher noise than the reference. Apart from that, the new PSD also shows different peaks, which needs to be further examined.

R&D (FilterCavity)
Print this report.
NaokiAritomi - 15:51, Monday 30 November 2020 (2297)Get code to link to this report
Comment to CCFC successfully locked (Click here to view original report: 2182)

I attached OLTF of CCFC and green lock. Note that I flipped the sign of measured data to match the measurement and theory. The measured phase is not consistent with theory.

Images attached to this comment
2297_20201130092424_20200827ccfcoltf.png
R&D (FilterCavity)
Print this report.
NaokiAritomi - 00:20, Friday 27 November 2020 (2296)Get code to link to this report
Optimization of CC PLL frequency for CCFC

In elog1727, I tuned CC PLL frequency from the fitting of CC separation frequency and CC PLL frequency, but the error of the fitting parameters is quite large with respect to optimal CC separation frequency 108 Hz. So this method is not precise to decide the correct detuning.

As written in elog2294, current CCFC error signal is not consistent with theoretical plot with optimal detuning, but instead it is similar to the theoretical plot with 25 Hz detuning.

If the current detuning is 25 Hz, we have to change the detuning by 29 Hz to obtain optimal detuning 54 Hz. Using the formula in elog1727, the CC PLL frequency has to be changed by 2*29 Hz/1.907605 =  30.41 Hz. Since the current CC PLL frequency is 6.99704303 MHz, optimal CC PLL frequency should be either 6.99707344 MHz or 6.99701262 MHz. I checked both cases by looking at CCFC error signal and found that 6.99701262 MHz is correct one (In DDS, 6.99701253 MHz was set). 

Here is the new CC PLL setting. I saved this setting as 20201126_dds3_CCFC.

channel   function frequency (MHz)                 binary number
CH0 CC PLL 20.99103760  1010 10111111 01010110 01011000
CH2/3 CC1/CCFC demod 13.99402518    111 00101010 00111001 10010000
      6.99701253      11 10010101 00011100 11001000

 

 

 

 

Attached plot shows CCFC error signal with different CCFC demodulation phase. Amplitude of the CCFC error signal is normalized with 83mV which is the amplitude of CCFC error signal when CCSB are off resonance of FC and CC1 is scanned.

Now the shape of CCFC error signal is similar to theoretical plot. In addtion to that, zero crossing point of blue curve in second plot is around 58Hz which is almost optimal detuning.

Images attached to this report
2296_20201126162236_20201126ccfcall.png 2296_20201127152518_20201126ccfc.png 2296_20201127152526_ccfcdemod.png
R&D (FilterCavity)
Print this report.
NaokiAritomi - 20:37, Thursday 26 November 2020 (2294)Get code to link to this report
Comment to CCFC error signal characterization (Click here to view original report: 2289)

CCFC error signal with 25 Hz detuning is very similar to the measurement.

Images attached to this comment
2294_20201126123653_20201125ccfc.png 2294_20201126123658_ccfc25hzdetuning.png
R&D (FilterCavity)
Print this report.
NaokiAritomi - 20:19, Thursday 26 November 2020 (2293)Get code to link to this report
Comment to Frequency tuning of coherent control sidebands (CCFC) (Click here to view original report: 1727)

Actually, the errors of the fitting parameters are -1907605 +/- 36859 and 13347486 +/- 257882. This error is quite large with respect to 108 Hz. We need to fine tune CC PLL frequency by looking at CCFC error signal.