z = 53.75mm, Pt = 4.85W : Thu, Apr 4, 2024 6-30-57 PM.txt
![Get code to link to this report](images/link.png)
edges at 48.5 and 79 mm : z_center = 63.75 mm
map at z_center : Thu, Apr 4, 2024 2-23-48 PM.txt
![Get code to link to this report](images/link.png)
I installed the 30mm diameter; 70mm length, 4/6 sapphire (T70#3) with pink tape at bottom towards IU.
Realigned DC
Pt = 26.3mW ; Pin = 30.7 mW
Pt = 4.87W ; long z scan (as for all previous z scan at x,y center ie 327,119 mm) : Thu, Apr 4, 2024 11-59-20 AM.txt
![Get code to link to this report](images/link.png)
z = 66.5mm, Pt = 4.83W : Wed, Apr 3, 2024 9-59-53 PM.txt
![Get code to link to this report](images/link.png)
I swapped the R and L cables for the OPO mixer.
During this process a cable for CC PLL monitor was quite damagedd but it's a bit tedious to replace alone and not priority for now so I left it as is for now.
I finalized the new box connections and little cable tidy up.
I could lock SHG after changing the DDS1 DAC1 phase from 20 to 240deg and the servo gain from 0.9 to 0.1 to avoid ringing. I measured about 30% HOM still present and 235mW of green produced.
The MZ/GRMC are heavily misaligned and I'll recover them tomorrow.
![Get code to link to this report](images/link.png)
I've installed my two supports, made using a 3D printer, for the LCD (from the screen of a monitor).
The beam of the setup is too low to path throw the LCD screen. So I added to more mirrors (2 inches) to get the beam higher, and I started to align them.
![Get code to link to this report](images/link.png)
z = 46.5m, Pt = 4.73W : Wed, Apr 3, 2024 5-28-27 PM.txt
![Get code to link to this report](images/link.png)
Installed T70#2 (30; 70 mm 5/6)
pink tape at bottom towardsiu
realigned DC
Pt = 27.4mW ; Pin = 31.5mW
long z scan Pt = 4.77W ; Wed, Apr 3, 2024 12-31-34 PM.txt
z edges at 36 and 77 mm : z_center = 56.5 mm
map at z_center : Wed, Apr 3, 2024 1-11-48 PM.txt
z = 46.5m, Pt = 4.73W : Wed, Apr 3, 2024 5-28-27 PM.txt
z = 66.5mm, Pt = 4.83W : Wed, Apr 3, 2024 9-59-53 PM.txt
![Get code to link to this report](images/link.png)
I finalized the mixers box and installed it in the bottom NIM rack.
A little issue is that the BNC mixers are a bit too large to close the box... However we still have quite some space in the NIM racks so I'll leave the box open for now.
I'm also taking this opportunity to change cables for more appropriate ones and passing them through the cable clamp.
One thing I'm not sure about is that OPO RF signal was sent to the 'L' port of one mixer which I thought was for LO..
I'll correct this.
I swapped the R and L cables for the OPO mixer.
During this process a cable for CC PLL monitor was quite damagedd but it's a bit tedious to replace alone and not priority for now so I left it as is for now.
I finalized the new box connections and little cable tidy up.
I could lock SHG after changing the DDS1 DAC1 phase from 20 to 240deg and the servo gain from 0.9 to 0.1 to avoid ringing. I measured about 30% HOM still present and 235mW of green produced.
The MZ/GRMC are heavily misaligned and I'll recover them tomorrow.
![Get code to link to this report](images/link.png)
Today I switched aircon of central building to cooling mode at 22degC.
Still need to be done for END.
![Get code to link to this report](images/link.png)
z = 47 mm ; Pt = 4.82 W ; Tue, Apr 2, 2024 6-27-00 PM.txt
![Get code to link to this report](images/link.png)
z = 67 mm, Pt = 4.81W : Tue, Apr 2, 2024 2-29-08 PM.txt
![Get code to link to this report](images/link.png)
this sample is t70#1.
Moved zIU to 40.01 mm
Pt = 25.5mW ; Pin = 30.0mW
There is definitely an issue with the holder : the sample is fixed by 2 set screws which can induce a slight tilt of the sample with respect to the input beam.... However there seems to exist a roll angle that compensate this...
Recentered DC with pink tape at bottom towards IU this time.
z scan with 4.81 W : Tue, Apr 2, 2024 10-17-23 AM.txt
z edges at 37 and 77 mm : center at 57mm
map at z_center : Tue, Apr 2, 2024 10-31-32 AM.txt
z = 67 mm, Pt = 4.81W : Tue, Apr 2, 2024 2-29-08 PM.txt
z = 47 mm ; Pt = 4.82 W ; Tue, Apr 2, 2024 6-27-00 PM.txt
![Get code to link to this report](images/link.png)
z = 54mm, Pt = 4.85W : Mon, Apr 1, 2024 11-58-05 PM.txt
this sample is t50#2
![Get code to link to this report](images/link.png)
DC was quite misaligned after changing sample..
after realignment
Pt = 26.1mW ; Pin = 30.0mW
rotated HWP to 28deg : Pt = 4.82W
long z scan : Mon, Apr 1, 2024 9-05-41 AM.txt
z edges at 29.5 and 58.5 mm : z_center = 44 mm
Pt = 4.84W
map at z_center : Mon, Apr 1, 2024 10-26-52 AM.txt
after visit break Pt~4.9W
z = 34 mm : Mon, Apr 1, 2024 4-44-36 PM.txt
z = 54mm, Pt = 4.85W : Mon, Apr 1, 2024 11-58-05 PM.txt
this sample is t50#2
![Get code to link to this report](images/link.png)
pt _ 4.65W
map at z_53 mm , Pt _ 4.65W; Sat, Mar 30, 2024 11-35-05 AM.txt
map at 33mm ' Sat, Mar 30, 2024 8-01-36 PM.txt
![Get code to link to this report](images/link.png)
laser : 7 A and HWP = 0deg : ~30mW
surface calib
z_iu = 69.18mm
z = 33 mm, (x,y) = (327.093,119.965)
checked calibration factor from 33 to 36mm, the max is at z=35mm ~13.9/W
z scan : Fri, Mar 29, 2024 11-29-17 AM.txt with Pt = 20.8mW (Pt = 21.4mW corresponds to Pin = 31.4mW)
bulk calibration
Pin = 31.1mW, Pt = 17.1mW to compute transmission
during measurement Pt = 17.5mW
moved z_iu by -0.32mm, z = 33.8mm gives R = 0.6413cm/W
z scan : Fri, Mar 29, 2024 2-27-42 PM.txt
T50#1 (30mm diameter, 50mm length, #4/6)
ziu = 48.61mm
dc was heavily misaligned in vertical... due to holder?
check if center is (x,y) 327,119 we can do +/-10mm in all direction without problem
also dc does not change during this so holder seems fine
Pt = 26.6mW, Pin = 31.2mW
can change z between 20 to 90 mm
HWP = 28deg : Pt = 4.83W
long z scan at x,y center : Fri, Mar 29, 2024 5-33-48 PM.txt
z edges are at 26 and 60 mm :Z_center = 43mm
10 mm radius map at z_center with 5mV gain and Pt = 4.81W : Fri, Mar 29, 2024 6-53-59 PM.txt
I placed a pink tape on top close to IU side
pt _ 4.65W
map at z_53 mm , Pt _ 4.65W; Sat, Mar 30, 2024 11-35-05 AM.txt
map at 33mm ' Sat, Mar 30, 2024 8-01-36 PM.txt
![Get code to link to this report](images/link.png)
Even though I already tried few days ago, this time unplugging/replugging the ethernet switch cables seems to have solve the ethernet connection issues of PCI and FC pcs.
![Get code to link to this report](images/link.png)
The data is being saved in the format below. Also, a header indicating the name of column has been added using "set dynamic attribute", and by enabling the header function in saving measurement to file.
Data | Column Number |
LC1 voltage | 0 |
LC2 voltage | 1 |
LC1 temperature | 2 |
LC2 temperature | 3 |
Mean Transmitted Power | 4 |
STD Tansmitted power | 5 |
Mean Main laser power | 6 |
STD Main laser power | 7 |
Mean Azimuth | 8 |
STD Azimuth | 9 |
Mean Ellipticity | 10 |
STD Ellipticity | 11 |
![Get code to link to this report](images/link.png)
In labview the averging filter was modified to include the standard deviation of data and also to save them simultaneously in file.
The data is being saved in the format below. Also, a header indicating the name of column has been added using "set dynamic attribute", and by enabling the header function in saving measurement to file.
Data | Column Number |
LC1 voltage | 0 |
LC2 voltage | 1 |
LC1 temperature | 2 |
LC2 temperature | 3 |
Mean Transmitted Power | 4 |
STD Tansmitted power | 5 |
Mean Main laser power | 6 |
STD Main laser power | 7 |
Mean Azimuth | 8 |
STD Azimuth | 9 |
Mean Ellipticity | 10 |
STD Ellipticity | 11 |
edges at 48.5 and 79 mm : z_center = 63.75 mm
map at z_center : Thu, Apr 4, 2024 2-23-48 PM.txt
z = 53.75mm, Pt = 4.85W : Thu, Apr 4, 2024 6-30-57 PM.txt
z = 73.75mm, Pt = 4.87W : Thu, Apr 4, 2024 10-28-30 PM.txt