Commissioning and Science Verification

Radio pointing offsets for single-beam receivers > Measurement

Revision history
  • 2016-09-01 - Kamazaki,T. - created
Table of contents

Preparations

  1. Login to COSMOS3 manager GUI with Group=asteXX (e.g. aste16) and Project=sbrfXX (e.g. sbrf16)
  2. Tune the receiver to be used
  3. Check Aspirin is in the WIND ON mode
    • Aspirin status can be confirmed by looking at the Aspirin Monitor window. Check the following three items.
      Current time = 2016/09/02 08:22:15    <--- The differences among these three times
      Aspirin time = 2016/09/02 08:22:14    <--- should be less than 60 seconds.
      Weather time = 2016/09/02 08:22:08    <---
      Aspirin Offset : (Az, El) = (   6.64,   0.42 )
      Aspirin status : ** WIND ON ** ( C22:on  C6:OFF WIND:on )    <--- should be "WIND_ON"
      Refraction Corr. status : OK                                 <--- should be "OK"
  4. Launch CONT qlook (qlookc) from cos3launcher
    • Make sure auto-pointing is disabled
    • Push the START button

Adjust initial pointing using either of the following observation tables

 
SaturnDASH345istx5d3.start
BAND8istx5b8.start
 

Check OPT images and start iOPT

  1. Check an OPT image
  2. Start iOPT on aste-c1c
    • iOPT calculates pointing offsets from optical images and records their results to files.

Measurement

  1. Run either of the following observation lists in the batch mode of the COSMOS3 manager GUI
     
    SaturnDASH345istd3.list
    BAND8istb8.list
     

Front page   New Page list Search Recent changes   Help   RSS of recent changes