MichaelPage - 16:23, Thursday 09 May 2024 (3556)
OPO replacement ATC beam setup
Logan, Michael
We once again found that the beam profile is showing some strange behavior with the asymmetry of the beam divergence. By using the profiler camera we see that it is probably due to misalignment of the Faraday isolator.
We took a Thorlabs IO-5-1064-HP from a box of miscalleneous optics for KAGRA filter cavity testing (next to Mitsuhashi Roberts linkage experiment, purchased end of FY23).
We put it on a micro tilt adjustment stage and made sure it could be adjusted to the beam height from the Mephisto (72-73 mm). Actually the Faraday isolator that was already there was just put on a pedestal post instead of an adjustment stage, so it is likely that the vertical alignment through the FI was not good, generating a visible higher order mode and causing the improper beam profile.