System recovery after Mitaka power outage

Eleonora, Matteo and Yuhang

All facilities and devices were recovered after the Mitaka power outrage.

The Mitaka power outage was done on 14th Nov., while the recovery work was done on 16th Nov.. Although everything was recovered, there were some issues we encountered during the recovery work.

1. The main power switch for the filter cavity arm lights were not found at the beginning. The reason is that this switch is located in the middle of filter cavity arm (I didn't know this at the beginning).

2. The sound issue of DDS. The solution is provided in elog1794.

3. To power on DAC, there is a DC voltage supply. At the beginning, I forgot to provide negative 18V. Then, while I increased the positive voltage, the maximum can reach only around 6.5V.

4. This time, none DDS channels flipped phase by 90deg.

5. The optimal temperature of SHG has been changed from 3.074 to 3.102. On the second day, the value changed back to 3.081.

6. The position of suspended mirror didn't change a lot. We don't need to change largly voltage offset sent to mirrors to recover FC flash.

7. PDH signal of FC lock has large 5kHz resonance even when FC is unlocked. It was figured out that this oscillation comes from SHG. And the problem is solved by reduce the gain of SHG to 2.1.

8. It was found that the input value of DGS doesn't change on medm. We checked medm sitemap/CDS. The situation is shown in the attached figure 1. It was figured out later that this is because I was using sine wave instead of square wave. After correcting to the good waveform, DGS worked again. The good sitemap/DGS is shown in the attached figure 2.

9. It was found that ratory pump in the middle arm has oil leakage.

Images attached to this report
2278_20201117075250_wechatimg40.jpeg 2278_20201117075315_good.png
Comments related to this report
MatteoLeonardi - 15:46, Tuesday 17 November 2020 (2279)
Some additional information regarding the recovery of the TAMA vacuum system after the power outage.

1) Two vacuum sensors were found not working. They are END station (arm side) and MID station (TAMA central side). The controller seems to work properly but it provide a "FAIL" message when reading the sensors.

2) The rotary pump in MID station has what it seems an oil leakage (see picture 1 and 2). I did not noticed this leakage when shutting it down on Friday but I did not payed particular attention to it, so I cannot tell if the problem was already present.

3) I could not find where to switch on the air dryer of South Arm. The West Arm air dryer are working.
RyutaroTakahashi - 12:57, Tuesday 24 November 2020 (2287)

Oil under the rotary pump was there from the old days. I replaced the rotary pump to new dry pump (ACP15). The TMP with the dry pump is working now at the mid point.