NaokiAritomi - 23:31, Thursday 23 January 2020 (1992)Get code to link to this report
FDS on 20200123

[Aritomi, Eleonora, Yuhang]

First we measured locking accuracy as shown in FIG. 1. Locking accuracy is 8 Hz which corresponds to 8.8 pm.

Then we changed green pump power from 50mW to 40mW so that we can reproduce FDS measurement before. OPO temperature and p pol PLL is as follows. 

green power (mW) 0 40
OPO temperature (kOhm) 7.2 7.2
p pol PLL (MHz) 300 190

Then we measured FDS around 400 Hz and 75 Hz detuning (FIG. 2 and FIG. 3). The parameters are as follows. Degradation budget is shown in FIG. 4.

It seems that locking accuracy is limiting squeezing level at low frequency.

sqz_dB = 16;                    % produced SQZ

L_rt = 150e-6;                    % FC losses

L_inj = 0.20;                     % Injection losses

L_ro = 0.11;                      % Readout losses

A0 = 0.05;                         % Squeezed field/filter cavity mode mismatch losses

C0 = 0.05;                         % Squeezed field/local oscillator mode mismatch losses

ERR_L =   8.8e-12;                  % Lock accuracy [m]

ERR_csi = 80e-3;                  % Phase noise[rad]

phi_Hom = [-4/180*pi, -27/180*pi];       % Homodyne angle [rad] 

det =  [423 446];                       % detuning frequency [Hz]

 

sqz_dB = 16;                    % produced SQZ

L_rt = 150e-6;                    % FC losses

L_inj = 0.20;                     % Injection losses

L_ro = 0.11;                      % Readout losses

A0 = 0.05;                         % Squeezed field/filter cavity mode mismatch losses

C0 = 0.05;                         % Squeezed field/local oscillator mode mismatch losses

ERR_L =   8.8e-12;                  % Lock accuracy [m]

ERR_csi = 80e-3;                  % Phase noise[rad]

phi_Hom = 90/180*pi;       % Homodyne angle [rad] 

det =  75;                       % detuning frequency [Hz]

 

Images attached to this report
1992_20200123153419_lockaccuracy20200123.png 1992_20200123154608_fds40020200123.png 1992_20200123154614_fds7520200123.png 1992_20200123154620_degradation20200123.png
Comments related to this report
EleonoraCapocasa - 11:41, Friday 24 January 2020 (1994)

Given the number Aritomi-san used (which are maybe too optimistic about MM), if we reduce the amount of initial squeezing  from 16 dB to 6 dB. The effect of lock accuracy is much reduced and in priciple we should be able to go below shot noise.