Warning: Undefined variable $s in /wwwsto01/wwwusers/gw-elog/osl/classes/DAO.php on line 959
NAOJ GW Elog Logbook
LOG-IN
Displaying report 1-1 of 1.
R&D (FilterCavity)
Print this report.
NaokiAritomi - 23:04, Thursday 24 October 2019 (1769)Get code to link to this report
IR locking accuracy

[Aritomi, Yuhang, Yaochin, Eleonora]

We measured IR error signal of filter cavity using BAB with TAMA PD which is same as CC1 PD in the reflection path from filter cavity. Pic. 1 shows IR error signal and IR transmission when filter cavity is scanned with AOM (setting is Pic. 2). Calculation method of calibration factor is same as entry 750.

Calibration:
peak-peak voltage of error signal: 26.2 mV
peak-peak time difference of error signal: 98 ms
slope = peak-peak voltage of error signal/peak-peak time difference of error signal = 26.2 mV/98 ms = 0.267 V/s
AOM scan frequency: 500 mHz
AOM deviation: 1 kHz
AOM scan speed for green: AOM deviation/AOM scan frequency = 2 kHz/s
AOM scan speed for IR: 2kHz/s /2 = 1 kHz/s
 
Calibration factor:
AOM scan speed for IR (Hz/s)/(2*slope(V/s)) = 1000(Hz/s)/(2*0.267(V/s)) = 1873 Hz/V

Note that factor of 2 in calibration factor is because slope of error signal on resonance is as twice as slope of peak-peak of error signal.

Then we measured spectrum of IR error signal when IR is locked on resonance and calculated rms (Pic.3). IR locking accuracy is 4.2 Hz which corresponds to 4.5 pm of locking accuracy. Pic. 4 shows squeezing degradation with this locking accuracy. In addition to mode mismatch, locking accuracy is also dominating. We need IR locking.

Images attached to this report
1769_20191024160409_error.jpg 1769_20191024160414_aom.jpg 1769_20210724223331_lockingaccuracy20191024.png 1769_20210724223338_degradation20191024.png
Comments related to this report
EleonoraCapocasa - 13:21, Sunday 27 October 2019 (1771)

Most of the RMS is accumualted below 10 Hz. We observed a correlation between the PDH error signal for IR (demodualtion of 15 Mhz sidebands) and the lock correction signal from RAMPEAUTO.

At the time of the measument we know the IR alignment was not good. This can be the cause of such behaviour.  We should repeat this measurement in good alignment condition.