EleonoraCapocasa - 13:18, Wednesday 16 October 2019 (1742)Get code to link to this report
H1 INPUT mirror coil doesn't respond but coil driver is fine

Yesterday I investigated the issue of not responding coil of INPUT mirror.

I made some tests by swapping the coil driver channels and I concluded that the coil driver is working fine and the problem is inside the vacuum chamber.

The not responding coil is H1 (top coil). This unbalance in the pitch driving will excite length. This can be maybe counteracted with horizontal coils. 

We should consider if to open the chamber for further investigation/coil repairing. 

Images attached to this report