KAGRA MIR (Absorption)
Print this report.
ManuelMarchio - 20:14, Friday 26 June 2015 (117)Get code to link to this report
Strange AC noise

In order to check if the spiky noise decreases after placing the clean booth,  long time acquisition of AC noise was performed (1 hour).

I couldn't find spikes, but we see a strange beaviour: the noise increase with time to a very high level (1mV std). Then another acquisition of 5 minutes was performed and it can be seen the noise suddenly decrease alone at a certain point. I'm trying to find a reason to this strange behaviour. The booth fan are ON, I think they never stop alone. Maybe some fan of the air conditioning or other noisy devices, I don't know.

------------------------------------------------------------------------------------------------------------------------------------

I used a vacuum cleaner to clean the floor inside the clean booth and around it. The floor is not smooth, it's not easy to remove all the dust. 

Images attached to this report
117_20150626125842_probeon1hour.jpg 117_20150626125850_probeon5min.jpg
Comments related to this report
ManuelMarchio - 17:14, Monday 29 June 2015 (119)

Today (Monday 29th) I acquired an hour of noise in the exactly same condition and I found 18microV of noise and no spiky noise at all. The only thing that changed since last time is that the system has been at rest for the weekend, about 48 hours.