KAGRA MIR (Absorption)
ManuelMarchio - 19:06, Friday 24 August 2018 (956)
Imaging unit alignment, acoustic noise reduction, reference samples scan and tama-size Sapphire scan

- Aligned the imaging unit. Using a blade at about 1cm from the crossing point. I moved the lens on the imaging unit until I saw a sharp image of the blade at the detector position.

- Using the surface reference sample I aligned the pump with the probe to maximize the signal. Then I maximized the signal finely adjusting  the position of the whole imaging unit.

- I noticed that there was some noise (around 300uV in the AC signal) with a constant phase. It came from the chopper vibration. So I put some sound absorbing panels and the noise reduced to about 50uV.

- I made a scan of the bulk reference sample with a pump power of 32mW

- I made a scan of the tama-size sapphire and I got the same absorption as before changing the pump size. 

- Compared the scans of the bulk reference sample with small pump and large pump. They give the same value (unexpected) but they have a different shape. To be understood.

Images attached to this report
956_20180824115832_34.png 956_20180824115837_16.png 956_20180824115845_23.png 956_20180824120022_bulkrefpump80pump36.png 956_20180824120536_tama2pump80pump36.png 956_20180824120921_a1.jpg 956_20180824120926_a2.jpg 956_20180824120931_a3.jpg 956_20180824120934_a4.jpg 956_20180824120938_a5.jpg 956_20180824120941_a6.jpg