KAGRA MIR (Absorption)
ManuelMarchio - 23:49, Wednesday 09 May 2018 (764)
Problem with the SURFACE reference sample

Before measuring the coating absorption of LMA samples and crystalline coating I wanted to test the surface reference sample in order to check if it had any damage. So I made a map of it.
The map shows a regular pattern of absorption that oscillates by a factor of 2.

I made some checks to understand the nature of this strange behavior. I report the steps in the attached pdf slides.

I'm going to look for a better alignment

Images attached to this report
764_20180509132306_surfrefproblem.png 764_20180509164639_43.png 764_20180509164644_30.png 764_20180509164648_44.png 764_20180509164655_26.png 764_20180509164701_35.png 764_20180509164706_45.png 764_20180509164714_00.png 764_20180509164720_57.png 764_20180509164725_53.png 764_20180509164730_18.png 764_20180509164738_36.png 764_20180509164809_41.png 764_20180509164813_43.png
Non-image files attached to this report