R&D (FilterCavity)
YuhangZhao - 19:01, Thursday 15 March 2018 (700)
Measurement of IR calibration factor

We measured the cavity bandwidth and error signal for IR with three different velocity of frequency scan. In this case, we can give a reasonable estimation of the calibration factor of IR. The result is 180.7 +/-5 Hz/V. We didn't consider error of invidual measurement, the standard deviation only comes from these three measurements results.

velocity bandwith calibration
200Hz/s 106Hz 176Hz/V
400Hz/s 112Hz 186Hz/V
80Hz/s 108Hz 180Hz/V

For a better parameter estimation, we will fit these measurment result.

Images attached to this report
700_20180315105946_ca1.png 700_20180315105951_ca2.png 700_20180315105957_ca3.png