For temperature control of LC we had implemented the PID control. But, before PID implementation, there was a feedback setup. Here, we were switching off the temperature control for 2s if the current applied was overshooting. But, this was creating an issue that the labview would consequently not save data or create a nan value in file.
We should have removed this after PID was implemented. In any case, it has been removed. Indeed the nan values or empty data don't appear anymore in labview graphical plots.
We will save retardation of LC tomorrow and finalise this conclusion.
only the part where the controller was switching off was removed. The part where we set the current of the controller at the time of overshooting still remains. Also, the controller is designed to remain switched off in case the current overshoots. So, the reenabling part remains as well.