Birefringence mitigation (General)
MarcEisenmann - 15:14, Wednesday 15 March 2023 (3203)
Comment to LC characterization and fit (Click here to view original report: 3202)

For reference I attach to this entry the retardation fit with and without taking into account the temperature of the LC.

Note that here and in previous entry I used the mean of the 3 data taken at every voltage step.

Taking into account the temperature in the fit reduce the residual rms by a factor 2 and the peak to peak by 50%.