KAGRA MIR (Absorption)
MarcEisenmann - 15:59, Thursday 17 March 2022 (2878)
annealed AZTEC absorption measurement started

Katsuki, Marc

This morning we checked again the pump beam profile to make sure the earthquake did not affect its properties.

It was fine so we checked the surface calibration (R = 16.64 at z = 35 mm and z_IU = 68mm) and bulk calibration (R_bulk = 0.6909 cm/W).

We inspected the sample with the strong green light and cleaned it with the ion gun.

We checked the centering and got X_center = 326.99mm and Y_center = 122.709 mm.

We increased the laser power to about 7.5 W and did a long z scan from which we got the 2 surfaces at z = 41.34 mm and 76.02mm making Z_center = 58.68 mm.

We could also estimate the absorption to be about 70 ppm/cm (roughly same as previous measurement).

We started a XY absorption measurement at Z_center with 0.25 mm step size, 0.5s waiting time and 10 order average/median filters that will last for about 16 h 30.