YuhangZhao - 18:42, Thursday 15 July 2021 (2624)
A comparison of FDS least square fit and mcmc fit (the same four parameters are set free in both cases)

To compare least square fit and mcmc fit in a fair way, it is necessary to make both of them have both four parameters free with the four parameters defined in elog2618.

The information of mcmc fit has been already summarized in elog2618. The fit of least square information is summarized in the attached four figures.

Figure 1 and 2 are FDS with detuning ~200Hz. Figure 3 and 4 are FDS with detuning ~70Hz.

The least square fit gives similar result with mcmc if detuning is around 200Hz. However, the least square fit gives not-expected and seems-unresonable result as figure 3 and 4. By just changing the fitting method from least square to mcmc, we extract information more precisely and more reasonably.

Images attached to this report