KAGRA MIR (Absorption)
MarcEisenmann - 13:57, Thursday 13 May 2021 (2492)
Comment to KASI sample absorption measurement (Click here to view original report: 2489)

Abe, Marc

Following the 3 maps measurements we performed again the bulk calibration where this time we moved the imagining unit by 0.32 mm in order to compensate the thickness difference between surface and bulk reference samples. We got the following result (also see last figure of this entry) :

AC_bulkref = 0.0731;
DC_bulkref = 4.164;
ACDC = 0.01755;
P_in = 29.5e-3;
P_t = 13.3e-3;
T_bulkref = P_t/P_in;
abs_bulkref = 1.04;
R_bulk = AC_bulkref/(DC_bulkref*sqrt(T_bulkref)*P_in*abs_bulkref) = 0.852 cm/W

Using this new calibration factor and using :

P_t = 6.25 W;
P_in = 7.322 W;

the absorption of this sample seems to be around 60 ppm/ cm for all 3 maps (see attached figures)

Today we will double check the bulk calibration as the change was quite larger than expected.

Images attached to this comment
2492_20210513065451_xy.jpg 2492_20210513065457_yz.jpg 2492_20210513065525_xz.jpg 2492_20210513065737_screenshotfrom20210513135724.png