R&D (FilterCavity)
YuhangZhao - 12:50, Tuesday 20 October 2020 (2243)
Calibration of AA spectrum (Input/End pitch/yaw) with oplev

To compare oplev and AA spectrum, the first step is to calibrate both of them. The calibration is done as follows:

1. Calibrate oplev signal. The oplev signal calibration method was developed by Eleonora and written in elog1874

2. Clibrate AA signal. The 4Hz sine wave was sent to Input/End pitch/yaw. By adjusting calibration factor of AA signal, AA's 4Hz peak was matched to oplev's peak. Then the calibration of AA signal was decided. The comparison is shown in the attached four figures.

  oplev AA
input pitch 0.038 0.018
input yaw 0.027 0.013
end pitch 0.026 0.017
end yaw 0.018 0.013

unit: urad/counts

Images attached to this report
2243_20201020071114_ipcali.png 2243_20201020071120_iycali.png 2243_20201020071127_epcali.png 2243_20201020071132_eycali.png