R&D (FilterCavity)
EleonoraCapocasa - 08:52, Thursday 30 January 2020 (2006)
Frequency dependent squeezing below shot noise at 70-100 Hz

[Aritomi, Yuhang, Raffaele, Matteo, Eleonora]

I did a quick fitting of the FDS measuremnt we did last friday (24/01).

We have about 1 dB of FDS sqz at low frequency and 3.5 dB at high frequency.

We see that at some point during the measurement the detuning changed from ~100 Hz to ~70 Hz.

Sqz degradation paramenters used for the fit:

sqz_dB = 8;                          % produced SQZ
L_rt = 150e-6;                      % FC losses
L_inj = 0.20;                         % Injection losses
L_ro = 0.11;                          % Readout losses
A0 = 0.05;                             % Squeezed field/filter cavity mode mismatch losses
C0 = 0.05;                             % Squeezed field/local oscillator mode mismatch losses
ERR_L =   8.8e-12;               % Lock accuracy [m]
ERR_csi = 80e-3;                 % Phase noise[rad]

Images attached to this report
2006_20200130005240_simvsfds.png