R&D (Cryogenic)
SatoshiTanioka - 12:01, Friday 22 November 2019 (1869)
Comment to The behavior of vacuum gauge (Click here to view original report: 1863)

I checked the output voltage from the vacuum gauge and it was 0.2V.
This indicates that there is an error in the B-A gauge or the sensor.

I did degas, but this not change the situation.

One possible solution is to clean up the gauge.