R&D (Cryogenic)
SatoshiTanioka - 09:28, Tuesday 19 November 2019 (1853)
Comment to Vacuum test (Click here to view original report: 1849)

This mornig the pressure was 9.8*10-1 Pa, though it was 5.0*10-3 Pa yesterday...
I am suspecting the residual water caused the problem.

I took a data during thus vacuum test.
The measured output voltage from ANELVA M-601GC has a relationship between the pressure which is described as

P = 102*V-12,

where P is the pressure and V is the output voltage.
I roughly checked the pressure using this relation and it was like 10-6 Pa, though the displayed pressure was ~10-3 Pa.
The output voltage was not correct for some reason,

I will check the measured voltage and do the test again without venting.