R&D (FilterCavity)
YuhangZhao - 00:39, Wednesday 11 September 2019 (1624)
comparison of PR optical lever spectrum with different yaw position of PR

Yuhang and Yuefan

According to the experience of Eleonora and Matteo, the yaw of PR has a very narrow range. Also, we moved -2000 counts of PR and we observed a stronger vibration of green beam. So we guess maybe PR is touching the intermedium mass.

So we checked how much we moved and them tried to move it little by little and then have a look the spectrum to see if there is any difference of peak values. The measurement was down from -1000 to 2000 counts while the original position is 750 counts. The spectrum is attached in the sequence from -1000 to 2000.

We could see that after 1300, the high frequency noise spectrum is a bit higher than the reference. And we could see at -200 and -1000, the spectrums have parts lower than the reference.

Images attached to this report
1624_20190910173023_20190910pr1000.png 1624_20190910173029_20190910pr200.png 1624_20190910173037_20190910pr200.png 1624_20190910173048_20190910prini.png 1624_20190910173054_20190910pr980.png 1624_20190910173059_20190910pr1300.png 1624_20190910173105_20190910pr1700.png 1624_20190910173109_20190910pr2000.png