R&D (FilterCavity)
YuhangZhao - 19:07, Saturday 29 June 2019 (1427)
Dark noise of coherent control 1 loop PD, error signal noise spectrum (free running) and optomechanical transfer function

Aritomi, EleonoraP, Yuhang

As described in the title, we characterized more about coherent control loop 1.

The dark noise is measured when there is not light. And other situation is the same(for example, PD is on, demodulation signal is on). The measurement point is error signal.

The free running error signal noise spectrum is measured after we checked that green phased shifter scanning and coherent error signal was fine.

Images attached to this report
1427_20190629120651_cc1darkerror.png 1427_20190629120702_cc1omtf.png
Non-image files attached to this report