R&D (FilterCavity)
EleonoraPolini - 18:01, Thursday 18 April 2019 (1311)
IR injection and reflection telescope update

I did another simulation for injection (-400 mm focal is not a common lens) and for the reflection (avoiding to change the already installed injection telescope into the homodyne).

The robustness for the injection telescope is really good, less than 3% moving the first lens in a range of +/- 5mm and less than 10% for the other one.

The robustness for the reflection telescope is not as good, we reach also 20% mismatch for +/- 5mm movement of one lens.

Images attached to this report
1311_20190418105921_injectionfin.png 1311_20190418105932_reflectionfin.png 1311_20190418105954_schemafin.png 1311_20190418110011_mismatchinjection1stlensfin.png 1311_20190418110021_mismatchinjection2ndlensfin.png 1311_20190418110036_mismatchreflection1stlensfin.png 1311_20190418110048_mismatchreflection2ndlensfin.png