R&D (FilterCavity)
YuhangZhao - 15:17, Monday 08 April 2019 (1279)
SHG reflection noise spectrum comparison between old and new servo

I set up a PD for SHG reflection. It can directly measure the green light noise spectrum. Then I did the measurement. The result is shown in the attached figure.

The new servo shows almost 10 times smaller noise at low frequency. While at high frequency(above 2kHz), the noise is five times higher than the old servo. The reason can be that we put notch around here.

Images attached to this report