R&D (FilterCavity)
NaokiAritomi - 03:52, Wednesday 27 March 2019 (1261)
Estimation of loss and phase noise
 
[Aritomi, Eleonora P., Matteo]
 
We measured squeezing and anti-squeezing with different green power.
 
Setting
center frequency: 100kHz
RBW: 1kHz
VBW: 30Hz
Sweep time: 2s
Scan of green phase: 2Hz with 1Vpp
 
Attached figure shows measured squeezing and anti-squeezing with different green power and calculated squeezing and anti-squeezing when loss = 30 %, phase noise = 150 mrad.
 
Although measured squeezing level is not precise due to phase noise, for sure squeeze level increased with lower pump power (lower anti squeezing). 
We confirmed that the reason why we don't have squeezing with 49mW green is phase noise , not too much loss.
 
Next step:
- Check the error signal of OPO reflection with new PD
Attached figure shows measured squeezing and anti-squeezing with different green power and   
calculated squeezing and anti-squeezing when loss = 30%, phase noise = 300mrad.
Images attached to this report
1261_20190730055222_sqzasqz.png