R&D (FilterCavity)
EleonoraCapocasa - 12:02, Wednesday 13 March 2019 (1249)
CC PLL pick-off realigned into fiber

[Aritomi, 2*Eleonora]

We have realigned the CC and ML pick-off into the fibers:

ML: 1.4 mW → 0.19 mW, coupling: 0.19*2/1.4 = 27%

CC: 3.3 mW → 0.5 mW, coupling: 1/3.3 = 30%
 
The results are similar to those of entry #1200.
 
We updated the wiki section about bench operation with this new values